Cargando…

Intra-Cardiac versus Transesophageal Echocardiographic Guidance for Left Atrial Appendage Occlusion with a Watchman FLX Device

This study aimed to compare the peri-procedural success and complication rate within a large registry of intra-cardiac echocardiography (ICE)- vs. transesophageal echocardiography (TEE)-guided left atrial appendage occlusion (LAAO) procedures with a Watchmann FLX device. Data from 772 LAAO procedure...

Descripción completa

Detalles Bibliográficos
Autores principales: Pastormerlo, Luigi Emilio, Tondo, Claudio, Fassini, Gaetano, Nicosia, Antonino, Ronco, Federico, Contarini, Marco, Giacchi, Giuseppe, Grasso, Carmelo, Casu, Gavino, Romeo, Maria Rita, Mazzone, Patrizio, Limite, Luca, Caramanno, Giuseppe, Geraci, Salvatore, Pagnotta, Paolo, Chiarito, Mauro, Tamburino, Corrado, Berti, Sergio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10607018/
https://www.ncbi.nlm.nih.gov/pubmed/37892796
http://dx.doi.org/10.3390/jcm12206658
Descripción
Sumario:This study aimed to compare the peri-procedural success and complication rate within a large registry of intra-cardiac echocardiography (ICE)- vs. transesophageal echocardiography (TEE)-guided left atrial appendage occlusion (LAAO) procedures with a Watchmann FLX device. Data from 772 LAAO procedures, performed at 26 Italian centers, were reviewed. Technical success was considered as the final implant of a Watchmann FLX device in LAA; the absence of pericardial tamponade, peri-procedural stroke and/or systemic embolism, major bleeding and device embolization during the procedure was defined as a procedural success. One-year stroke and major bleeding rates were evaluated as outcome. ICE-guided LAA occlusion was performed in 149 patients, while TEE was used in 623 patients. Baseline characteristics were similar between the ICE and TEE groups. The technical success was 100% in both groups. Procedural success was also extremely high (98.5%), and was comparable between ICE (98.7%) and TEE (98.5%). ICE was associated with a slightly longer procedural time (73 ± 31 vs. 61.9 ± 36 min, p = 0.042) and shorter hospital stay (5.3 ± 4 vs. 5.8 ± 6 days, p = 0.028) compared to the TEE group. At one year, stroke and major bleeding rates did not differ between the ICE and TEE groups. A Watchmann FLX device showed high technical and procedural success rate, and ICE guidance does not appear inferior to TEE.