Cargando…

Accuracy of the Injection-Based Occlusion Tool Utilizing Saline and Glucose Solution in Cryoballoon Ablation Guided by a Novel Dielectric Imaging System

Introduction: The aim of this study is to assess the accuracy of the injection-based occlusion (IBO) tool utilizing saline and glucose solution in verifying pulmonary vein (PV) occlusion during cryoballoon ablation guided by a novel dielectric system (KODEX–EPD system). Methods: In this retrospectiv...

Descripción completa

Detalles Bibliográficos
Autores principales: Mo, Binfeng, Yuan, Jiali, Lian, Xiaoming, Cai, Xingxing, Wang, Qunshan, Li, Yigang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10607856/
https://www.ncbi.nlm.nih.gov/pubmed/37887874
http://dx.doi.org/10.3390/jcdd10100427
_version_ 1785127641843499008
author Mo, Binfeng
Yuan, Jiali
Lian, Xiaoming
Cai, Xingxing
Wang, Qunshan
Li, Yigang
author_facet Mo, Binfeng
Yuan, Jiali
Lian, Xiaoming
Cai, Xingxing
Wang, Qunshan
Li, Yigang
author_sort Mo, Binfeng
collection PubMed
description Introduction: The aim of this study is to assess the accuracy of the injection-based occlusion (IBO) tool utilizing saline and glucose solution in verifying pulmonary vein (PV) occlusion during cryoballoon ablation guided by a novel dielectric system (KODEX–EPD system). Methods: In this retrospective study, we enrolled 34 consecutive patients with paroxysmal atrial fibrillation (AF) who underwent their initial cryoballoon ablation procedure guided by the KODEX-EPD system. PV occlusion was firstly assessed by the IBO tool utilizing saline or glucose solution and then verified by direct contrast angiography. Patients were divided into two groups according to the fluid used in the IBO tool: the Saline Group and the Glucose Group. Results: The overall procedure time and fluoroscopy time were comparable between the Saline Group and the Glucose Group (113.7 ± 18.3 vs. 108.4 ± 15.9 min; p = 0.375 and 10.1 ± 3.7 vs. 9.3 ± 3.5 min; p = 0.559). The IBO tool was utilized a total of 138 times in the Saline Group and 135 times in the Glucose Group. When assessing PV occlusion, the IBO tool using saline demonstrated a sensitivity of 92.6% and a specificity of 95.2% compared to angiography. Similarly, the IBO tool utilizing glucose solution showed a sensitivity of 93.2% and a specificity of 96.1%. Conclusions: The IBO tool utilizing non-contrast fluid, saline and glucose solution, demonstrates a high level of sensitivity and specificity in accurately predicting PV occlusion during cryoablation procedures. Both the saline and glucose solutions used in the IBO tool show promising results in effectively assessing PV occlusion.
format Online
Article
Text
id pubmed-10607856
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-106078562023-10-28 Accuracy of the Injection-Based Occlusion Tool Utilizing Saline and Glucose Solution in Cryoballoon Ablation Guided by a Novel Dielectric Imaging System Mo, Binfeng Yuan, Jiali Lian, Xiaoming Cai, Xingxing Wang, Qunshan Li, Yigang J Cardiovasc Dev Dis Article Introduction: The aim of this study is to assess the accuracy of the injection-based occlusion (IBO) tool utilizing saline and glucose solution in verifying pulmonary vein (PV) occlusion during cryoballoon ablation guided by a novel dielectric system (KODEX–EPD system). Methods: In this retrospective study, we enrolled 34 consecutive patients with paroxysmal atrial fibrillation (AF) who underwent their initial cryoballoon ablation procedure guided by the KODEX-EPD system. PV occlusion was firstly assessed by the IBO tool utilizing saline or glucose solution and then verified by direct contrast angiography. Patients were divided into two groups according to the fluid used in the IBO tool: the Saline Group and the Glucose Group. Results: The overall procedure time and fluoroscopy time were comparable between the Saline Group and the Glucose Group (113.7 ± 18.3 vs. 108.4 ± 15.9 min; p = 0.375 and 10.1 ± 3.7 vs. 9.3 ± 3.5 min; p = 0.559). The IBO tool was utilized a total of 138 times in the Saline Group and 135 times in the Glucose Group. When assessing PV occlusion, the IBO tool using saline demonstrated a sensitivity of 92.6% and a specificity of 95.2% compared to angiography. Similarly, the IBO tool utilizing glucose solution showed a sensitivity of 93.2% and a specificity of 96.1%. Conclusions: The IBO tool utilizing non-contrast fluid, saline and glucose solution, demonstrates a high level of sensitivity and specificity in accurately predicting PV occlusion during cryoablation procedures. Both the saline and glucose solutions used in the IBO tool show promising results in effectively assessing PV occlusion. MDPI 2023-10-17 /pmc/articles/PMC10607856/ /pubmed/37887874 http://dx.doi.org/10.3390/jcdd10100427 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Mo, Binfeng
Yuan, Jiali
Lian, Xiaoming
Cai, Xingxing
Wang, Qunshan
Li, Yigang
Accuracy of the Injection-Based Occlusion Tool Utilizing Saline and Glucose Solution in Cryoballoon Ablation Guided by a Novel Dielectric Imaging System
title Accuracy of the Injection-Based Occlusion Tool Utilizing Saline and Glucose Solution in Cryoballoon Ablation Guided by a Novel Dielectric Imaging System
title_full Accuracy of the Injection-Based Occlusion Tool Utilizing Saline and Glucose Solution in Cryoballoon Ablation Guided by a Novel Dielectric Imaging System
title_fullStr Accuracy of the Injection-Based Occlusion Tool Utilizing Saline and Glucose Solution in Cryoballoon Ablation Guided by a Novel Dielectric Imaging System
title_full_unstemmed Accuracy of the Injection-Based Occlusion Tool Utilizing Saline and Glucose Solution in Cryoballoon Ablation Guided by a Novel Dielectric Imaging System
title_short Accuracy of the Injection-Based Occlusion Tool Utilizing Saline and Glucose Solution in Cryoballoon Ablation Guided by a Novel Dielectric Imaging System
title_sort accuracy of the injection-based occlusion tool utilizing saline and glucose solution in cryoballoon ablation guided by a novel dielectric imaging system
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10607856/
https://www.ncbi.nlm.nih.gov/pubmed/37887874
http://dx.doi.org/10.3390/jcdd10100427
work_keys_str_mv AT mobinfeng accuracyoftheinjectionbasedocclusiontoolutilizingsalineandglucosesolutionincryoballoonablationguidedbyanoveldielectricimagingsystem
AT yuanjiali accuracyoftheinjectionbasedocclusiontoolutilizingsalineandglucosesolutionincryoballoonablationguidedbyanoveldielectricimagingsystem
AT lianxiaoming accuracyoftheinjectionbasedocclusiontoolutilizingsalineandglucosesolutionincryoballoonablationguidedbyanoveldielectricimagingsystem
AT caixingxing accuracyoftheinjectionbasedocclusiontoolutilizingsalineandglucosesolutionincryoballoonablationguidedbyanoveldielectricimagingsystem
AT wangqunshan accuracyoftheinjectionbasedocclusiontoolutilizingsalineandglucosesolutionincryoballoonablationguidedbyanoveldielectricimagingsystem
AT liyigang accuracyoftheinjectionbasedocclusiontoolutilizingsalineandglucosesolutionincryoballoonablationguidedbyanoveldielectricimagingsystem