Use of CdZnTe Quasi-Hemispherical Detectors under the Impact of Low Temperatures and High-Gamma Radiation Fluxes
This article discusses the possibilities and limitations of CdZnTe (CZT) quasi-hemispherical detectors operating over a wide temperature range and at high-gamma radiation fluxes. The results of the negative influence of low- and high-operating temperatures, as well as high-radiation fluxes on spectr...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10611207/ https://www.ncbi.nlm.nih.gov/pubmed/37896472 http://dx.doi.org/10.3390/s23208378 |
Sumario: | This article discusses the possibilities and limitations of CdZnTe (CZT) quasi-hemispherical detectors operating over a wide temperature range and at high-gamma radiation fluxes. The results of the negative influence of low- and high-operating temperatures, as well as high-radiation fluxes on spectrometric characteristics of CZT detectors and possible ways to eliminate performance deterioration, are presented. The impact of infrared (IR) illumination parameters, such as wavelength and irradiation intensity on the spectroscopy performance of detectors, was investigated. A correctly chosen IR illumination wavelength and intensity were shown to significantly improve the energy resolution of CZT quasi-hemispherical detectors, allowing their stable operation in high-gamma radiation fluxes and extend the operating temperature range toward low temperatures. The influences of bias voltage values and temperatures on the quasi-hemispherical CZT detectors’ operating ability at high-gamma radiation fluxes were studied. |
---|