Comparison of image quality, arterial depiction, and radiation dose between two rapid kVp-switching dual-energy CT scanners in CT angiography at 40-keV

PURPOSE: To compare the quantitative parameters and qualitative image quality of dual-energy CT angiography (CTA) between two rapid kVp-switching dual-energy CT scanners. MATERIALS AND METHODS: Between May 2021 and March 2022, 79 participants underwent whole-body CTA using either Discovery CT750 HD...

Descripción completa

Detalles Bibliográficos
Autores principales: Kaga, Tetsuro, Noda, Yoshifumi, Nagata, Shoma, Kawai, Nobuyuki, Miyoshi, Toshiharu, Hyodo, Fuminori, Kato, Hiroki, Matsuo, Masayuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Nature Singapore 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10613589/
https://www.ncbi.nlm.nih.gov/pubmed/37212946
http://dx.doi.org/10.1007/s11604-023-01448-5
_version_ 1785128861330046976
author Kaga, Tetsuro
Noda, Yoshifumi
Nagata, Shoma
Kawai, Nobuyuki
Miyoshi, Toshiharu
Hyodo, Fuminori
Kato, Hiroki
Matsuo, Masayuki
author_facet Kaga, Tetsuro
Noda, Yoshifumi
Nagata, Shoma
Kawai, Nobuyuki
Miyoshi, Toshiharu
Hyodo, Fuminori
Kato, Hiroki
Matsuo, Masayuki
author_sort Kaga, Tetsuro
collection PubMed
description PURPOSE: To compare the quantitative parameters and qualitative image quality of dual-energy CT angiography (CTA) between two rapid kVp-switching dual-energy CT scanners. MATERIALS AND METHODS: Between May 2021 and March 2022, 79 participants underwent whole-body CTA using either Discovery CT750 HD (Group A, n = 38) or Revolution CT Apex (Group B, n = 41). All data were reconstructed at 40-keV and with adaptive statistical iterative reconstruction-Veo of 40%. The two groups were compared in terms of CT numbers of the thoracic and abdominal aorta, and the iliac artery, background noise, signal-to-noise ratio (SNR) of the artery, CT dose-index volume (CTDI(vol)), and qualitative scores for image noise, sharpness, diagnostic acceptability, and arterial depictions. RESULTS: The median CT number of the abdominal aorta (p = 0.04) and SNR of the thoracic aorta (p = 0.02) were higher in Group B than in Group A, while no difference was observed in the other CT numbers and SNRs of the artery (p = 0.09–0.23). The background noises at the thoracic (p = 0.11), abdominal (p = 0.85), and pelvic (p = 0.85) regions were comparable between the two groups. CTDI(vol) was lower in Group B than in Group A (p = 0.006). All qualitative scores were higher in Group B than in Group A (p < 0.001–0.04). The arterial depictions were nearly identical in both two groups (p = 0.005–1.0). CONCLUSION: In dual-energy CTA at 40-keV, Revolution CT Apex improved qualitative image quality and reduced radiation dose.
format Online
Article
Text
id pubmed-10613589
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Springer Nature Singapore
record_format MEDLINE/PubMed
spelling pubmed-106135892023-10-31 Comparison of image quality, arterial depiction, and radiation dose between two rapid kVp-switching dual-energy CT scanners in CT angiography at 40-keV Kaga, Tetsuro Noda, Yoshifumi Nagata, Shoma Kawai, Nobuyuki Miyoshi, Toshiharu Hyodo, Fuminori Kato, Hiroki Matsuo, Masayuki Jpn J Radiol Original Article PURPOSE: To compare the quantitative parameters and qualitative image quality of dual-energy CT angiography (CTA) between two rapid kVp-switching dual-energy CT scanners. MATERIALS AND METHODS: Between May 2021 and March 2022, 79 participants underwent whole-body CTA using either Discovery CT750 HD (Group A, n = 38) or Revolution CT Apex (Group B, n = 41). All data were reconstructed at 40-keV and with adaptive statistical iterative reconstruction-Veo of 40%. The two groups were compared in terms of CT numbers of the thoracic and abdominal aorta, and the iliac artery, background noise, signal-to-noise ratio (SNR) of the artery, CT dose-index volume (CTDI(vol)), and qualitative scores for image noise, sharpness, diagnostic acceptability, and arterial depictions. RESULTS: The median CT number of the abdominal aorta (p = 0.04) and SNR of the thoracic aorta (p = 0.02) were higher in Group B than in Group A, while no difference was observed in the other CT numbers and SNRs of the artery (p = 0.09–0.23). The background noises at the thoracic (p = 0.11), abdominal (p = 0.85), and pelvic (p = 0.85) regions were comparable between the two groups. CTDI(vol) was lower in Group B than in Group A (p = 0.006). All qualitative scores were higher in Group B than in Group A (p < 0.001–0.04). The arterial depictions were nearly identical in both two groups (p = 0.005–1.0). CONCLUSION: In dual-energy CTA at 40-keV, Revolution CT Apex improved qualitative image quality and reduced radiation dose. Springer Nature Singapore 2023-05-22 2023 /pmc/articles/PMC10613589/ /pubmed/37212946 http://dx.doi.org/10.1007/s11604-023-01448-5 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article
Kaga, Tetsuro
Noda, Yoshifumi
Nagata, Shoma
Kawai, Nobuyuki
Miyoshi, Toshiharu
Hyodo, Fuminori
Kato, Hiroki
Matsuo, Masayuki
Comparison of image quality, arterial depiction, and radiation dose between two rapid kVp-switching dual-energy CT scanners in CT angiography at 40-keV
title Comparison of image quality, arterial depiction, and radiation dose between two rapid kVp-switching dual-energy CT scanners in CT angiography at 40-keV
title_full Comparison of image quality, arterial depiction, and radiation dose between two rapid kVp-switching dual-energy CT scanners in CT angiography at 40-keV
title_fullStr Comparison of image quality, arterial depiction, and radiation dose between two rapid kVp-switching dual-energy CT scanners in CT angiography at 40-keV
title_full_unstemmed Comparison of image quality, arterial depiction, and radiation dose between two rapid kVp-switching dual-energy CT scanners in CT angiography at 40-keV
title_short Comparison of image quality, arterial depiction, and radiation dose between two rapid kVp-switching dual-energy CT scanners in CT angiography at 40-keV
title_sort comparison of image quality, arterial depiction, and radiation dose between two rapid kvp-switching dual-energy ct scanners in ct angiography at 40-kev
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10613589/
https://www.ncbi.nlm.nih.gov/pubmed/37212946
http://dx.doi.org/10.1007/s11604-023-01448-5
work_keys_str_mv AT kagatetsuro comparisonofimagequalityarterialdepictionandradiationdosebetweentworapidkvpswitchingdualenergyctscannersinctangiographyat40kev
AT nodayoshifumi comparisonofimagequalityarterialdepictionandradiationdosebetweentworapidkvpswitchingdualenergyctscannersinctangiographyat40kev
AT nagatashoma comparisonofimagequalityarterialdepictionandradiationdosebetweentworapidkvpswitchingdualenergyctscannersinctangiographyat40kev
AT kawainobuyuki comparisonofimagequalityarterialdepictionandradiationdosebetweentworapidkvpswitchingdualenergyctscannersinctangiographyat40kev
AT miyoshitoshiharu comparisonofimagequalityarterialdepictionandradiationdosebetweentworapidkvpswitchingdualenergyctscannersinctangiographyat40kev
AT hyodofuminori comparisonofimagequalityarterialdepictionandradiationdosebetweentworapidkvpswitchingdualenergyctscannersinctangiographyat40kev
AT katohiroki comparisonofimagequalityarterialdepictionandradiationdosebetweentworapidkvpswitchingdualenergyctscannersinctangiographyat40kev
AT matsuomasayuki comparisonofimagequalityarterialdepictionandradiationdosebetweentworapidkvpswitchingdualenergyctscannersinctangiographyat40kev