Finite element analysis comparison of Type 42A2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame

BACKGROUND: At present, not all Type AO/OTA 42A2 open fractures can be treated by external fixation brackets, not to mention the inconvenience of this technique in clinical practice. External titanium alloy locking plates, which are lightweight and easy-to-operate, can be used as an alternative trea...

Descripción completa

Detalles Bibliográficos
Autores principales: Fang, Shitao, Zhang, Ling, Yang, Yunqi, Wang, Yun, Guo, Jinkun, Mi, Lei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10619248/
https://www.ncbi.nlm.nih.gov/pubmed/37907959
http://dx.doi.org/10.1186/s13018-023-04307-1
_version_ 1785129946151124992
author Fang, Shitao
Zhang, Ling
Yang, Yunqi
Wang, Yun
Guo, Jinkun
Mi, Lei
author_facet Fang, Shitao
Zhang, Ling
Yang, Yunqi
Wang, Yun
Guo, Jinkun
Mi, Lei
author_sort Fang, Shitao
collection PubMed
description BACKGROUND: At present, not all Type AO/OTA 42A2 open fractures can be treated by external fixation brackets, not to mention the inconvenience of this technique in clinical practice. External titanium alloy locking plates, which are lightweight and easy-to-operate, can be used as an alternative treatment option for such patients. However, there are few reports of finite element biomechanical analysis on the titanium alloy locking plates and fixation brackets being placed on the medial side of the tibial fracture. In this study, the biomechanical properties of titanium alloy locking plates and fixation brackets for treating Type AO/OTA 42A2 fractures were compared by applying the finite element method, and the results provided data support for the clinical application of the external titanium alloy locking plate technique. METHODS: Type AO/OTA 42A2 fracture models were constructed using CT data of a male volunteer for two external fixation techniques, namely the external titanium alloy locking plate technique and the external fixation bracket technique, according to commonly-used clinical protocols. Then, the four-point bending, axial compression, clockwise rotation and counterclockwise rotation tests under the maximum load were simulated in finite element analysis software. The stress distribution, peak stress and overall tibial displacement data for the two different external fixation techniques were obtained and compared. RESULTS: In the four different test conditions (i.e., four-point bending, axial compression, clockwise torsion, counterclockwise torsion) under the maximum load, the two external fixation techniques showed obvious von Mises stress concentration at the contacts between the screw and tibia, between the screw and titanium alloy locking plate, between the self-tapping self-drilling needle and tibia, between the self-tapping self-drilling needle and the external fixation device, as well as around the fracture end and around the cortical bone at the upper and lower ends of the tibia. The peak stress was ranged 26.67–558.77 MPa, all below the yield stress strength of titanium alloy. The peak tibial displacement of the external titanium alloy locking plate model was smaller than that of the fixation bracket model. In terms of structural stability, the external titanium alloy locking plate technique was superior to the external fixation bracket technique. CONCLUSIONS: When fixing Type AO/OTA 42A2 fractures, external titanium alloy locking plates are not only lightweight and easy-to-operate, but also have better performance in terms of axial compression, bending and torsion resistance. According to the finite element biomechanical analysis, external titanium alloy locking plates are superior to traditional external fixation brackets in treating Type AO/OTA 42A2 fractures and can better meet the needs of clinical application.
format Online
Article
Text
id pubmed-10619248
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-106192482023-11-02 Finite element analysis comparison of Type 42A2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame Fang, Shitao Zhang, Ling Yang, Yunqi Wang, Yun Guo, Jinkun Mi, Lei J Orthop Surg Res Research Article BACKGROUND: At present, not all Type AO/OTA 42A2 open fractures can be treated by external fixation brackets, not to mention the inconvenience of this technique in clinical practice. External titanium alloy locking plates, which are lightweight and easy-to-operate, can be used as an alternative treatment option for such patients. However, there are few reports of finite element biomechanical analysis on the titanium alloy locking plates and fixation brackets being placed on the medial side of the tibial fracture. In this study, the biomechanical properties of titanium alloy locking plates and fixation brackets for treating Type AO/OTA 42A2 fractures were compared by applying the finite element method, and the results provided data support for the clinical application of the external titanium alloy locking plate technique. METHODS: Type AO/OTA 42A2 fracture models were constructed using CT data of a male volunteer for two external fixation techniques, namely the external titanium alloy locking plate technique and the external fixation bracket technique, according to commonly-used clinical protocols. Then, the four-point bending, axial compression, clockwise rotation and counterclockwise rotation tests under the maximum load were simulated in finite element analysis software. The stress distribution, peak stress and overall tibial displacement data for the two different external fixation techniques were obtained and compared. RESULTS: In the four different test conditions (i.e., four-point bending, axial compression, clockwise torsion, counterclockwise torsion) under the maximum load, the two external fixation techniques showed obvious von Mises stress concentration at the contacts between the screw and tibia, between the screw and titanium alloy locking plate, between the self-tapping self-drilling needle and tibia, between the self-tapping self-drilling needle and the external fixation device, as well as around the fracture end and around the cortical bone at the upper and lower ends of the tibia. The peak stress was ranged 26.67–558.77 MPa, all below the yield stress strength of titanium alloy. The peak tibial displacement of the external titanium alloy locking plate model was smaller than that of the fixation bracket model. In terms of structural stability, the external titanium alloy locking plate technique was superior to the external fixation bracket technique. CONCLUSIONS: When fixing Type AO/OTA 42A2 fractures, external titanium alloy locking plates are not only lightweight and easy-to-operate, but also have better performance in terms of axial compression, bending and torsion resistance. According to the finite element biomechanical analysis, external titanium alloy locking plates are superior to traditional external fixation brackets in treating Type AO/OTA 42A2 fractures and can better meet the needs of clinical application. BioMed Central 2023-10-31 /pmc/articles/PMC10619248/ /pubmed/37907959 http://dx.doi.org/10.1186/s13018-023-04307-1 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research Article
Fang, Shitao
Zhang, Ling
Yang, Yunqi
Wang, Yun
Guo, Jinkun
Mi, Lei
Finite element analysis comparison of Type 42A2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame
title Finite element analysis comparison of Type 42A2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame
title_full Finite element analysis comparison of Type 42A2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame
title_fullStr Finite element analysis comparison of Type 42A2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame
title_full_unstemmed Finite element analysis comparison of Type 42A2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame
title_short Finite element analysis comparison of Type 42A2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame
title_sort finite element analysis comparison of type 42a2 fracture fixed with external titanium alloy locking plate and traditional external fixation frame
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10619248/
https://www.ncbi.nlm.nih.gov/pubmed/37907959
http://dx.doi.org/10.1186/s13018-023-04307-1
work_keys_str_mv AT fangshitao finiteelementanalysiscomparisonoftype42a2fracturefixedwithexternaltitaniumalloylockingplateandtraditionalexternalfixationframe
AT zhangling finiteelementanalysiscomparisonoftype42a2fracturefixedwithexternaltitaniumalloylockingplateandtraditionalexternalfixationframe
AT yangyunqi finiteelementanalysiscomparisonoftype42a2fracturefixedwithexternaltitaniumalloylockingplateandtraditionalexternalfixationframe
AT wangyun finiteelementanalysiscomparisonoftype42a2fracturefixedwithexternaltitaniumalloylockingplateandtraditionalexternalfixationframe
AT guojinkun finiteelementanalysiscomparisonoftype42a2fracturefixedwithexternaltitaniumalloylockingplateandtraditionalexternalfixationframe
AT milei finiteelementanalysiscomparisonoftype42a2fracturefixedwithexternaltitaniumalloylockingplateandtraditionalexternalfixationframe