Magnetic resonance imaging burn injury from metal eyelets on patient’s clothing
This case report describes an incident which occurred following a cardiac MRI scan. The patient was permitted to wear their own jogging bottoms for the scan, which had two metal eyelets on them. The day after the MRI examination the patient called the MRI department to inform them that there was a l...
Autores principales: | Prescott, Sarah, Jarvest, Bruce, Poole, Harry |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The British Institute of Radiology.
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10621571/ https://www.ncbi.nlm.nih.gov/pubmed/37928698 http://dx.doi.org/10.1259/bjrcr.20230088 |