Cargando…
The effect of combining different sampling tools on the performance of electromagnetic navigational bronchoscopy for the evaluation of peripheral lung lesions and factors associated with its diagnostic yield
BACKGROUND: We assessed the performance of Electromagnetic navigational bronchoscopy (ENB) as a standalone diagnostic technique and the performance of different sampling tools used during the procedure. METHODS: We recruited 160 consecutive patients who underwent ENB for peripheral lung lesions (PLL...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10634141/ https://www.ncbi.nlm.nih.gov/pubmed/37940942 http://dx.doi.org/10.1186/s12890-023-02711-1 |
Sumario: | BACKGROUND: We assessed the performance of Electromagnetic navigational bronchoscopy (ENB) as a standalone diagnostic technique and the performance of different sampling tools used during the procedure. METHODS: We recruited 160 consecutive patients who underwent ENB for peripheral lung lesions (PLL) at a tertiary care centre. The diagnostic performance of ENB and sampling tools was assessed using a logistic regression model and a ROC-curve in which the dependent variable was diagnostic success. A multivariate model was built to predict diagnostic success before performing ENB to select the best candidates for the procedure. RESULTS: Most patients with PLLs in the study were male (65%), with a mean age of 67.9 years. The yield was 66% when the most common techniques were used together as suction catheter + transbronchial biopsy forceps (TBBx) + bronchoalveolar lavage + bronchial washing (p < 0.001) and increased to 69% when transbronchial needle aspiration (TBNA) and cytology brush were added (p < 0.001). Adding diagnostic techniques such as TBBx and TBNA resulted in an increase in diagnostic performance, with a statistically significant trend (p = 0.002). The logistic model area-under the ROC-curve for diagnostic success during ENB was 0.83 (95% CI:0.75–0.90; p < 0.001), and a logit value ≥ 0.12 was associated with ≥ 50% probability of diagnostic success. CONCLUSIONS: ENB, as a stand-alone diagnostic tool for the evaluation of PLLs when performed by experienced operators using a multi-modality technique, has a good diagnostic yield. The probability of having a diagnostic ENB could be assessed using the proposed model. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12890-023-02711-1. |
---|