Cargando…

Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru

Managing contaminated waste in blood banks is a crucial process that must end with the safe disposal of blood products. In this sense, practical methods and indicators must be used to guarantee the operability of autoclave equipment and the sterility of treated waste to mitigate the risk of biologic...

Descripción completa

Detalles Bibliográficos
Autores principales: Lagos-Palomino, Lesdy, Rueda-Torres, Lenin, Sanchez-Holguin, Gloria, Soncco-Llulluy, Fernando, Rosales-Rimache, Jaime
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10637551/
https://www.ncbi.nlm.nih.gov/pubmed/37960832
http://dx.doi.org/10.1097/MD.0000000000035293
_version_ 1785146521596985344
author Lagos-Palomino, Lesdy
Rueda-Torres, Lenin
Sanchez-Holguin, Gloria
Soncco-Llulluy, Fernando
Rosales-Rimache, Jaime
author_facet Lagos-Palomino, Lesdy
Rueda-Torres, Lenin
Sanchez-Holguin, Gloria
Soncco-Llulluy, Fernando
Rosales-Rimache, Jaime
author_sort Lagos-Palomino, Lesdy
collection PubMed
description Managing contaminated waste in blood banks is a crucial process that must end with the safe disposal of blood products. In this sense, practical methods and indicators must be used to guarantee the operability of autoclave equipment and the sterility of treated waste to mitigate the risk of biological exposure and hospital biosafety. We designed a cross-sectional study to evaluate the performance the autoclaving process with Bowie & Dick test and biological indicator at the Hemotherapy and Blood Bank Unit of the Cayetano Heredia Hospital in Peru. Fifty autoclaving processes were carried out independently for the Bowie & Dick tests and biological indicators based on Geobacillus stearothermophilus spores. Autoclaving was programmed at 134°C for 3.5 minutes for the Bowie & Dick test, while, for the biological indicator, it was programmed at 121°C for 5 minutes. Both in the presence of contaminated waste. The autoclaving process evaluated by the Bowie & Dick test had a compliance rate of 80% (CI95: 66.3%–90.0%), while, by the biological indicator, 90% (CI95: 78.2%–96.7%). We did not find significant differences between the results of both tests (P = .689; Fisher exact test). The compliance rate in the autoclaving process within the blood bank of a Peruvian Hospital is acceptable; however, there are actions for continuous improvement, especially at the vacuum level in the autoclaving process.
format Online
Article
Text
id pubmed-10637551
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-106375512023-11-15 Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru Lagos-Palomino, Lesdy Rueda-Torres, Lenin Sanchez-Holguin, Gloria Soncco-Llulluy, Fernando Rosales-Rimache, Jaime Medicine (Baltimore) 6600 Managing contaminated waste in blood banks is a crucial process that must end with the safe disposal of blood products. In this sense, practical methods and indicators must be used to guarantee the operability of autoclave equipment and the sterility of treated waste to mitigate the risk of biological exposure and hospital biosafety. We designed a cross-sectional study to evaluate the performance the autoclaving process with Bowie & Dick test and biological indicator at the Hemotherapy and Blood Bank Unit of the Cayetano Heredia Hospital in Peru. Fifty autoclaving processes were carried out independently for the Bowie & Dick tests and biological indicators based on Geobacillus stearothermophilus spores. Autoclaving was programmed at 134°C for 3.5 minutes for the Bowie & Dick test, while, for the biological indicator, it was programmed at 121°C for 5 minutes. Both in the presence of contaminated waste. The autoclaving process evaluated by the Bowie & Dick test had a compliance rate of 80% (CI95: 66.3%–90.0%), while, by the biological indicator, 90% (CI95: 78.2%–96.7%). We did not find significant differences between the results of both tests (P = .689; Fisher exact test). The compliance rate in the autoclaving process within the blood bank of a Peruvian Hospital is acceptable; however, there are actions for continuous improvement, especially at the vacuum level in the autoclaving process. Lippincott Williams & Wilkins 2023-11-10 /pmc/articles/PMC10637551/ /pubmed/37960832 http://dx.doi.org/10.1097/MD.0000000000035293 Text en Copyright © 2023 the Author(s). Published by Wolters Kluwer Health, Inc. https://creativecommons.org/licenses/by-nc/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial License 4.0 (CCBY-NC) (https://creativecommons.org/licenses/by-nc/4.0/) , where it is permissible to download, share, remix, transform, and buildup the work provided it is properly cited. The work cannot be used commercially without permission from the journal.
spellingShingle 6600
Lagos-Palomino, Lesdy
Rueda-Torres, Lenin
Sanchez-Holguin, Gloria
Soncco-Llulluy, Fernando
Rosales-Rimache, Jaime
Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru
title Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru
title_full Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru
title_fullStr Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru
title_full_unstemmed Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru
title_short Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru
title_sort performance evaluation of the sterilization process with bowie & dick test and biological indicator in the quality control of a blood bank in peru
topic 6600
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10637551/
https://www.ncbi.nlm.nih.gov/pubmed/37960832
http://dx.doi.org/10.1097/MD.0000000000035293
work_keys_str_mv AT lagospalominolesdy performanceevaluationofthesterilizationprocesswithbowiedicktestandbiologicalindicatorinthequalitycontrolofabloodbankinperu
AT ruedatorreslenin performanceevaluationofthesterilizationprocesswithbowiedicktestandbiologicalindicatorinthequalitycontrolofabloodbankinperu
AT sanchezholguingloria performanceevaluationofthesterilizationprocesswithbowiedicktestandbiologicalindicatorinthequalitycontrolofabloodbankinperu
AT sonccollulluyfernando performanceevaluationofthesterilizationprocesswithbowiedicktestandbiologicalindicatorinthequalitycontrolofabloodbankinperu
AT rosalesrimachejaime performanceevaluationofthesterilizationprocesswithbowiedicktestandbiologicalindicatorinthequalitycontrolofabloodbankinperu