The diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population
PURPOSE: To compare the diagnostic accuracy of the photopic negative response (PhNR) elicited by red-blue (RB) and white-white (WW) stimuli, for detection of retinal ganglion cell (RGC) dysfunction in a heterogeneous clinical cohort. METHODS: Adults referred for electrophysiological investigations w...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10638186/ https://www.ncbi.nlm.nih.gov/pubmed/37889400 http://dx.doi.org/10.1007/s10633-023-09956-5 |
_version_ | 1785146566447726592 |
---|---|
author | Leo, Shaun M. Neveu, Magella M. Yu-Wai-Man, Patrick Mahroo, Omar A. Robson, Anthony G. |
author_facet | Leo, Shaun M. Neveu, Magella M. Yu-Wai-Man, Patrick Mahroo, Omar A. Robson, Anthony G. |
author_sort | Leo, Shaun M. |
collection | PubMed |
description | PURPOSE: To compare the diagnostic accuracy of the photopic negative response (PhNR) elicited by red-blue (RB) and white-white (WW) stimuli, for detection of retinal ganglion cell (RGC) dysfunction in a heterogeneous clinical cohort. METHODS: Adults referred for electrophysiological investigations were recruited consecutively for this single-centre, prospective, paired diagnostic accuracy study. PhNRs were recorded to red flashes (1.5 cd·s·m(−2)) on a blue background (10 cd·m(−2)) and to white flashes on a white background (the latter being the ISCEV standard LA 3 stimulus). PhNR results were compared with a reference test battery assessing RGC/optic nerve structure and function including optical coherence tomography (OCT) retinal nerve fibre layer thickness and mean RGC volume measurements, fundus photography, pattern electroretinography and visual evoked potentials. Primary outcome measures were differences in sensitivity and specificity of the two PhNR methods. RESULTS: Two hundred and forty-three participants were initially enrolled, with 200 (median age 54; range 18–95; female 65%) meeting inclusion criteria. Sensitivity was 53% (95% confidence intervals [CI] 39% to 68%) and 62% (95% CI 48% to 76%), for WW and RB PhNRs, respectively. Specificity was 80% (95% CI 74% to 86%) and 78% (95% CI 72% to 85%), respectively. There was a statistically significant difference between sensitivities (p = 0.046) but not specificities (p = 0.08) of the two methods. Receiver operator characteristic (ROC) area under the curve (AUC) values were 0.73 for WW and 0.74 for RB PhNRs. CONCLUSION: PhNRs to red flashes on a blue background may be more sensitive than white-on-white stimuli, but there is no significant difference between specificities. This study highlights the value and potential convenience of using white-on-white stimuli, already used widely for routine ERG assessment. |
format | Online Article Text |
id | pubmed-10638186 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-106381862023-11-14 The diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population Leo, Shaun M. Neveu, Magella M. Yu-Wai-Man, Patrick Mahroo, Omar A. Robson, Anthony G. Doc Ophthalmol Original Research Article PURPOSE: To compare the diagnostic accuracy of the photopic negative response (PhNR) elicited by red-blue (RB) and white-white (WW) stimuli, for detection of retinal ganglion cell (RGC) dysfunction in a heterogeneous clinical cohort. METHODS: Adults referred for electrophysiological investigations were recruited consecutively for this single-centre, prospective, paired diagnostic accuracy study. PhNRs were recorded to red flashes (1.5 cd·s·m(−2)) on a blue background (10 cd·m(−2)) and to white flashes on a white background (the latter being the ISCEV standard LA 3 stimulus). PhNR results were compared with a reference test battery assessing RGC/optic nerve structure and function including optical coherence tomography (OCT) retinal nerve fibre layer thickness and mean RGC volume measurements, fundus photography, pattern electroretinography and visual evoked potentials. Primary outcome measures were differences in sensitivity and specificity of the two PhNR methods. RESULTS: Two hundred and forty-three participants were initially enrolled, with 200 (median age 54; range 18–95; female 65%) meeting inclusion criteria. Sensitivity was 53% (95% confidence intervals [CI] 39% to 68%) and 62% (95% CI 48% to 76%), for WW and RB PhNRs, respectively. Specificity was 80% (95% CI 74% to 86%) and 78% (95% CI 72% to 85%), respectively. There was a statistically significant difference between sensitivities (p = 0.046) but not specificities (p = 0.08) of the two methods. Receiver operator characteristic (ROC) area under the curve (AUC) values were 0.73 for WW and 0.74 for RB PhNRs. CONCLUSION: PhNRs to red flashes on a blue background may be more sensitive than white-on-white stimuli, but there is no significant difference between specificities. This study highlights the value and potential convenience of using white-on-white stimuli, already used widely for routine ERG assessment. Springer Berlin Heidelberg 2023-10-27 2023 /pmc/articles/PMC10638186/ /pubmed/37889400 http://dx.doi.org/10.1007/s10633-023-09956-5 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Original Research Article Leo, Shaun M. Neveu, Magella M. Yu-Wai-Man, Patrick Mahroo, Omar A. Robson, Anthony G. The diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population |
title | The diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population |
title_full | The diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population |
title_fullStr | The diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population |
title_full_unstemmed | The diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population |
title_short | The diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population |
title_sort | diagnostic accuracy of photopic negative responses evoked by broadband and chromatic stimuli in a clinically heterogeneous population |
topic | Original Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10638186/ https://www.ncbi.nlm.nih.gov/pubmed/37889400 http://dx.doi.org/10.1007/s10633-023-09956-5 |
work_keys_str_mv | AT leoshaunm thediagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT neveumagellam thediagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT yuwaimanpatrick thediagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT mahrooomara thediagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT robsonanthonyg thediagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT leoshaunm diagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT neveumagellam diagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT yuwaimanpatrick diagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT mahrooomara diagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation AT robsonanthonyg diagnosticaccuracyofphotopicnegativeresponsesevokedbybroadbandandchromaticstimuliinaclinicallyheterogeneouspopulation |