Implications of humeral short-stem diametral sizing on implant stability

BACKGROUND: Shoulder arthroplasty humeral stem design has evolved to include various shapes, coatings, lengths, sizes, and fixation methods. While necessary to accommodate patient anatomy characteristics, this creates a surgical paradox of choice. The relationship between the surgeon’s selection of...

Descripción completa

Detalles Bibliográficos
Autores principales: Reeves, Jacob M., Spangenberg, Gregory W., Elwell, Josie A., Stewart, Ben, Vanasse, Tom, Roche, Chris, Faber, Kenneth J., Langohr, G. Daniel G.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10638575/
https://www.ncbi.nlm.nih.gov/pubmed/37969503
http://dx.doi.org/10.1016/j.jseint.2023.06.023
_version_ 1785146590504157184
author Reeves, Jacob M.
Spangenberg, Gregory W.
Elwell, Josie A.
Stewart, Ben
Vanasse, Tom
Roche, Chris
Faber, Kenneth J.
Langohr, G. Daniel G.
author_facet Reeves, Jacob M.
Spangenberg, Gregory W.
Elwell, Josie A.
Stewart, Ben
Vanasse, Tom
Roche, Chris
Faber, Kenneth J.
Langohr, G. Daniel G.
author_sort Reeves, Jacob M.
collection PubMed
description BACKGROUND: Shoulder arthroplasty humeral stem design has evolved to include various shapes, coatings, lengths, sizes, and fixation methods. While necessary to accommodate patient anatomy characteristics, this creates a surgical paradox of choice. The relationship between the surgeon’s selection of short-stem implant size and construct stiffness, resistance to subsidence and micromotion has not been assessed. METHODS: Eight paired cadaveric humeri were reconstructed with surgeon-selected (SS) and 2-mm diametrically larger (SS+2) short-stemmed press-fit implants. Each reconstruction was subjected to 2000 cycles of 90° forward flexion loading, and stem subsidence and micromotion were measured using optical tracking. Compressive stiffness of the stem-bone reconstruction was then assessed by applying a load in-line with the stem axis that resulted in 5 mm of stem subsidence. RESULTS: Increasing stem size by 2 mm resulted in the construct stiffness more than doubling compared to SS stems (−741 ± 243 N/mm vs. −334 ± 120 N/mm; P = .003; power = 0.971). These larger stems also subsided significantly less than their SS counterparts (SS: 1.2 ± 0.6 mm; SS+2: 0.5 ± 0.5 mm; P = .029; power = 0.66), though there were no significant changes in micromotion (SS: 169 ± 59 μm; SS+2: 187 ± 52 μm; P = .506; power = 0.094). CONCLUSIONS: The results of this study highlight the importance of proper short-stem sizing, as a relatively small 2 mm increase in diametral size was observed to significantly impact construct stiffness, which could increase the risk of stress shielding and implant loosening. Future work should focus on developing tools that objectively quantify bone quality and aid surgeons in selecting the appropriate size short-stem humeral implants for a particular patient.
format Online
Article
Text
id pubmed-10638575
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-106385752023-11-15 Implications of humeral short-stem diametral sizing on implant stability Reeves, Jacob M. Spangenberg, Gregory W. Elwell, Josie A. Stewart, Ben Vanasse, Tom Roche, Chris Faber, Kenneth J. Langohr, G. Daniel G. JSES Int Shoulder BACKGROUND: Shoulder arthroplasty humeral stem design has evolved to include various shapes, coatings, lengths, sizes, and fixation methods. While necessary to accommodate patient anatomy characteristics, this creates a surgical paradox of choice. The relationship between the surgeon’s selection of short-stem implant size and construct stiffness, resistance to subsidence and micromotion has not been assessed. METHODS: Eight paired cadaveric humeri were reconstructed with surgeon-selected (SS) and 2-mm diametrically larger (SS+2) short-stemmed press-fit implants. Each reconstruction was subjected to 2000 cycles of 90° forward flexion loading, and stem subsidence and micromotion were measured using optical tracking. Compressive stiffness of the stem-bone reconstruction was then assessed by applying a load in-line with the stem axis that resulted in 5 mm of stem subsidence. RESULTS: Increasing stem size by 2 mm resulted in the construct stiffness more than doubling compared to SS stems (−741 ± 243 N/mm vs. −334 ± 120 N/mm; P = .003; power = 0.971). These larger stems also subsided significantly less than their SS counterparts (SS: 1.2 ± 0.6 mm; SS+2: 0.5 ± 0.5 mm; P = .029; power = 0.66), though there were no significant changes in micromotion (SS: 169 ± 59 μm; SS+2: 187 ± 52 μm; P = .506; power = 0.094). CONCLUSIONS: The results of this study highlight the importance of proper short-stem sizing, as a relatively small 2 mm increase in diametral size was observed to significantly impact construct stiffness, which could increase the risk of stress shielding and implant loosening. Future work should focus on developing tools that objectively quantify bone quality and aid surgeons in selecting the appropriate size short-stem humeral implants for a particular patient. Elsevier 2023-07-24 /pmc/articles/PMC10638575/ /pubmed/37969503 http://dx.doi.org/10.1016/j.jseint.2023.06.023 Text en © 2023 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Shoulder
Reeves, Jacob M.
Spangenberg, Gregory W.
Elwell, Josie A.
Stewart, Ben
Vanasse, Tom
Roche, Chris
Faber, Kenneth J.
Langohr, G. Daniel G.
Implications of humeral short-stem diametral sizing on implant stability
title Implications of humeral short-stem diametral sizing on implant stability
title_full Implications of humeral short-stem diametral sizing on implant stability
title_fullStr Implications of humeral short-stem diametral sizing on implant stability
title_full_unstemmed Implications of humeral short-stem diametral sizing on implant stability
title_short Implications of humeral short-stem diametral sizing on implant stability
title_sort implications of humeral short-stem diametral sizing on implant stability
topic Shoulder
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10638575/
https://www.ncbi.nlm.nih.gov/pubmed/37969503
http://dx.doi.org/10.1016/j.jseint.2023.06.023
work_keys_str_mv AT reevesjacobm implicationsofhumeralshortstemdiametralsizingonimplantstability
AT spangenberggregoryw implicationsofhumeralshortstemdiametralsizingonimplantstability
AT elwelljosiea implicationsofhumeralshortstemdiametralsizingonimplantstability
AT stewartben implicationsofhumeralshortstemdiametralsizingonimplantstability
AT vanassetom implicationsofhumeralshortstemdiametralsizingonimplantstability
AT rochechris implicationsofhumeralshortstemdiametralsizingonimplantstability
AT faberkennethj implicationsofhumeralshortstemdiametralsizingonimplantstability
AT langohrgdanielg implicationsofhumeralshortstemdiametralsizingonimplantstability