Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy
PURPOSE: The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energ...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10642597/ https://www.ncbi.nlm.nih.gov/pubmed/37969143 http://dx.doi.org/10.4103/jmp.jmp_27_23 |
_version_ | 1785146997852864512 |
---|---|
author | Chirdchid, Thamonwan Ruangchan, Sirinya Sanghangthum, Taweap |
author_facet | Chirdchid, Thamonwan Ruangchan, Sirinya Sanghangthum, Taweap |
author_sort | Chirdchid, Thamonwan |
collection | PubMed |
description | PURPOSE: The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energy computed tomography (DECT)-based dose calculations. MATERIALS AND METHODS: The CT calibration curve of SECT and DECT data was generated using the stoichiometric calibration method. WET measurement was performed for RSP validation using a Giraffe dosimeter (IBA dosimetry) in various substitute tissues (Gammex) and real pig tissues. The thorax (008A, CIRS) and head (731-HN, CIRS) phantoms were used to generate proton plans. The dosimetric evaluations of SECT and DECT-based plans were performed using the gamma analysis with 1%/1 mm and the dose–volume histograms (DVHs) comparison. RESULTS: For RSP validation of substitute tissues, the largest percent WET difference between measurement and calculation was observed up to 17.9% (4 mm) in lung tissue, using SECT based. In real pig tissues, the average WET difference was 2.3% ± 2.1% and 2.5% ± 2.3% for SECT and DECT, respectively. The average gamma passed of about 92.1% for the lung and 96.8% for the head regions was reported. For the lung region, the DVH of the target dose was observed with a higher predicted dose in SECT than in DECT, while results in the head region were in good agreement for both SECT and DECT. CONCLUSION: The performed dosimetric comparison indicates the dose differences between SECT and DECT. The impact of the CT calibration curve is more pronounced for the thorax region. |
format | Online Article Text |
id | pubmed-10642597 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-106425972023-11-15 Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy Chirdchid, Thamonwan Ruangchan, Sirinya Sanghangthum, Taweap J Med Phys Original Article PURPOSE: The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energy computed tomography (DECT)-based dose calculations. MATERIALS AND METHODS: The CT calibration curve of SECT and DECT data was generated using the stoichiometric calibration method. WET measurement was performed for RSP validation using a Giraffe dosimeter (IBA dosimetry) in various substitute tissues (Gammex) and real pig tissues. The thorax (008A, CIRS) and head (731-HN, CIRS) phantoms were used to generate proton plans. The dosimetric evaluations of SECT and DECT-based plans were performed using the gamma analysis with 1%/1 mm and the dose–volume histograms (DVHs) comparison. RESULTS: For RSP validation of substitute tissues, the largest percent WET difference between measurement and calculation was observed up to 17.9% (4 mm) in lung tissue, using SECT based. In real pig tissues, the average WET difference was 2.3% ± 2.1% and 2.5% ± 2.3% for SECT and DECT, respectively. The average gamma passed of about 92.1% for the lung and 96.8% for the head regions was reported. For the lung region, the DVH of the target dose was observed with a higher predicted dose in SECT than in DECT, while results in the head region were in good agreement for both SECT and DECT. CONCLUSION: The performed dosimetric comparison indicates the dose differences between SECT and DECT. The impact of the CT calibration curve is more pronounced for the thorax region. Wolters Kluwer - Medknow 2023 2023-09-18 /pmc/articles/PMC10642597/ /pubmed/37969143 http://dx.doi.org/10.4103/jmp.jmp_27_23 Text en Copyright: © 2023 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Chirdchid, Thamonwan Ruangchan, Sirinya Sanghangthum, Taweap Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy |
title | Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy |
title_full | Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy |
title_fullStr | Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy |
title_full_unstemmed | Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy |
title_short | Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy |
title_sort | dosimetric comparison between single-energy computed tomography and dual-energy computed tomography relative to stopping power estimation in proton therapy |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10642597/ https://www.ncbi.nlm.nih.gov/pubmed/37969143 http://dx.doi.org/10.4103/jmp.jmp_27_23 |
work_keys_str_mv | AT chirdchidthamonwan dosimetriccomparisonbetweensingleenergycomputedtomographyanddualenergycomputedtomographyrelativetostoppingpowerestimationinprotontherapy AT ruangchansirinya dosimetriccomparisonbetweensingleenergycomputedtomographyanddualenergycomputedtomographyrelativetostoppingpowerestimationinprotontherapy AT sanghangthumtaweap dosimetriccomparisonbetweensingleenergycomputedtomographyanddualenergycomputedtomographyrelativetostoppingpowerestimationinprotontherapy |