Evaluation of International Atomic Energy Agency Technical Report Series-483 Detector-specific Output Correction Factor for Various Collimator Systems

AIM: In this study, a 6MV flattening filter (FF) and 6MV FF Free (FFF) photon beam small-field output factors (OF) were measured with various collimators using different detectors. The corrected OFs were compared with the treatment planning system (TPS) calculated OFs. MATERIALS AND METHODS: OF meas...

Descripción completa

Detalles Bibliográficos
Autores principales: Kannan, Mageshraja, Saminathan, Sathiyan, Chandraraj, Varatharaj, Raj, D. Gowtham, Ganesh, K. M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10642599/
https://www.ncbi.nlm.nih.gov/pubmed/37969152
http://dx.doi.org/10.4103/jmp.jmp_59_23
_version_ 1785146998322626560
author Kannan, Mageshraja
Saminathan, Sathiyan
Chandraraj, Varatharaj
Raj, D. Gowtham
Ganesh, K. M.
author_facet Kannan, Mageshraja
Saminathan, Sathiyan
Chandraraj, Varatharaj
Raj, D. Gowtham
Ganesh, K. M.
author_sort Kannan, Mageshraja
collection PubMed
description AIM: In this study, a 6MV flattening filter (FF) and 6MV FF Free (FFF) photon beam small-field output factors (OF) were measured with various collimators using different detectors. The corrected OFs were compared with the treatment planning system (TPS) calculated OFs. MATERIALS AND METHODS: OF measurements were performed with four different types of collimators: Varian Millennium multi-leaf collimator (MLC), Elekta Agility MLC, Apex micro-MLC (mMLC) and a stereotactic cone. Ten detectors (four ionization chambers and six diodes) were used to perform the OF measurements at a depth of 10 cm with a source-to-surface distance of 90 cm. The corrected OF was calculated from the measurements. The corrected OFs were compared with existing TPS-generated OFs. RESULTS: The use of detector-specific output correction factor (OCF) in the PTW diode P detector reduced the OF uncertainty by <4.1% for 1 cm × 1 cm S(clin). The corrected OF was compared with TPS calculated OF; the maximum variation with the IBA CC01 chamber was 3.75%, 3.72%, 1.16%, and 0.90% for 5 mm stereotactic cone, 0.49 cm × 0.49 cm Apex mMLC, 1 cm × 1 cm Agility MLC, and 1 cm × 1 cm Millennium MLC, respectively. CONCLUSION: The technical report series-483 protocol recommends that detector-specific OCF should be used to calculate the corrected OF from the measured OF. The implementation of OCF in the TPS commissioning will reduce the small-field OF variation by <3% for any type of detector.
format Online
Article
Text
id pubmed-10642599
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-106425992023-11-15 Evaluation of International Atomic Energy Agency Technical Report Series-483 Detector-specific Output Correction Factor for Various Collimator Systems Kannan, Mageshraja Saminathan, Sathiyan Chandraraj, Varatharaj Raj, D. Gowtham Ganesh, K. M. J Med Phys Original Article AIM: In this study, a 6MV flattening filter (FF) and 6MV FF Free (FFF) photon beam small-field output factors (OF) were measured with various collimators using different detectors. The corrected OFs were compared with the treatment planning system (TPS) calculated OFs. MATERIALS AND METHODS: OF measurements were performed with four different types of collimators: Varian Millennium multi-leaf collimator (MLC), Elekta Agility MLC, Apex micro-MLC (mMLC) and a stereotactic cone. Ten detectors (four ionization chambers and six diodes) were used to perform the OF measurements at a depth of 10 cm with a source-to-surface distance of 90 cm. The corrected OF was calculated from the measurements. The corrected OFs were compared with existing TPS-generated OFs. RESULTS: The use of detector-specific output correction factor (OCF) in the PTW diode P detector reduced the OF uncertainty by <4.1% for 1 cm × 1 cm S(clin). The corrected OF was compared with TPS calculated OF; the maximum variation with the IBA CC01 chamber was 3.75%, 3.72%, 1.16%, and 0.90% for 5 mm stereotactic cone, 0.49 cm × 0.49 cm Apex mMLC, 1 cm × 1 cm Agility MLC, and 1 cm × 1 cm Millennium MLC, respectively. CONCLUSION: The technical report series-483 protocol recommends that detector-specific OCF should be used to calculate the corrected OF from the measured OF. The implementation of OCF in the TPS commissioning will reduce the small-field OF variation by <3% for any type of detector. Wolters Kluwer - Medknow 2023 2023-09-18 /pmc/articles/PMC10642599/ /pubmed/37969152 http://dx.doi.org/10.4103/jmp.jmp_59_23 Text en Copyright: © 2023 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Kannan, Mageshraja
Saminathan, Sathiyan
Chandraraj, Varatharaj
Raj, D. Gowtham
Ganesh, K. M.
Evaluation of International Atomic Energy Agency Technical Report Series-483 Detector-specific Output Correction Factor for Various Collimator Systems
title Evaluation of International Atomic Energy Agency Technical Report Series-483 Detector-specific Output Correction Factor for Various Collimator Systems
title_full Evaluation of International Atomic Energy Agency Technical Report Series-483 Detector-specific Output Correction Factor for Various Collimator Systems
title_fullStr Evaluation of International Atomic Energy Agency Technical Report Series-483 Detector-specific Output Correction Factor for Various Collimator Systems
title_full_unstemmed Evaluation of International Atomic Energy Agency Technical Report Series-483 Detector-specific Output Correction Factor for Various Collimator Systems
title_short Evaluation of International Atomic Energy Agency Technical Report Series-483 Detector-specific Output Correction Factor for Various Collimator Systems
title_sort evaluation of international atomic energy agency technical report series-483 detector-specific output correction factor for various collimator systems
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10642599/
https://www.ncbi.nlm.nih.gov/pubmed/37969152
http://dx.doi.org/10.4103/jmp.jmp_59_23
work_keys_str_mv AT kannanmageshraja evaluationofinternationalatomicenergyagencytechnicalreportseries483detectorspecificoutputcorrectionfactorforvariouscollimatorsystems
AT saminathansathiyan evaluationofinternationalatomicenergyagencytechnicalreportseries483detectorspecificoutputcorrectionfactorforvariouscollimatorsystems
AT chandrarajvaratharaj evaluationofinternationalatomicenergyagencytechnicalreportseries483detectorspecificoutputcorrectionfactorforvariouscollimatorsystems
AT rajdgowtham evaluationofinternationalatomicenergyagencytechnicalreportseries483detectorspecificoutputcorrectionfactorforvariouscollimatorsystems
AT ganeshkm evaluationofinternationalatomicenergyagencytechnicalreportseries483detectorspecificoutputcorrectionfactorforvariouscollimatorsystems