Development of a prediction model for target positioning by using diaphragm waveforms extracted from CBCT projection images

PURPOSE: To develop a prediction model (PM) for target positioning using diaphragm waveforms extracted from CBCT projection images. METHODS: Nineteen patients with lung cancer underwent orthogonal rotational kV x‐ray imaging lasting 70 s. IR markers placed on their abdominal surfaces and an implante...

Descripción completa

Detalles Bibliográficos
Autores principales: Sakurai, Yuta, Ambo, Shintaro, Nakamura, Mitsuhiro, Iramina, Hiraku, Iizuka, Yusuke, Mitsuyoshi, Takamasa, Matsuo, Yukinori, Mizowaki, Takashi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10647967/
https://www.ncbi.nlm.nih.gov/pubmed/37543990
http://dx.doi.org/10.1002/acm2.14112
_version_ 1785135228840312832
author Sakurai, Yuta
Ambo, Shintaro
Nakamura, Mitsuhiro
Iramina, Hiraku
Iizuka, Yusuke
Mitsuyoshi, Takamasa
Matsuo, Yukinori
Mizowaki, Takashi
author_facet Sakurai, Yuta
Ambo, Shintaro
Nakamura, Mitsuhiro
Iramina, Hiraku
Iizuka, Yusuke
Mitsuyoshi, Takamasa
Matsuo, Yukinori
Mizowaki, Takashi
author_sort Sakurai, Yuta
collection PubMed
description PURPOSE: To develop a prediction model (PM) for target positioning using diaphragm waveforms extracted from CBCT projection images. METHODS: Nineteen patients with lung cancer underwent orthogonal rotational kV x‐ray imaging lasting 70 s. IR markers placed on their abdominal surfaces and an implanted gold marker located nearest to the tumor were considered as external surrogates and the target, respectively. Four different types of regression‐based PM were trained using surrogate motions and target positions for the first 60 s, as follows: Scenario A: Based on the clinical scenario, 3D target positions extracted from projection images were used as they were (PM(CL)). Scenario B: The short‐arc 4D‐CBCT waveform exhibiting eight target positions was obtained by averaging the target positions in Scenario A. The waveform was repeated for 60 s (W (4D‐CBCT)) by adapting to the respiratory phase of the external surrogate. W (4D‐CBCT) was used as the target positions (PM(4D‐CBCT)). Scenario C: The Amsterdam Shroud (AS) signal, which depicted the diaphragm motion in the superior–inferior direction was extracted from the orthogonal projection images. The amplitude and phase of W (4D‐CBCT) were corrected based on the AS signal. The AS‐corrected W (4D‐CBCT) was used as the target positions (PM(AS‐4D‐CBCT)). Scenario D: The AS signal was extracted from single projection images. Other processes were the same as in Scenario C. The prediction errors were calculated for the remaining 10 s. RESULTS: The 3D prediction error within 3 mm was 77.3% for PM(4D‐CBCT), which was 12.8% lower than that for PM(CL). Using the diaphragm waveforms, the percentage of errors within 3 mm improved by approximately 7% to 84.0%‐85.3% for PM(AS‐4D‐CBCT) in Scenarios C and D, respectively. Statistically significant differences were observed between the prediction errors of PM(4D‐CBCT) and PM(AS‐4D‐CBCT). CONCLUSION: PM(AS‐4D‐CBCT) outperformed PM(4D‐CBCT), proving the efficacy of the AS signal‐based correction. PM(AS‐4D‐CBCT) would make it possible to predict target positions from 4D‐CBCT images without gold markers.
format Online
Article
Text
id pubmed-10647967
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-106479672023-08-06 Development of a prediction model for target positioning by using diaphragm waveforms extracted from CBCT projection images Sakurai, Yuta Ambo, Shintaro Nakamura, Mitsuhiro Iramina, Hiraku Iizuka, Yusuke Mitsuyoshi, Takamasa Matsuo, Yukinori Mizowaki, Takashi J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: To develop a prediction model (PM) for target positioning using diaphragm waveforms extracted from CBCT projection images. METHODS: Nineteen patients with lung cancer underwent orthogonal rotational kV x‐ray imaging lasting 70 s. IR markers placed on their abdominal surfaces and an implanted gold marker located nearest to the tumor were considered as external surrogates and the target, respectively. Four different types of regression‐based PM were trained using surrogate motions and target positions for the first 60 s, as follows: Scenario A: Based on the clinical scenario, 3D target positions extracted from projection images were used as they were (PM(CL)). Scenario B: The short‐arc 4D‐CBCT waveform exhibiting eight target positions was obtained by averaging the target positions in Scenario A. The waveform was repeated for 60 s (W (4D‐CBCT)) by adapting to the respiratory phase of the external surrogate. W (4D‐CBCT) was used as the target positions (PM(4D‐CBCT)). Scenario C: The Amsterdam Shroud (AS) signal, which depicted the diaphragm motion in the superior–inferior direction was extracted from the orthogonal projection images. The amplitude and phase of W (4D‐CBCT) were corrected based on the AS signal. The AS‐corrected W (4D‐CBCT) was used as the target positions (PM(AS‐4D‐CBCT)). Scenario D: The AS signal was extracted from single projection images. Other processes were the same as in Scenario C. The prediction errors were calculated for the remaining 10 s. RESULTS: The 3D prediction error within 3 mm was 77.3% for PM(4D‐CBCT), which was 12.8% lower than that for PM(CL). Using the diaphragm waveforms, the percentage of errors within 3 mm improved by approximately 7% to 84.0%‐85.3% for PM(AS‐4D‐CBCT) in Scenarios C and D, respectively. Statistically significant differences were observed between the prediction errors of PM(4D‐CBCT) and PM(AS‐4D‐CBCT). CONCLUSION: PM(AS‐4D‐CBCT) outperformed PM(4D‐CBCT), proving the efficacy of the AS signal‐based correction. PM(AS‐4D‐CBCT) would make it possible to predict target positions from 4D‐CBCT images without gold markers. John Wiley and Sons Inc. 2023-08-06 /pmc/articles/PMC10647967/ /pubmed/37543990 http://dx.doi.org/10.1002/acm2.14112 Text en © 2023 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Sakurai, Yuta
Ambo, Shintaro
Nakamura, Mitsuhiro
Iramina, Hiraku
Iizuka, Yusuke
Mitsuyoshi, Takamasa
Matsuo, Yukinori
Mizowaki, Takashi
Development of a prediction model for target positioning by using diaphragm waveforms extracted from CBCT projection images
title Development of a prediction model for target positioning by using diaphragm waveforms extracted from CBCT projection images
title_full Development of a prediction model for target positioning by using diaphragm waveforms extracted from CBCT projection images
title_fullStr Development of a prediction model for target positioning by using diaphragm waveforms extracted from CBCT projection images
title_full_unstemmed Development of a prediction model for target positioning by using diaphragm waveforms extracted from CBCT projection images
title_short Development of a prediction model for target positioning by using diaphragm waveforms extracted from CBCT projection images
title_sort development of a prediction model for target positioning by using diaphragm waveforms extracted from cbct projection images
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10647967/
https://www.ncbi.nlm.nih.gov/pubmed/37543990
http://dx.doi.org/10.1002/acm2.14112
work_keys_str_mv AT sakuraiyuta developmentofapredictionmodelfortargetpositioningbyusingdiaphragmwaveformsextractedfromcbctprojectionimages
AT amboshintaro developmentofapredictionmodelfortargetpositioningbyusingdiaphragmwaveformsextractedfromcbctprojectionimages
AT nakamuramitsuhiro developmentofapredictionmodelfortargetpositioningbyusingdiaphragmwaveformsextractedfromcbctprojectionimages
AT iraminahiraku developmentofapredictionmodelfortargetpositioningbyusingdiaphragmwaveformsextractedfromcbctprojectionimages
AT iizukayusuke developmentofapredictionmodelfortargetpositioningbyusingdiaphragmwaveformsextractedfromcbctprojectionimages
AT mitsuyoshitakamasa developmentofapredictionmodelfortargetpositioningbyusingdiaphragmwaveformsextractedfromcbctprojectionimages
AT matsuoyukinori developmentofapredictionmodelfortargetpositioningbyusingdiaphragmwaveformsextractedfromcbctprojectionimages
AT mizowakitakashi developmentofapredictionmodelfortargetpositioningbyusingdiaphragmwaveformsextractedfromcbctprojectionimages