Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume

INTRODUCTION: The objective of this laboratory study was to measure the effect of decreased lung compliance and endotracheal tube (ETT) leakage on measured exhaled tidal volume at the airway and at the ventilator, in a research study with a test lung. METHODS: The subjects were infant, adult and ped...

Descripción completa

Detalles Bibliográficos
Autores principales: Al-Majed, Sami I, Thompson, John E, Watson, Kenneth F, Randolph, Adrienne G
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2004
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1065056/
https://www.ncbi.nlm.nih.gov/pubmed/15566583
http://dx.doi.org/10.1186/cc2954
_version_ 1782123336115945472
author Al-Majed, Sami I
Thompson, John E
Watson, Kenneth F
Randolph, Adrienne G
author_facet Al-Majed, Sami I
Thompson, John E
Watson, Kenneth F
Randolph, Adrienne G
author_sort Al-Majed, Sami I
collection PubMed
description INTRODUCTION: The objective of this laboratory study was to measure the effect of decreased lung compliance and endotracheal tube (ETT) leakage on measured exhaled tidal volume at the airway and at the ventilator, in a research study with a test lung. METHODS: The subjects were infant, adult and pediatric test lungs. In the test lung model, lung compliances were set to normal and to levels seen in acute respiratory distress syndrome. Set tidal volume was 6 ml/kg across a range of simulated weights and ETT sizes. Data were recorded from both the ventilator light-emitting diode display and the CO(2)SMO Plus monitor display by a single observer. Effective tidal volume was calculated from a standard equation. RESULTS: In all test lung models, exhaled tidal volume measured at the airway decreased markedly with decreasing lung compliance, but measurement at the ventilator showed minimal change. In the absence of a simulated ETT leak, calculation of the effective tidal volume led to measurements very similar to exhaled tidal volume measured at the ETT. With a simulated ETT tube leak, the effective tidal volume markedly overestimated tidal volume measured at the airway. CONCLUSION: Previous investigators have emphasized the need to measure tidal volume at the ETT for all children. When ETT leakage is minimal, it seems from our simulated lung models that calculation of effective tidal volume would give similar readings to tidal volume measured at the airway, even in small patients. Future studies of tidal volume measurement accuracy in mechanically ventilated children should control for the degree of ETT leakage.
format Text
id pubmed-1065056
institution National Center for Biotechnology Information
language English
publishDate 2004
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-10650562005-03-16 Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume Al-Majed, Sami I Thompson, John E Watson, Kenneth F Randolph, Adrienne G Crit Care Research INTRODUCTION: The objective of this laboratory study was to measure the effect of decreased lung compliance and endotracheal tube (ETT) leakage on measured exhaled tidal volume at the airway and at the ventilator, in a research study with a test lung. METHODS: The subjects were infant, adult and pediatric test lungs. In the test lung model, lung compliances were set to normal and to levels seen in acute respiratory distress syndrome. Set tidal volume was 6 ml/kg across a range of simulated weights and ETT sizes. Data were recorded from both the ventilator light-emitting diode display and the CO(2)SMO Plus monitor display by a single observer. Effective tidal volume was calculated from a standard equation. RESULTS: In all test lung models, exhaled tidal volume measured at the airway decreased markedly with decreasing lung compliance, but measurement at the ventilator showed minimal change. In the absence of a simulated ETT leak, calculation of the effective tidal volume led to measurements very similar to exhaled tidal volume measured at the ETT. With a simulated ETT tube leak, the effective tidal volume markedly overestimated tidal volume measured at the airway. CONCLUSION: Previous investigators have emphasized the need to measure tidal volume at the ETT for all children. When ETT leakage is minimal, it seems from our simulated lung models that calculation of effective tidal volume would give similar readings to tidal volume measured at the airway, even in small patients. Future studies of tidal volume measurement accuracy in mechanically ventilated children should control for the degree of ETT leakage. BioMed Central 2004 2004-10-06 /pmc/articles/PMC1065056/ /pubmed/15566583 http://dx.doi.org/10.1186/cc2954 Text en Copyright © 2004 Al-Majed et al.; licensee BioMed Central Ltd.
spellingShingle Research
Al-Majed, Sami I
Thompson, John E
Watson, Kenneth F
Randolph, Adrienne G
Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
title Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
title_full Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
title_fullStr Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
title_full_unstemmed Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
title_short Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
title_sort effect of lung compliance and endotracheal tube leakage on measurement of tidal volume
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1065056/
https://www.ncbi.nlm.nih.gov/pubmed/15566583
http://dx.doi.org/10.1186/cc2954
work_keys_str_mv AT almajedsamii effectoflungcomplianceandendotrachealtubeleakageonmeasurementoftidalvolume
AT thompsonjohne effectoflungcomplianceandendotrachealtubeleakageonmeasurementoftidalvolume
AT watsonkennethf effectoflungcomplianceandendotrachealtubeleakageonmeasurementoftidalvolume
AT randolphadrienneg effectoflungcomplianceandendotrachealtubeleakageonmeasurementoftidalvolume