True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study

SentriO Oxy™ is a newly available, Food and Drug Administration-approved oxygenation mask system that provides high oxygenation, even on low-flow (5–10 L/min) oxygen. This study aimed to accurately measure the intratracheal fraction of inspired oxygen (FiO(2)) using SentriO Oxy™ masks under relative...

Descripción completa

Detalles Bibliográficos
Autores principales: Chiang, Cheng, Teng, Wei-Nung, Chiang, Ting-Yun, Huang, Chao-Lan, Lin, Shi-Pin, Chang, Wen-Kuei, Ting, Chien-Kun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Netherlands 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10651707/
https://www.ncbi.nlm.nih.gov/pubmed/37828296
http://dx.doi.org/10.1007/s10877-023-01076-4
_version_ 1785136050336694272
author Chiang, Cheng
Teng, Wei-Nung
Chiang, Ting-Yun
Huang, Chao-Lan
Lin, Shi-Pin
Chang, Wen-Kuei
Ting, Chien-Kun
author_facet Chiang, Cheng
Teng, Wei-Nung
Chiang, Ting-Yun
Huang, Chao-Lan
Lin, Shi-Pin
Chang, Wen-Kuei
Ting, Chien-Kun
author_sort Chiang, Cheng
collection PubMed
description SentriO Oxy™ is a newly available, Food and Drug Administration-approved oxygenation mask system that provides high oxygenation, even on low-flow (5–10 L/min) oxygen. This study aimed to accurately measure the intratracheal fraction of inspired oxygen (FiO(2)) using SentriO Oxy™ masks under relatively low oxygen flow rates. A manikin-ventilator-test lung simulation system was used. We measured FiO(2) at the level of the carina, 5 minutes after applying 45 different respiratory parameter combinations using SentriO Oxy™ masks. Tidal volume (TV) was set to 300, 500, and 700 mL; respiratory rate (RR) was set to 8, 12, 16, 20, and 24 breaths per minute; and oxygen flow rate was set to 6, 8, and 10 L/min. Our hypothesis was that FiO(2) would be proportional to the difference between oxygen flow rate and minute ventilation. FiO(2) measured by smaller TV, lower RR, or higher oxygen flows revealed a significantly higher value, confirming our hypothesis. In addition, using linear regression analysis, we found that TV, RR, and oxygen flow were all significant factors influencing the measured FiO(2). Our experiment proposed two prediction equations considering the oxygen flow rate, TV, and RR. The results of our study may provide information and prediction of FiO(2) for clinicians to use SentriO Oxy™ masks during sedative anesthetic procedures under low oxygen flow rates.
format Online
Article
Text
id pubmed-10651707
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Springer Netherlands
record_format MEDLINE/PubMed
spelling pubmed-106517072023-10-13 True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study Chiang, Cheng Teng, Wei-Nung Chiang, Ting-Yun Huang, Chao-Lan Lin, Shi-Pin Chang, Wen-Kuei Ting, Chien-Kun J Clin Monit Comput Original Research SentriO Oxy™ is a newly available, Food and Drug Administration-approved oxygenation mask system that provides high oxygenation, even on low-flow (5–10 L/min) oxygen. This study aimed to accurately measure the intratracheal fraction of inspired oxygen (FiO(2)) using SentriO Oxy™ masks under relatively low oxygen flow rates. A manikin-ventilator-test lung simulation system was used. We measured FiO(2) at the level of the carina, 5 minutes after applying 45 different respiratory parameter combinations using SentriO Oxy™ masks. Tidal volume (TV) was set to 300, 500, and 700 mL; respiratory rate (RR) was set to 8, 12, 16, 20, and 24 breaths per minute; and oxygen flow rate was set to 6, 8, and 10 L/min. Our hypothesis was that FiO(2) would be proportional to the difference between oxygen flow rate and minute ventilation. FiO(2) measured by smaller TV, lower RR, or higher oxygen flows revealed a significantly higher value, confirming our hypothesis. In addition, using linear regression analysis, we found that TV, RR, and oxygen flow were all significant factors influencing the measured FiO(2). Our experiment proposed two prediction equations considering the oxygen flow rate, TV, and RR. The results of our study may provide information and prediction of FiO(2) for clinicians to use SentriO Oxy™ masks during sedative anesthetic procedures under low oxygen flow rates. Springer Netherlands 2023-10-13 2023 /pmc/articles/PMC10651707/ /pubmed/37828296 http://dx.doi.org/10.1007/s10877-023-01076-4 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Research
Chiang, Cheng
Teng, Wei-Nung
Chiang, Ting-Yun
Huang, Chao-Lan
Lin, Shi-Pin
Chang, Wen-Kuei
Ting, Chien-Kun
True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study
title True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study
title_full True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study
title_fullStr True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study
title_full_unstemmed True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study
title_short True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study
title_sort true intratracheal oxygen concentration delivered by sentrio oxy™ masks under various respiratory conditions: a bench study
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10651707/
https://www.ncbi.nlm.nih.gov/pubmed/37828296
http://dx.doi.org/10.1007/s10877-023-01076-4
work_keys_str_mv AT chiangcheng trueintratrachealoxygenconcentrationdeliveredbysentriooxymasksundervariousrespiratoryconditionsabenchstudy
AT tengweinung trueintratrachealoxygenconcentrationdeliveredbysentriooxymasksundervariousrespiratoryconditionsabenchstudy
AT chiangtingyun trueintratrachealoxygenconcentrationdeliveredbysentriooxymasksundervariousrespiratoryconditionsabenchstudy
AT huangchaolan trueintratrachealoxygenconcentrationdeliveredbysentriooxymasksundervariousrespiratoryconditionsabenchstudy
AT linshipin trueintratrachealoxygenconcentrationdeliveredbysentriooxymasksundervariousrespiratoryconditionsabenchstudy
AT changwenkuei trueintratrachealoxygenconcentrationdeliveredbysentriooxymasksundervariousrespiratoryconditionsabenchstudy
AT tingchienkun trueintratrachealoxygenconcentrationdeliveredbysentriooxymasksundervariousrespiratoryconditionsabenchstudy