Cargando…

Aortic rupture during STABILISE (stent-assisted balloon-induced intimal disruption and relamination in aortic dissection repair) technique

The STABILISE (stent-assisted balloon-induced intimal disruption and relamination in aortic dissection repair) technique has shown promising results for treating type B aortic dissections, but the potential exists for fatal adverse effects. We present a case of infrarenal aortic rupture while using...

Descripción completa

Detalles Bibliográficos
Autores principales: Hsu, Ashley C., DiBartolomeo, Alexander D., Han, Sukgu M., Fleischman, Fernando, Magee, Gregory A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10665655/
https://www.ncbi.nlm.nih.gov/pubmed/38023324
http://dx.doi.org/10.1016/j.jvscit.2023.101338
Descripción
Sumario:The STABILISE (stent-assisted balloon-induced intimal disruption and relamination in aortic dissection repair) technique has shown promising results for treating type B aortic dissections, but the potential exists for fatal adverse effects. We present a case of infrarenal aortic rupture while using a compliant balloon to balloon mold the true lumen inside previously placed bare metal stents during the STABILISE technique. Caution is advised for providers who wish to perform the STABILISE technique, and we recommend using a semi-compliant balloon sized to the smallest total aortic diameter to mitigate the risk of rupture.