Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring
Our aim is to develop a blood pressure (BP) measurement technology that could be integrated into a finger-worn pulse oximeter, eliminating the need for a brachial cuff. We present a miniature cuffless tonometric finger probe system that uses the oscillometric method to measure BP. Our approach uses...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10665806/ https://www.ncbi.nlm.nih.gov/pubmed/38026187 http://dx.doi.org/10.1016/j.isci.2023.108295 |
Sumario: | Our aim is to develop a blood pressure (BP) measurement technology that could be integrated into a finger-worn pulse oximeter, eliminating the need for a brachial cuff. We present a miniature cuffless tonometric finger probe system that uses the oscillometric method to measure BP. Our approach uses a motorized press that is used to apply pressure to the fingertip to measure BP. We verified the functionality of the device in a clinical trial (n = 43) resulting in systolic and diastolic pressures ((mean ± SD) mmHg) of (−3.5 ± 8.4) mmHg and (−4.0 ± 4.4) mmHg, respectively. Comparison was made with manual auscultation (n = 26) and automated cuff oscillometry (n = 18). In addition to BP, we demonstrated the ability of the device to assess arterial stiffness (n = 18) and detect atrial fibrillation (n = 6). We were able to introduce a sufficiently small device that could be used for convenient ambulatory measurements with minimal discomfort. |
---|