Cargando…
Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring
Our aim is to develop a blood pressure (BP) measurement technology that could be integrated into a finger-worn pulse oximeter, eliminating the need for a brachial cuff. We present a miniature cuffless tonometric finger probe system that uses the oscillometric method to measure BP. Our approach uses...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10665806/ https://www.ncbi.nlm.nih.gov/pubmed/38026187 http://dx.doi.org/10.1016/j.isci.2023.108295 |
_version_ | 1785138907296301056 |
---|---|
author | Panula, Tuukka Sirkiä, Jukka-Pekka Koivisto, Tero Pänkäälä, Mikko Niiranen, Teemu Kantola, Ilkka Kaisti, Matti |
author_facet | Panula, Tuukka Sirkiä, Jukka-Pekka Koivisto, Tero Pänkäälä, Mikko Niiranen, Teemu Kantola, Ilkka Kaisti, Matti |
author_sort | Panula, Tuukka |
collection | PubMed |
description | Our aim is to develop a blood pressure (BP) measurement technology that could be integrated into a finger-worn pulse oximeter, eliminating the need for a brachial cuff. We present a miniature cuffless tonometric finger probe system that uses the oscillometric method to measure BP. Our approach uses a motorized press that is used to apply pressure to the fingertip to measure BP. We verified the functionality of the device in a clinical trial (n = 43) resulting in systolic and diastolic pressures ((mean ± SD) mmHg) of (−3.5 ± 8.4) mmHg and (−4.0 ± 4.4) mmHg, respectively. Comparison was made with manual auscultation (n = 26) and automated cuff oscillometry (n = 18). In addition to BP, we demonstrated the ability of the device to assess arterial stiffness (n = 18) and detect atrial fibrillation (n = 6). We were able to introduce a sufficiently small device that could be used for convenient ambulatory measurements with minimal discomfort. |
format | Online Article Text |
id | pubmed-10665806 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-106658062023-10-24 Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring Panula, Tuukka Sirkiä, Jukka-Pekka Koivisto, Tero Pänkäälä, Mikko Niiranen, Teemu Kantola, Ilkka Kaisti, Matti iScience Article Our aim is to develop a blood pressure (BP) measurement technology that could be integrated into a finger-worn pulse oximeter, eliminating the need for a brachial cuff. We present a miniature cuffless tonometric finger probe system that uses the oscillometric method to measure BP. Our approach uses a motorized press that is used to apply pressure to the fingertip to measure BP. We verified the functionality of the device in a clinical trial (n = 43) resulting in systolic and diastolic pressures ((mean ± SD) mmHg) of (−3.5 ± 8.4) mmHg and (−4.0 ± 4.4) mmHg, respectively. Comparison was made with manual auscultation (n = 26) and automated cuff oscillometry (n = 18). In addition to BP, we demonstrated the ability of the device to assess arterial stiffness (n = 18) and detect atrial fibrillation (n = 6). We were able to introduce a sufficiently small device that could be used for convenient ambulatory measurements with minimal discomfort. Elsevier 2023-10-24 /pmc/articles/PMC10665806/ /pubmed/38026187 http://dx.doi.org/10.1016/j.isci.2023.108295 Text en © 2023 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Article Panula, Tuukka Sirkiä, Jukka-Pekka Koivisto, Tero Pänkäälä, Mikko Niiranen, Teemu Kantola, Ilkka Kaisti, Matti Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring |
title | Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring |
title_full | Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring |
title_fullStr | Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring |
title_full_unstemmed | Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring |
title_short | Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring |
title_sort | development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10665806/ https://www.ncbi.nlm.nih.gov/pubmed/38026187 http://dx.doi.org/10.1016/j.isci.2023.108295 |
work_keys_str_mv | AT panulatuukka developmentandclinicalvalidationofaminiaturizedfingerprobeforbedsidehemodynamicmonitoring AT sirkiajukkapekka developmentandclinicalvalidationofaminiaturizedfingerprobeforbedsidehemodynamicmonitoring AT koivistotero developmentandclinicalvalidationofaminiaturizedfingerprobeforbedsidehemodynamicmonitoring AT pankaalamikko developmentandclinicalvalidationofaminiaturizedfingerprobeforbedsidehemodynamicmonitoring AT niiranenteemu developmentandclinicalvalidationofaminiaturizedfingerprobeforbedsidehemodynamicmonitoring AT kantolailkka developmentandclinicalvalidationofaminiaturizedfingerprobeforbedsidehemodynamicmonitoring AT kaistimatti developmentandclinicalvalidationofaminiaturizedfingerprobeforbedsidehemodynamicmonitoring |