Optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted NMR experiments

INTRODUCTION: Diffusion weighting in optically detected magnetic resonance experiments involving diamond nitrogen-vacancy (NV) centers can provide valuable microstructural information. Bi-planar gradient coils employed for diffusion weighting afford excellent spatial access, essential for integratin...

Descripción completa

Detalles Bibliográficos
Autores principales: Amrein, Philipp, Bruckmaier, Fleming, Jia, Feng, Bucher, Dominik B., Zaitsev, Maxim, Littin, Sebastian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10667408/
https://www.ncbi.nlm.nih.gov/pubmed/37578612
http://dx.doi.org/10.1007/s10334-023-01111-0
_version_ 1785139242281730048
author Amrein, Philipp
Bruckmaier, Fleming
Jia, Feng
Bucher, Dominik B.
Zaitsev, Maxim
Littin, Sebastian
author_facet Amrein, Philipp
Bruckmaier, Fleming
Jia, Feng
Bucher, Dominik B.
Zaitsev, Maxim
Littin, Sebastian
author_sort Amrein, Philipp
collection PubMed
description INTRODUCTION: Diffusion weighting in optically detected magnetic resonance experiments involving diamond nitrogen-vacancy (NV) centers can provide valuable microstructural information. Bi-planar gradient coils employed for diffusion weighting afford excellent spatial access, essential for integrating the NV-NMR components. Nevertheless, owing to the polar tilt of roughly [Formula: see text] of the diamond NV center, the primary magnetic field direction must be taken into account accordingly. METHODS: To determine the most effective bi-planar gradient coil configurations, we conducted an investigation into the impact of various factors, including the square side length, surface separation, and surface orientation. This was accomplished by generating over 500 bi-planar surface configurations using automated methods. RESULTS: We successfully generated and evaluated coil layouts in terms of sensitivity and field accuracy. Interestingly, inclined bi-planar orientations close to the NV–NMR setup’s requirement, showed higher sensitivity for the transverse gradient channels than horizontal or vertical orientations. We fabricated a suitable solution as a three-channel bi-planar double-layered PCB system and experimentally validated the sensitivities at [Formula: see text] and [Formula: see text] for the transverse [Formula: see text] and [Formula: see text] gradients, and [Formula: see text] for the [Formula: see text] gradient. DISCUSSION: We found that the chosen relative bi-planar tilt of [Formula: see text] represents a reasonable compromise in terms of overall performance and allows for easier coil implementation with a straight, horizontal alignment within the overall experimental setup.
format Online
Article
Text
id pubmed-10667408
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-106674082023-08-14 Optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted NMR experiments Amrein, Philipp Bruckmaier, Fleming Jia, Feng Bucher, Dominik B. Zaitsev, Maxim Littin, Sebastian MAGMA Research Article INTRODUCTION: Diffusion weighting in optically detected magnetic resonance experiments involving diamond nitrogen-vacancy (NV) centers can provide valuable microstructural information. Bi-planar gradient coils employed for diffusion weighting afford excellent spatial access, essential for integrating the NV-NMR components. Nevertheless, owing to the polar tilt of roughly [Formula: see text] of the diamond NV center, the primary magnetic field direction must be taken into account accordingly. METHODS: To determine the most effective bi-planar gradient coil configurations, we conducted an investigation into the impact of various factors, including the square side length, surface separation, and surface orientation. This was accomplished by generating over 500 bi-planar surface configurations using automated methods. RESULTS: We successfully generated and evaluated coil layouts in terms of sensitivity and field accuracy. Interestingly, inclined bi-planar orientations close to the NV–NMR setup’s requirement, showed higher sensitivity for the transverse gradient channels than horizontal or vertical orientations. We fabricated a suitable solution as a three-channel bi-planar double-layered PCB system and experimentally validated the sensitivities at [Formula: see text] and [Formula: see text] for the transverse [Formula: see text] and [Formula: see text] gradients, and [Formula: see text] for the [Formula: see text] gradient. DISCUSSION: We found that the chosen relative bi-planar tilt of [Formula: see text] represents a reasonable compromise in terms of overall performance and allows for easier coil implementation with a straight, horizontal alignment within the overall experimental setup. Springer International Publishing 2023-08-14 2023 /pmc/articles/PMC10667408/ /pubmed/37578612 http://dx.doi.org/10.1007/s10334-023-01111-0 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Research Article
Amrein, Philipp
Bruckmaier, Fleming
Jia, Feng
Bucher, Dominik B.
Zaitsev, Maxim
Littin, Sebastian
Optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted NMR experiments
title Optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted NMR experiments
title_full Optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted NMR experiments
title_fullStr Optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted NMR experiments
title_full_unstemmed Optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted NMR experiments
title_short Optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted NMR experiments
title_sort optimal bi-planar gradient coil configurations for diamond nitrogen-vacancy based diffusion-weighted nmr experiments
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10667408/
https://www.ncbi.nlm.nih.gov/pubmed/37578612
http://dx.doi.org/10.1007/s10334-023-01111-0
work_keys_str_mv AT amreinphilipp optimalbiplanargradientcoilconfigurationsfordiamondnitrogenvacancybaseddiffusionweightednmrexperiments
AT bruckmaierfleming optimalbiplanargradientcoilconfigurationsfordiamondnitrogenvacancybaseddiffusionweightednmrexperiments
AT jiafeng optimalbiplanargradientcoilconfigurationsfordiamondnitrogenvacancybaseddiffusionweightednmrexperiments
AT bucherdominikb optimalbiplanargradientcoilconfigurationsfordiamondnitrogenvacancybaseddiffusionweightednmrexperiments
AT zaitsevmaxim optimalbiplanargradientcoilconfigurationsfordiamondnitrogenvacancybaseddiffusionweightednmrexperiments
AT littinsebastian optimalbiplanargradientcoilconfigurationsfordiamondnitrogenvacancybaseddiffusionweightednmrexperiments