A Pilot Study Examining the Dielectric Response of Human Forearm Tissues
This work aims to describe the dielectric behaviors of four main tissues in the human forearm using mathematical modelling, including fat, muscle, blood and bone. Multi-frequency bioimpedance analysis (MF-BIA) was initially performed using the finite element method (FEM) with a 3D forearm model to e...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10669245/ https://www.ncbi.nlm.nih.gov/pubmed/37998136 http://dx.doi.org/10.3390/bios13110961 |
_version_ | 1785139650413723648 |
---|---|
author | Yu, Yang Kalra, Anubha Manju Anand, Gautam Lowe, Andrew |
author_facet | Yu, Yang Kalra, Anubha Manju Anand, Gautam Lowe, Andrew |
author_sort | Yu, Yang |
collection | PubMed |
description | This work aims to describe the dielectric behaviors of four main tissues in the human forearm using mathematical modelling, including fat, muscle, blood and bone. Multi-frequency bioimpedance analysis (MF-BIA) was initially performed using the finite element method (FEM) with a 3D forearm model to estimate impedance spectra from 10 kHz to 1 MHz, followed by a pilot study involving two healthy subjects to characterize the response of actual forearm tissues from 1 kHz to 349 kHz. Both the simulation and experimental results were fitted to a single-dispersion Cole model (SDCM) and a multi-dispersion Cole model (MDCM) to determine the Cole parameters for each tissue. Cole-type responses of both simulated and actual human forearms were observed. A paired t-test based on the root mean squared error (RMSE) values indicated that both Cole models performed comparably in fitting both simulated and measured bioimpedance data. However, MDCM exhibited higher accuracy, with a correlation coefficient (R(2)) of 0.99 and 0.89, RMSE of 0.22 [Formula: see text] and 0.56 [Formula: see text] , mean difference (mean [Formula: see text] standard deviation) of 0.00 [Formula: see text] 0.23 [Formula: see text] and −0.28 [Formula: see text] 0.23 [Formula: see text] , and mean absolute error (MAE) of 0.0007 [Formula: see text] and 0.2789 [Formula: see text] for the real part and imaginary part of impedance, respectively. Determining the electrical response of multi-tissues can be helpful in developing physiological monitoring of an organ or a section of the human body through MF-BIA and hemodynamic monitoring by filtering out the impedance contributions from the surrounding tissues to blood-flow-induced impedance variations. |
format | Online Article Text |
id | pubmed-10669245 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-106692452023-10-29 A Pilot Study Examining the Dielectric Response of Human Forearm Tissues Yu, Yang Kalra, Anubha Manju Anand, Gautam Lowe, Andrew Biosensors (Basel) Article This work aims to describe the dielectric behaviors of four main tissues in the human forearm using mathematical modelling, including fat, muscle, blood and bone. Multi-frequency bioimpedance analysis (MF-BIA) was initially performed using the finite element method (FEM) with a 3D forearm model to estimate impedance spectra from 10 kHz to 1 MHz, followed by a pilot study involving two healthy subjects to characterize the response of actual forearm tissues from 1 kHz to 349 kHz. Both the simulation and experimental results were fitted to a single-dispersion Cole model (SDCM) and a multi-dispersion Cole model (MDCM) to determine the Cole parameters for each tissue. Cole-type responses of both simulated and actual human forearms were observed. A paired t-test based on the root mean squared error (RMSE) values indicated that both Cole models performed comparably in fitting both simulated and measured bioimpedance data. However, MDCM exhibited higher accuracy, with a correlation coefficient (R(2)) of 0.99 and 0.89, RMSE of 0.22 [Formula: see text] and 0.56 [Formula: see text] , mean difference (mean [Formula: see text] standard deviation) of 0.00 [Formula: see text] 0.23 [Formula: see text] and −0.28 [Formula: see text] 0.23 [Formula: see text] , and mean absolute error (MAE) of 0.0007 [Formula: see text] and 0.2789 [Formula: see text] for the real part and imaginary part of impedance, respectively. Determining the electrical response of multi-tissues can be helpful in developing physiological monitoring of an organ or a section of the human body through MF-BIA and hemodynamic monitoring by filtering out the impedance contributions from the surrounding tissues to blood-flow-induced impedance variations. MDPI 2023-10-29 /pmc/articles/PMC10669245/ /pubmed/37998136 http://dx.doi.org/10.3390/bios13110961 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Yu, Yang Kalra, Anubha Manju Anand, Gautam Lowe, Andrew A Pilot Study Examining the Dielectric Response of Human Forearm Tissues |
title | A Pilot Study Examining the Dielectric Response of Human Forearm Tissues |
title_full | A Pilot Study Examining the Dielectric Response of Human Forearm Tissues |
title_fullStr | A Pilot Study Examining the Dielectric Response of Human Forearm Tissues |
title_full_unstemmed | A Pilot Study Examining the Dielectric Response of Human Forearm Tissues |
title_short | A Pilot Study Examining the Dielectric Response of Human Forearm Tissues |
title_sort | pilot study examining the dielectric response of human forearm tissues |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10669245/ https://www.ncbi.nlm.nih.gov/pubmed/37998136 http://dx.doi.org/10.3390/bios13110961 |
work_keys_str_mv | AT yuyang apilotstudyexaminingthedielectricresponseofhumanforearmtissues AT kalraanubhamanju apilotstudyexaminingthedielectricresponseofhumanforearmtissues AT anandgautam apilotstudyexaminingthedielectricresponseofhumanforearmtissues AT loweandrew apilotstudyexaminingthedielectricresponseofhumanforearmtissues AT yuyang pilotstudyexaminingthedielectricresponseofhumanforearmtissues AT kalraanubhamanju pilotstudyexaminingthedielectricresponseofhumanforearmtissues AT anandgautam pilotstudyexaminingthedielectricresponseofhumanforearmtissues AT loweandrew pilotstudyexaminingthedielectricresponseofhumanforearmtissues |