Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study

The locking plate may provide improved fixation in osteoporotic bone; however, it has been reported to fail due to varus collapse or screw perforation of the articular surface, especially in osteoporotic bone with medial cortex comminution. Using bone graft as an intramedullary strut together with p...

Descripción completa

Detalles Bibliográficos
Autores principales: Hsiao, Chih-Kun, Chiu, Yen-Wei, Hsiao, Hao-Yuan, Tsai, Yi-Jung, Lee, Cheng-Hung, Yen, Cheng-Yo, Tu, Yuan-Kun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10672415/
https://www.ncbi.nlm.nih.gov/pubmed/38004250
http://dx.doi.org/10.3390/life13112110
_version_ 1785140385909047296
author Hsiao, Chih-Kun
Chiu, Yen-Wei
Hsiao, Hao-Yuan
Tsai, Yi-Jung
Lee, Cheng-Hung
Yen, Cheng-Yo
Tu, Yuan-Kun
author_facet Hsiao, Chih-Kun
Chiu, Yen-Wei
Hsiao, Hao-Yuan
Tsai, Yi-Jung
Lee, Cheng-Hung
Yen, Cheng-Yo
Tu, Yuan-Kun
author_sort Hsiao, Chih-Kun
collection PubMed
description The locking plate may provide improved fixation in osteoporotic bone; however, it has been reported to fail due to varus collapse or screw perforation of the articular surface, especially in osteoporotic bone with medial cortex comminution. Using bone graft as an intramedullary strut together with plate fixation may result in a stronger construct. However, the drawbacks of bone grafts include limited supply, high cost, and infection risk. PMMA (so-called bone cement) has been widely used for implant fixation due to its good mechanical properties, fabricability, and biocompatibility. The risk of donor-site infection and the drawbacks of allografting may be overcome by considering PMMA struts as alternatives to fibular grafts for humeral intramedullary grafting surgeries. However, the potential effects of intramedullary PMMA strut on the dynamic behaviour of osteoporotic humerus fractures remain unclear. This study aimed to investigate the influence of an intramedullary PMMA strut on the stability of unstable proximal humeral fractures in an osteoporotic synthetic model. Two fixation techniques, a locking plate alone (non-strut group) and the same fixation augmented with an intramedullary PMMA strut (with-strut group), were cyclically tested in 20 artificial humeral models. Axially cyclic testing was performed to 450 N for 10,000 cycles, intercyclic motion, cumulated fragment migration, and residual deformation of the constructs were determined at periodic cyclic intervals, and the groups were compared. Results showed that adding an intramedullary PMMA strut could decrease 1.6 times intercyclic motion, 2 times cumulated fracture gap migration, and 1.8 times residual deformation from non-strut fixation. During cycling, neither screw pull-out, cut-through, nor implant failure was observed in the strut-augmented group. We concluded that the plate-strut mechanism could enhance the cyclic stability of the fixation and minimize the residual displacement of the fragment in treating osteoporotic proximal humeral unstable fractures. The PMMA strut has the potential to substitute donor bone and serve as an intramedullary support when used in combination with locking plate fixation. The intramedullary support with bone cement can be considered a solution in the treatment of osteoporotic proximal humeral fractures, especially when there is medial comminution.
format Online
Article
Text
id pubmed-10672415
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-106724152023-10-24 Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study Hsiao, Chih-Kun Chiu, Yen-Wei Hsiao, Hao-Yuan Tsai, Yi-Jung Lee, Cheng-Hung Yen, Cheng-Yo Tu, Yuan-Kun Life (Basel) Article The locking plate may provide improved fixation in osteoporotic bone; however, it has been reported to fail due to varus collapse or screw perforation of the articular surface, especially in osteoporotic bone with medial cortex comminution. Using bone graft as an intramedullary strut together with plate fixation may result in a stronger construct. However, the drawbacks of bone grafts include limited supply, high cost, and infection risk. PMMA (so-called bone cement) has been widely used for implant fixation due to its good mechanical properties, fabricability, and biocompatibility. The risk of donor-site infection and the drawbacks of allografting may be overcome by considering PMMA struts as alternatives to fibular grafts for humeral intramedullary grafting surgeries. However, the potential effects of intramedullary PMMA strut on the dynamic behaviour of osteoporotic humerus fractures remain unclear. This study aimed to investigate the influence of an intramedullary PMMA strut on the stability of unstable proximal humeral fractures in an osteoporotic synthetic model. Two fixation techniques, a locking plate alone (non-strut group) and the same fixation augmented with an intramedullary PMMA strut (with-strut group), were cyclically tested in 20 artificial humeral models. Axially cyclic testing was performed to 450 N for 10,000 cycles, intercyclic motion, cumulated fragment migration, and residual deformation of the constructs were determined at periodic cyclic intervals, and the groups were compared. Results showed that adding an intramedullary PMMA strut could decrease 1.6 times intercyclic motion, 2 times cumulated fracture gap migration, and 1.8 times residual deformation from non-strut fixation. During cycling, neither screw pull-out, cut-through, nor implant failure was observed in the strut-augmented group. We concluded that the plate-strut mechanism could enhance the cyclic stability of the fixation and minimize the residual displacement of the fragment in treating osteoporotic proximal humeral unstable fractures. The PMMA strut has the potential to substitute donor bone and serve as an intramedullary support when used in combination with locking plate fixation. The intramedullary support with bone cement can be considered a solution in the treatment of osteoporotic proximal humeral fractures, especially when there is medial comminution. MDPI 2023-10-24 /pmc/articles/PMC10672415/ /pubmed/38004250 http://dx.doi.org/10.3390/life13112110 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Hsiao, Chih-Kun
Chiu, Yen-Wei
Hsiao, Hao-Yuan
Tsai, Yi-Jung
Lee, Cheng-Hung
Yen, Cheng-Yo
Tu, Yuan-Kun
Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study
title Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study
title_full Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study
title_fullStr Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study
title_full_unstemmed Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study
title_short Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study
title_sort cyclic stability of locking plate augmented with intramedullary polymethyl methacrylate (pmma) strut fixation for osteoporotic humeral fractures: a biomechanical study
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10672415/
https://www.ncbi.nlm.nih.gov/pubmed/38004250
http://dx.doi.org/10.3390/life13112110
work_keys_str_mv AT hsiaochihkun cyclicstabilityoflockingplateaugmentedwithintramedullarypolymethylmethacrylatepmmastrutfixationforosteoporotichumeralfracturesabiomechanicalstudy
AT chiuyenwei cyclicstabilityoflockingplateaugmentedwithintramedullarypolymethylmethacrylatepmmastrutfixationforosteoporotichumeralfracturesabiomechanicalstudy
AT hsiaohaoyuan cyclicstabilityoflockingplateaugmentedwithintramedullarypolymethylmethacrylatepmmastrutfixationforosteoporotichumeralfracturesabiomechanicalstudy
AT tsaiyijung cyclicstabilityoflockingplateaugmentedwithintramedullarypolymethylmethacrylatepmmastrutfixationforosteoporotichumeralfracturesabiomechanicalstudy
AT leechenghung cyclicstabilityoflockingplateaugmentedwithintramedullarypolymethylmethacrylatepmmastrutfixationforosteoporotichumeralfracturesabiomechanicalstudy
AT yenchengyo cyclicstabilityoflockingplateaugmentedwithintramedullarypolymethylmethacrylatepmmastrutfixationforosteoporotichumeralfracturesabiomechanicalstudy
AT tuyuankun cyclicstabilityoflockingplateaugmentedwithintramedullarypolymethylmethacrylatepmmastrutfixationforosteoporotichumeralfracturesabiomechanicalstudy