Cargando…

Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use

Continuous glucose monitors (CGMs) are valuable tools for improving glycemic control, yet their accuracy might be influenced by physical activity. This study sought to assess the accuracy of the three latest factory-calibrated CGM systems available in Europe at the time the study was conducted, both...

Descripción completa

Detalles Bibliográficos
Autores principales: Lundemose, Sissel Banner, Laugesen, Christian, Ranjan, Ajenthen Gayathri, Nørgaard, Kirsten
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10675113/
https://www.ncbi.nlm.nih.gov/pubmed/38005642
http://dx.doi.org/10.3390/s23229256
_version_ 1785149783242964992
author Lundemose, Sissel Banner
Laugesen, Christian
Ranjan, Ajenthen Gayathri
Nørgaard, Kirsten
author_facet Lundemose, Sissel Banner
Laugesen, Christian
Ranjan, Ajenthen Gayathri
Nørgaard, Kirsten
author_sort Lundemose, Sissel Banner
collection PubMed
description Continuous glucose monitors (CGMs) are valuable tools for improving glycemic control, yet their accuracy might be influenced by physical activity. This study sought to assess the accuracy of the three latest factory-calibrated CGM systems available in Europe at the time the study was conducted, both during aerobic exercise and in typical daily scenarios. The accuracy evaluation, based on metrics such as the median absolute relative difference (MARD) and point and rate error-grid analyses (PEGA and REGA), involved 13 adults with type 1 diabetes. Participants wore all sensors during a 1 h in-clinic exercise session followed by a subsequent 3-day home period, with blood glucose measurements serving as reference values in both contexts. During exercise, no statistically significant differences in MARD were observed (Dexcom G6: 12.6%, Guardian 4: 10.7%, and Freestyle Libre 2: 17.2%; p = 0.31), and similarly, no significant differences emerged in PEGA-zone-AB (100%, 100%, 96.8%; p = 0.37). Nevertheless, Freestyle Libre 2 showed comparatively diminished accuracy in estimating glucose trends during exercise (REGA-zone-AB: 100%, 93.0%, 73.3%; p = 0.0003). In the home environment, Freestyle Libre 2 exhibited a significantly higher MARD when compared to the other systems (10.2%, 11.9%, 16.7%, p = 0.02). Overall, Dexcom G6 and Guardian 4 demonstrated superior accuracy in both exercise and daily life scenarios compared to Freestyle Libre 2.
format Online
Article
Text
id pubmed-10675113
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-106751132023-11-18 Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use Lundemose, Sissel Banner Laugesen, Christian Ranjan, Ajenthen Gayathri Nørgaard, Kirsten Sensors (Basel) Communication Continuous glucose monitors (CGMs) are valuable tools for improving glycemic control, yet their accuracy might be influenced by physical activity. This study sought to assess the accuracy of the three latest factory-calibrated CGM systems available in Europe at the time the study was conducted, both during aerobic exercise and in typical daily scenarios. The accuracy evaluation, based on metrics such as the median absolute relative difference (MARD) and point and rate error-grid analyses (PEGA and REGA), involved 13 adults with type 1 diabetes. Participants wore all sensors during a 1 h in-clinic exercise session followed by a subsequent 3-day home period, with blood glucose measurements serving as reference values in both contexts. During exercise, no statistically significant differences in MARD were observed (Dexcom G6: 12.6%, Guardian 4: 10.7%, and Freestyle Libre 2: 17.2%; p = 0.31), and similarly, no significant differences emerged in PEGA-zone-AB (100%, 100%, 96.8%; p = 0.37). Nevertheless, Freestyle Libre 2 showed comparatively diminished accuracy in estimating glucose trends during exercise (REGA-zone-AB: 100%, 93.0%, 73.3%; p = 0.0003). In the home environment, Freestyle Libre 2 exhibited a significantly higher MARD when compared to the other systems (10.2%, 11.9%, 16.7%, p = 0.02). Overall, Dexcom G6 and Guardian 4 demonstrated superior accuracy in both exercise and daily life scenarios compared to Freestyle Libre 2. MDPI 2023-11-18 /pmc/articles/PMC10675113/ /pubmed/38005642 http://dx.doi.org/10.3390/s23229256 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Communication
Lundemose, Sissel Banner
Laugesen, Christian
Ranjan, Ajenthen Gayathri
Nørgaard, Kirsten
Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use
title Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use
title_full Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use
title_fullStr Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use
title_full_unstemmed Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use
title_short Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use
title_sort factory-calibrated continuous glucose monitoring systems in type 1 diabetes: accuracy during in-clinic exercise and home use
topic Communication
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10675113/
https://www.ncbi.nlm.nih.gov/pubmed/38005642
http://dx.doi.org/10.3390/s23229256
work_keys_str_mv AT lundemosesisselbanner factorycalibratedcontinuousglucosemonitoringsystemsintype1diabetesaccuracyduringinclinicexerciseandhomeuse
AT laugesenchristian factorycalibratedcontinuousglucosemonitoringsystemsintype1diabetesaccuracyduringinclinicexerciseandhomeuse
AT ranjanajenthengayathri factorycalibratedcontinuousglucosemonitoringsystemsintype1diabetesaccuracyduringinclinicexerciseandhomeuse
AT nørgaardkirsten factorycalibratedcontinuousglucosemonitoringsystemsintype1diabetesaccuracyduringinclinicexerciseandhomeuse