Cargando…
648. Validation of a Synthetic Multiplexed External Control for Monitoring the Performance and Detection of Multiple Respiratory Viral and Bacterial Pathogens for a Point of Care Nucleic Acid Testing Panel
BACKGROUND: Advancements in nucleic acid testing has enabled point of care testing solutions to provide clinicians with greatly improved time to diagnosis and patient care. However, complex technologies and regulatory requirements demand close monitoring of assay performance to accurately identify s...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10678055/ http://dx.doi.org/10.1093/ofid/ofad500.711 |
_version_ | 1785150275187638272 |
---|---|
author | Sivaji, Nithya Farrell, Emma Spenlinhauer, Tania R Gordon, Joan Cronin, Monica Graham, Brad Bastar, Jeffrey |
author_facet | Sivaji, Nithya Farrell, Emma Spenlinhauer, Tania R Gordon, Joan Cronin, Monica Graham, Brad Bastar, Jeffrey |
author_sort | Sivaji, Nithya |
collection | PubMed |
description | BACKGROUND: Advancements in nucleic acid testing has enabled point of care testing solutions to provide clinicians with greatly improved time to diagnosis and patient care. However, complex technologies and regulatory requirements demand close monitoring of assay performance to accurately identify shifts, trends, and random errors to ensure accurate patient reporting. To support these requirements, MMQCI has developed a synthetic, easy-to-use, external control that simultaneously monitors the accurate identification of 15 respiratory pathogens (11 viral, 4 bacterial targets) detected with the BIOFIRE(®) SPOTFIRE(®) Respiratory (R) Panel. METHODS: The multiplex control contains genome segments of all pathogens detected by the SPOTFIRE R Panel. Sequences were designed in silico, ligated into MMQCI vectors, and transformed to create stable frozen clone stocks. In vitro RNA transcripts were generated, quantified, and formulated in a proprietary matrix to stabilize and carry the RNA through the entire test process. Validation studies were conducted to demonstrate precision across multiple reagents, operators and sites. Failure-mode, shipping and stability studies were conducted to confirm robustness and reliability. Three unique lots of the SPOTFIRE RSP Positive and Negative controls were manufactured to include variability in key components and tested across 6 different sites, incorporating multiple reagent lots, operators and instruments. A total of 361 runs were performed across the positive and negative controls using the SPOTFIRE R Panel. RESULTS: Of the 361 total runs across the 6 sites, 351 were valid and used for analysis. A total of 171/172 positive controls resulted in correct calls (99.3%), and all 179 negative controls were called correctly (100%). A regression analysis was performed to confirm real-time stability for 12 months when stored at temperatures ranging from 2(o)C to 25(o)C. [Figure: see text] CONCLUSION: Validation studies demonstrated reproducibility, and overall precision and robustness of SPOTFIRE RSP Positive and Negative Controls. The SPOTFIRE RSP Positive and Negative Controls provide a stable, ease to use solution for training, verification and routine monitoring of the performance of the BIOFIRE(®) SPOTFIRE(®) Respiratory Solution. DISCLOSURES: Nithya Sivaji, MS, Maine Molecular Quality Controls Inc.: employer Emma Farrell, n/a, Maine Molecular Quality Controls Inc.: employer Tania R. Spenlinhauer, PhD, Maine Molecular Quality Controls Inc.: employer Joan Gordon, n/a, Maine Molecular Quality Controls, Inc: President|Maine Molecular Quality Controls, Inc: Ownership Interest Monica Cronin, n/a, bioMerieux: Stocks/Bonds Brad Graham, n/a, bioMerieiuux: Employee Jeffrey Bastar, n/a, bioMerieux: Employee|bioMerieux: Stocks/Bonds |
format | Online Article Text |
id | pubmed-10678055 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-106780552023-11-27 648. Validation of a Synthetic Multiplexed External Control for Monitoring the Performance and Detection of Multiple Respiratory Viral and Bacterial Pathogens for a Point of Care Nucleic Acid Testing Panel Sivaji, Nithya Farrell, Emma Spenlinhauer, Tania R Gordon, Joan Cronin, Monica Graham, Brad Bastar, Jeffrey Open Forum Infect Dis Abstract BACKGROUND: Advancements in nucleic acid testing has enabled point of care testing solutions to provide clinicians with greatly improved time to diagnosis and patient care. However, complex technologies and regulatory requirements demand close monitoring of assay performance to accurately identify shifts, trends, and random errors to ensure accurate patient reporting. To support these requirements, MMQCI has developed a synthetic, easy-to-use, external control that simultaneously monitors the accurate identification of 15 respiratory pathogens (11 viral, 4 bacterial targets) detected with the BIOFIRE(®) SPOTFIRE(®) Respiratory (R) Panel. METHODS: The multiplex control contains genome segments of all pathogens detected by the SPOTFIRE R Panel. Sequences were designed in silico, ligated into MMQCI vectors, and transformed to create stable frozen clone stocks. In vitro RNA transcripts were generated, quantified, and formulated in a proprietary matrix to stabilize and carry the RNA through the entire test process. Validation studies were conducted to demonstrate precision across multiple reagents, operators and sites. Failure-mode, shipping and stability studies were conducted to confirm robustness and reliability. Three unique lots of the SPOTFIRE RSP Positive and Negative controls were manufactured to include variability in key components and tested across 6 different sites, incorporating multiple reagent lots, operators and instruments. A total of 361 runs were performed across the positive and negative controls using the SPOTFIRE R Panel. RESULTS: Of the 361 total runs across the 6 sites, 351 were valid and used for analysis. A total of 171/172 positive controls resulted in correct calls (99.3%), and all 179 negative controls were called correctly (100%). A regression analysis was performed to confirm real-time stability for 12 months when stored at temperatures ranging from 2(o)C to 25(o)C. [Figure: see text] CONCLUSION: Validation studies demonstrated reproducibility, and overall precision and robustness of SPOTFIRE RSP Positive and Negative Controls. The SPOTFIRE RSP Positive and Negative Controls provide a stable, ease to use solution for training, verification and routine monitoring of the performance of the BIOFIRE(®) SPOTFIRE(®) Respiratory Solution. DISCLOSURES: Nithya Sivaji, MS, Maine Molecular Quality Controls Inc.: employer Emma Farrell, n/a, Maine Molecular Quality Controls Inc.: employer Tania R. Spenlinhauer, PhD, Maine Molecular Quality Controls Inc.: employer Joan Gordon, n/a, Maine Molecular Quality Controls, Inc: President|Maine Molecular Quality Controls, Inc: Ownership Interest Monica Cronin, n/a, bioMerieux: Stocks/Bonds Brad Graham, n/a, bioMerieiuux: Employee Jeffrey Bastar, n/a, bioMerieux: Employee|bioMerieux: Stocks/Bonds Oxford University Press 2023-11-27 /pmc/articles/PMC10678055/ http://dx.doi.org/10.1093/ofid/ofad500.711 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of Infectious Diseases Society of America. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Abstract Sivaji, Nithya Farrell, Emma Spenlinhauer, Tania R Gordon, Joan Cronin, Monica Graham, Brad Bastar, Jeffrey 648. Validation of a Synthetic Multiplexed External Control for Monitoring the Performance and Detection of Multiple Respiratory Viral and Bacterial Pathogens for a Point of Care Nucleic Acid Testing Panel |
title | 648. Validation of a Synthetic Multiplexed External Control for Monitoring the Performance and Detection of Multiple Respiratory Viral and Bacterial Pathogens for a Point of Care Nucleic Acid Testing Panel |
title_full | 648. Validation of a Synthetic Multiplexed External Control for Monitoring the Performance and Detection of Multiple Respiratory Viral and Bacterial Pathogens for a Point of Care Nucleic Acid Testing Panel |
title_fullStr | 648. Validation of a Synthetic Multiplexed External Control for Monitoring the Performance and Detection of Multiple Respiratory Viral and Bacterial Pathogens for a Point of Care Nucleic Acid Testing Panel |
title_full_unstemmed | 648. Validation of a Synthetic Multiplexed External Control for Monitoring the Performance and Detection of Multiple Respiratory Viral and Bacterial Pathogens for a Point of Care Nucleic Acid Testing Panel |
title_short | 648. Validation of a Synthetic Multiplexed External Control for Monitoring the Performance and Detection of Multiple Respiratory Viral and Bacterial Pathogens for a Point of Care Nucleic Acid Testing Panel |
title_sort | 648. validation of a synthetic multiplexed external control for monitoring the performance and detection of multiple respiratory viral and bacterial pathogens for a point of care nucleic acid testing panel |
topic | Abstract |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10678055/ http://dx.doi.org/10.1093/ofid/ofad500.711 |
work_keys_str_mv | AT sivajinithya 648validationofasyntheticmultiplexedexternalcontrolformonitoringtheperformanceanddetectionofmultiplerespiratoryviralandbacterialpathogensforapointofcarenucleicacidtestingpanel AT farrellemma 648validationofasyntheticmultiplexedexternalcontrolformonitoringtheperformanceanddetectionofmultiplerespiratoryviralandbacterialpathogensforapointofcarenucleicacidtestingpanel AT spenlinhauertaniar 648validationofasyntheticmultiplexedexternalcontrolformonitoringtheperformanceanddetectionofmultiplerespiratoryviralandbacterialpathogensforapointofcarenucleicacidtestingpanel AT gordonjoan 648validationofasyntheticmultiplexedexternalcontrolformonitoringtheperformanceanddetectionofmultiplerespiratoryviralandbacterialpathogensforapointofcarenucleicacidtestingpanel AT croninmonica 648validationofasyntheticmultiplexedexternalcontrolformonitoringtheperformanceanddetectionofmultiplerespiratoryviralandbacterialpathogensforapointofcarenucleicacidtestingpanel AT grahambrad 648validationofasyntheticmultiplexedexternalcontrolformonitoringtheperformanceanddetectionofmultiplerespiratoryviralandbacterialpathogensforapointofcarenucleicacidtestingpanel AT bastarjeffrey 648validationofasyntheticmultiplexedexternalcontrolformonitoringtheperformanceanddetectionofmultiplerespiratoryviralandbacterialpathogensforapointofcarenucleicacidtestingpanel |