The effect of breast shielding outside the field of view on breast entrance surface dose in axial X-ray examinations: a phantom study

PURPOSE: The purpose of this study was to evaluate the effect of outside-field-of-view (FOV) lead shielding on the entrance surface dose (ESD) of the breast on an anthropomorphic X-ray phantom for a variety of axial skeleton X-ray examinations. METHODS: Using an anthropomorphic phantom and radiation...

Descripción completa

Detalles Bibliográficos
Autores principales: Hurley, Lauren, Alashban, Yazeed, Albeshan, Salman, England, Andrew, McEntee, Mark F.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Galenos Publishing 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10679606/
https://www.ncbi.nlm.nih.gov/pubmed/37129301
http://dx.doi.org/10.4274/dir.2023.232126
Descripción
Sumario:PURPOSE: The purpose of this study was to evaluate the effect of outside-field-of-view (FOV) lead shielding on the entrance surface dose (ESD) of the breast on an anthropomorphic X-ray phantom for a variety of axial skeleton X-ray examinations. METHODS: Using an anthropomorphic phantom and radiation dosimeter, the ESD of the breast was measured with and without outside-FOV shielding in anterior-posterior (AP) abdomen, AP cervical spine, occipitomental 30° (OM30) facial bones, AP lumbar spine, and lateral lumbar spine radiography. The effect of several exposure parameters, including a low milliampere-seconds technique, grid use, automatic exposure control use, wraparound lead (WAL) use, trolley use, and X-ray table use, on the ESD of the breast with and without outside-FOV shielding was investigated. The mean ESD (μSv) and standard deviation for each radiographic protocol were calculated. A one-tailed Student’s t-test was carried out to evaluate whether ESD to the breast was reduced with the use of outside-FOV shielding. RESULTS: A total of 920 breast ESD measurements were recorded across the different protocol parameters. The largest decrease in mean ESD of the breast with outside-FOV shielding was 0.002 μSv (P = 0.084), recorded in the AP abdomen on the table with a grid, OM30 on the table with a grid, OM30 standard protocol on the trolley, and OM30 on the trolley with WAL protocols. This decrease was found to be statistically non-significant. CONCLUSION: This study found no significant decrease in the ESD of the breast with the use of outside-FOV shielding for the AP abdomen, AP cervical spine, OM30 facial bones, AP lumbar spine, or lateral lumbar spine radiography across a range of protocols.