Cargando…
Rapid atrial pacing above the maximum sensor rate: a case report
BACKGROUND: While ventricular-based timing modes are known to cause elevated atrial pacing above the lower rate when intrinsic atrioventricular (AV) conduction is shorter than programmed AV delay, there is one case report in 2015 by Jafri et al. where rapid atrial pacing was induced in an Abbott dev...
Autores principales: | Bodziock, George M, Kozak, Patrick M, Pruitt, Carrie, Dillon, John F, Bhave, Prashant D |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10686531/ https://www.ncbi.nlm.nih.gov/pubmed/38034939 http://dx.doi.org/10.1093/ehjcr/ytad586 |
Ejemplares similares
-
Arrhythmia Following Pulse Generator Change: What Is the Mechanism?
por: Kozak, Patrick M., et al.
Publicado: (2022) -
Left Atrial Appendage Occlusion Device Buckling: An Avoidable Deployment Complication Identified and Corrected Using Intraprocedural Transesophageal Echocardiography
por: Pruitt, Zachary, et al.
Publicado: (2022) -
Sensors for rate responsive pacing
por: Dell'Orto, Simonetta, et al.
Publicado: (2004) -
Prevention and Treatment of Atrioesophageal Fistula Related to Catheter Ablation for Atrial Fibrillation
por: Bodziock, George M., et al.
Publicado: (2019) -
Rapid Atrial Pacing Promotes Atrial Fibrillation Substrate in Unanesthetized Instrumented Rats
por: Mulla, Wesam, et al.
Publicado: (2019)