A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design
BACKGROUND: Although a new generation of tomographs with a longer axial field-of-view called total-body PET have been developed, they are not widely utilized due to their high cost compared to conventional scanners. The newly designed walk-through total-body PET scanner is introduced as a high-throu...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10689648/ https://www.ncbi.nlm.nih.gov/pubmed/38036794 http://dx.doi.org/10.1186/s40658-023-00593-0 |
_version_ | 1785152402610978816 |
---|---|
author | Dadgar, Meysam Maebe, Jens Abi Akl, Maya Vervenne, Boris Vandenberghe, Stefaan |
author_facet | Dadgar, Meysam Maebe, Jens Abi Akl, Maya Vervenne, Boris Vandenberghe, Stefaan |
author_sort | Dadgar, Meysam |
collection | PubMed |
description | BACKGROUND: Although a new generation of tomographs with a longer axial field-of-view called total-body PET have been developed, they are not widely utilized due to their high cost compared to conventional scanners. The newly designed walk-through total-body PET scanner is introduced as a high-throughput and cost-efficient alternative to total-body PET scanners, by making use of a flat panel geometry and lower cost, depth-of-interaction capable, monolithic BGO detectors. The main aim of the presented study is to evaluate through Monte Carlo simulation the system characteristics of the walk-through total-body PET scanner by comparing it with a Quadra-like total-body PET of similar attributes to the Siemens Biograph Vision Quadra. METHODS: The walk-through total-body PET is comprised of two flat detector panels, spaced 50 cm apart. Each panel, 70 [Formula: see text] 106 cm[Formula: see text] in size, consists of 280 BGO-based monolithic detectors. The Quadra-like TB-PET has been simulated based on the characteristics of the Biograph Vision Quadra, one of the most common total-body PET scanners with 106 cm of axial field-of-view, which is constructed with pixelated LSO scintillation crystals. The spatial resolution, sensitivity, count rate performance, scatter fractions, and image quality of both scanners are simulated in the GATE simulation toolkit for comparison. RESULTS: Due to the DOI-capable detectors used in the walk-through total-body PET, the values of the spatial resolution of this scanner were all below 2 mm along directions parallel to the panels, and reached a maximum of 3.36 mm in the direction perpendicular to the panels. This resolution is a large improvement compared to the values of the Quadra-like TB-PET. The walk-through total-body PET uses its maximum sensitivity (154 cps/kBq) for data acquisition and image reconstruction. CONCLUSION: Based on the combination of very good spatial resolution and high sensitivity of the walk-through total-body PET, along with a 2.2 times lower scintillation crystal volume and 1.8 times lower SiPM surface, this scanner can be a very cost-efficient alternative for total-body PET scanners in cases where concomitant CT is not required. |
format | Online Article Text |
id | pubmed-10689648 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-106896482023-12-02 A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design Dadgar, Meysam Maebe, Jens Abi Akl, Maya Vervenne, Boris Vandenberghe, Stefaan EJNMMI Phys Original Research BACKGROUND: Although a new generation of tomographs with a longer axial field-of-view called total-body PET have been developed, they are not widely utilized due to their high cost compared to conventional scanners. The newly designed walk-through total-body PET scanner is introduced as a high-throughput and cost-efficient alternative to total-body PET scanners, by making use of a flat panel geometry and lower cost, depth-of-interaction capable, monolithic BGO detectors. The main aim of the presented study is to evaluate through Monte Carlo simulation the system characteristics of the walk-through total-body PET scanner by comparing it with a Quadra-like total-body PET of similar attributes to the Siemens Biograph Vision Quadra. METHODS: The walk-through total-body PET is comprised of two flat detector panels, spaced 50 cm apart. Each panel, 70 [Formula: see text] 106 cm[Formula: see text] in size, consists of 280 BGO-based monolithic detectors. The Quadra-like TB-PET has been simulated based on the characteristics of the Biograph Vision Quadra, one of the most common total-body PET scanners with 106 cm of axial field-of-view, which is constructed with pixelated LSO scintillation crystals. The spatial resolution, sensitivity, count rate performance, scatter fractions, and image quality of both scanners are simulated in the GATE simulation toolkit for comparison. RESULTS: Due to the DOI-capable detectors used in the walk-through total-body PET, the values of the spatial resolution of this scanner were all below 2 mm along directions parallel to the panels, and reached a maximum of 3.36 mm in the direction perpendicular to the panels. This resolution is a large improvement compared to the values of the Quadra-like TB-PET. The walk-through total-body PET uses its maximum sensitivity (154 cps/kBq) for data acquisition and image reconstruction. CONCLUSION: Based on the combination of very good spatial resolution and high sensitivity of the walk-through total-body PET, along with a 2.2 times lower scintillation crystal volume and 1.8 times lower SiPM surface, this scanner can be a very cost-efficient alternative for total-body PET scanners in cases where concomitant CT is not required. Springer International Publishing 2023-12-01 /pmc/articles/PMC10689648/ /pubmed/38036794 http://dx.doi.org/10.1186/s40658-023-00593-0 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Original Research Dadgar, Meysam Maebe, Jens Abi Akl, Maya Vervenne, Boris Vandenberghe, Stefaan A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design |
title | A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design |
title_full | A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design |
title_fullStr | A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design |
title_full_unstemmed | A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design |
title_short | A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design |
title_sort | simulation study of the system characteristics for a long axial fov pet design based on monolithic bgo flat panels compared with a pixelated lso cylindrical design |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10689648/ https://www.ncbi.nlm.nih.gov/pubmed/38036794 http://dx.doi.org/10.1186/s40658-023-00593-0 |
work_keys_str_mv | AT dadgarmeysam asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT maebejens asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT abiaklmaya asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT vervenneboris asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT vandenberghestefaan asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT dadgarmeysam simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT maebejens simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT abiaklmaya simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT vervenneboris simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign AT vandenberghestefaan simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign |