A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design

BACKGROUND: Although a new generation of tomographs with a longer axial field-of-view called total-body PET have been developed, they are not widely utilized due to their high cost compared to conventional scanners. The newly designed walk-through total-body PET scanner is introduced as a high-throu...

Descripción completa

Detalles Bibliográficos
Autores principales: Dadgar, Meysam, Maebe, Jens, Abi Akl, Maya, Vervenne, Boris, Vandenberghe, Stefaan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10689648/
https://www.ncbi.nlm.nih.gov/pubmed/38036794
http://dx.doi.org/10.1186/s40658-023-00593-0
_version_ 1785152402610978816
author Dadgar, Meysam
Maebe, Jens
Abi Akl, Maya
Vervenne, Boris
Vandenberghe, Stefaan
author_facet Dadgar, Meysam
Maebe, Jens
Abi Akl, Maya
Vervenne, Boris
Vandenberghe, Stefaan
author_sort Dadgar, Meysam
collection PubMed
description BACKGROUND: Although a new generation of tomographs with a longer axial field-of-view called total-body PET have been developed, they are not widely utilized due to their high cost compared to conventional scanners. The newly designed walk-through total-body PET scanner is introduced as a high-throughput and cost-efficient alternative to total-body PET scanners, by making use of a flat panel geometry and lower cost, depth-of-interaction capable, monolithic BGO detectors. The main aim of the presented study is to evaluate through Monte Carlo simulation the system characteristics of the walk-through total-body PET scanner by comparing it with a Quadra-like total-body PET of similar attributes to the Siemens Biograph Vision Quadra. METHODS: The walk-through total-body PET is comprised of two flat detector panels, spaced 50 cm apart. Each panel, 70 [Formula: see text] 106 cm[Formula: see text] in size, consists of 280 BGO-based monolithic detectors. The Quadra-like TB-PET has been simulated based on the characteristics of the Biograph Vision Quadra, one of the most common total-body PET scanners with 106 cm of axial field-of-view, which is constructed with pixelated LSO scintillation crystals. The spatial resolution, sensitivity, count rate performance, scatter fractions, and image quality of both scanners are simulated in the GATE simulation toolkit for comparison. RESULTS: Due to the DOI-capable detectors used in the walk-through total-body PET, the values of the spatial resolution of this scanner were all below 2 mm along directions parallel to the panels, and reached a maximum of 3.36 mm in the direction perpendicular to the panels. This resolution is a large improvement compared to the values of the Quadra-like TB-PET. The walk-through total-body PET uses its maximum sensitivity (154 cps/kBq) for data acquisition and image reconstruction. CONCLUSION: Based on the combination of very good spatial resolution and high sensitivity of the walk-through total-body PET, along with a 2.2 times lower scintillation crystal volume and 1.8 times lower SiPM surface, this scanner can be a very cost-efficient alternative for total-body PET scanners in cases where concomitant CT is not required.
format Online
Article
Text
id pubmed-10689648
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-106896482023-12-02 A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design Dadgar, Meysam Maebe, Jens Abi Akl, Maya Vervenne, Boris Vandenberghe, Stefaan EJNMMI Phys Original Research BACKGROUND: Although a new generation of tomographs with a longer axial field-of-view called total-body PET have been developed, they are not widely utilized due to their high cost compared to conventional scanners. The newly designed walk-through total-body PET scanner is introduced as a high-throughput and cost-efficient alternative to total-body PET scanners, by making use of a flat panel geometry and lower cost, depth-of-interaction capable, monolithic BGO detectors. The main aim of the presented study is to evaluate through Monte Carlo simulation the system characteristics of the walk-through total-body PET scanner by comparing it with a Quadra-like total-body PET of similar attributes to the Siemens Biograph Vision Quadra. METHODS: The walk-through total-body PET is comprised of two flat detector panels, spaced 50 cm apart. Each panel, 70 [Formula: see text] 106 cm[Formula: see text] in size, consists of 280 BGO-based monolithic detectors. The Quadra-like TB-PET has been simulated based on the characteristics of the Biograph Vision Quadra, one of the most common total-body PET scanners with 106 cm of axial field-of-view, which is constructed with pixelated LSO scintillation crystals. The spatial resolution, sensitivity, count rate performance, scatter fractions, and image quality of both scanners are simulated in the GATE simulation toolkit for comparison. RESULTS: Due to the DOI-capable detectors used in the walk-through total-body PET, the values of the spatial resolution of this scanner were all below 2 mm along directions parallel to the panels, and reached a maximum of 3.36 mm in the direction perpendicular to the panels. This resolution is a large improvement compared to the values of the Quadra-like TB-PET. The walk-through total-body PET uses its maximum sensitivity (154 cps/kBq) for data acquisition and image reconstruction. CONCLUSION: Based on the combination of very good spatial resolution and high sensitivity of the walk-through total-body PET, along with a 2.2 times lower scintillation crystal volume and 1.8 times lower SiPM surface, this scanner can be a very cost-efficient alternative for total-body PET scanners in cases where concomitant CT is not required. Springer International Publishing 2023-12-01 /pmc/articles/PMC10689648/ /pubmed/38036794 http://dx.doi.org/10.1186/s40658-023-00593-0 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Research
Dadgar, Meysam
Maebe, Jens
Abi Akl, Maya
Vervenne, Boris
Vandenberghe, Stefaan
A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design
title A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design
title_full A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design
title_fullStr A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design
title_full_unstemmed A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design
title_short A simulation study of the system characteristics for a long axial FOV PET design based on monolithic BGO flat panels compared with a pixelated LSO cylindrical design
title_sort simulation study of the system characteristics for a long axial fov pet design based on monolithic bgo flat panels compared with a pixelated lso cylindrical design
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10689648/
https://www.ncbi.nlm.nih.gov/pubmed/38036794
http://dx.doi.org/10.1186/s40658-023-00593-0
work_keys_str_mv AT dadgarmeysam asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT maebejens asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT abiaklmaya asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT vervenneboris asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT vandenberghestefaan asimulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT dadgarmeysam simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT maebejens simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT abiaklmaya simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT vervenneboris simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign
AT vandenberghestefaan simulationstudyofthesystemcharacteristicsforalongaxialfovpetdesignbasedonmonolithicbgoflatpanelscomparedwithapixelatedlsocylindricaldesign