Additional Plate Fixation of Hinge Fractures After Varisation Distal Femoral Osteotomies Provides Favorable Torsional Stability: A Biomechanical Study

BACKGROUND: Hinge fractures are considered risk factors for delayed or nonunion of the osteotomy gap in distal femoral osteotomies (DFOs). Limited evidence exists regarding the treatment of hinge fractures after DFO, which could improve stability and thus bone healing. PURPOSE: To (1) examine the ef...

Descripción completa

Detalles Bibliográficos
Autores principales: Peez, Christian, Grosse-Allermann, Arian, Deichsel, Adrian, Raschke, Michael J., Glasbrenner, Johannes, Briese, Thorben, Wermers, Jens, Herbst, Elmar, Kittl, Christoph
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10691291/
https://www.ncbi.nlm.nih.gov/pubmed/37936394
http://dx.doi.org/10.1177/03635465231206947
_version_ 1785152708699750400
author Peez, Christian
Grosse-Allermann, Arian
Deichsel, Adrian
Raschke, Michael J.
Glasbrenner, Johannes
Briese, Thorben
Wermers, Jens
Herbst, Elmar
Kittl, Christoph
author_facet Peez, Christian
Grosse-Allermann, Arian
Deichsel, Adrian
Raschke, Michael J.
Glasbrenner, Johannes
Briese, Thorben
Wermers, Jens
Herbst, Elmar
Kittl, Christoph
author_sort Peez, Christian
collection PubMed
description BACKGROUND: Hinge fractures are considered risk factors for delayed or nonunion of the osteotomy gap in distal femoral osteotomies (DFOs). Limited evidence exists regarding the treatment of hinge fractures after DFO, which could improve stability and thus bone healing. PURPOSE: To (1) examine the effect of hinge fractures on the biomechanical properties of the bone-implant construct, (2) evaluate the biomechanical advantages of an additional fixation of a hinge fracture, and (3) test the biomechanical properties of different types of varisation DFOs. STUDY DESIGN: Controlled laboratory study. METHODS: A total of 32 fresh-frozen human distal femora equally underwent medial closing wedge DFO or lateral opening wedge DFO using a unilateral locking compression plate. The following conditions were serially tested: (1) preserved hinge; (2) hinge fracture along the osteotomy plane; (3) screw fixation of the hinge fracture; and (4) locking T-plate fixation of the hinge fracture. Using a servo-hydraulic materials testing machine, we subjected each construct to 15 cycles of axial compression (400 N; 20 N/s) and internal and external rotational loads (10 N·m; 0.5 N·m/s) to evaluate the stiffness. The axial and torsional hinge displacement was recorded using a 3-dimensional optical measuring system. Repeated-measures 1-way analysis of variance and post hoc Bonferroni correction were used for multiple comparisons. Statistical significance was set at P < .05. RESULTS: Independent from the type of osteotomy, a fractured hinge significantly (P < .001) increased rotational displacement and reduced stiffness of the bone-implant construct, resulting in ≥1.92 mm increased displacement and ≥70% reduced stiffness in each rotational direction, while the axial stiffness remained unchanged. For both procedures, neither a screw nor a plate could restore intact rotational stiffness (P < .01), while only the plate was able to restore intact rotational displacement. However, the plate always performed better compared with the screw, with significantly higher and lower values for stiffness (+38% to +53%; P < .05) and displacement (–55% to −72%; P < .01), respectively, in ≥1 rotational direction. At the same time, the type of osteotomy did not significantly affect axial and torsional stability. CONCLUSION: Hinge fractures after medial closing wedge DFO and lateral opening wedge DFO caused decreased bone-implant construct rotational stiffness and increased fracture-site displacement. In contrast, the axial stiffness remained unchanged in the cadaveric model. CLINICAL RELEVANCE: When considering an osteosynthesis of a hinge fracture in a DFO, an additional plate fixation was the construct with the highest stiffness and least displacement, which could restore intact hinge rotational displacement.
format Online
Article
Text
id pubmed-10691291
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-106912912023-12-02 Additional Plate Fixation of Hinge Fractures After Varisation Distal Femoral Osteotomies Provides Favorable Torsional Stability: A Biomechanical Study Peez, Christian Grosse-Allermann, Arian Deichsel, Adrian Raschke, Michael J. Glasbrenner, Johannes Briese, Thorben Wermers, Jens Herbst, Elmar Kittl, Christoph Am J Sports Med Articles BACKGROUND: Hinge fractures are considered risk factors for delayed or nonunion of the osteotomy gap in distal femoral osteotomies (DFOs). Limited evidence exists regarding the treatment of hinge fractures after DFO, which could improve stability and thus bone healing. PURPOSE: To (1) examine the effect of hinge fractures on the biomechanical properties of the bone-implant construct, (2) evaluate the biomechanical advantages of an additional fixation of a hinge fracture, and (3) test the biomechanical properties of different types of varisation DFOs. STUDY DESIGN: Controlled laboratory study. METHODS: A total of 32 fresh-frozen human distal femora equally underwent medial closing wedge DFO or lateral opening wedge DFO using a unilateral locking compression plate. The following conditions were serially tested: (1) preserved hinge; (2) hinge fracture along the osteotomy plane; (3) screw fixation of the hinge fracture; and (4) locking T-plate fixation of the hinge fracture. Using a servo-hydraulic materials testing machine, we subjected each construct to 15 cycles of axial compression (400 N; 20 N/s) and internal and external rotational loads (10 N·m; 0.5 N·m/s) to evaluate the stiffness. The axial and torsional hinge displacement was recorded using a 3-dimensional optical measuring system. Repeated-measures 1-way analysis of variance and post hoc Bonferroni correction were used for multiple comparisons. Statistical significance was set at P < .05. RESULTS: Independent from the type of osteotomy, a fractured hinge significantly (P < .001) increased rotational displacement and reduced stiffness of the bone-implant construct, resulting in ≥1.92 mm increased displacement and ≥70% reduced stiffness in each rotational direction, while the axial stiffness remained unchanged. For both procedures, neither a screw nor a plate could restore intact rotational stiffness (P < .01), while only the plate was able to restore intact rotational displacement. However, the plate always performed better compared with the screw, with significantly higher and lower values for stiffness (+38% to +53%; P < .05) and displacement (–55% to −72%; P < .01), respectively, in ≥1 rotational direction. At the same time, the type of osteotomy did not significantly affect axial and torsional stability. CONCLUSION: Hinge fractures after medial closing wedge DFO and lateral opening wedge DFO caused decreased bone-implant construct rotational stiffness and increased fracture-site displacement. In contrast, the axial stiffness remained unchanged in the cadaveric model. CLINICAL RELEVANCE: When considering an osteosynthesis of a hinge fracture in a DFO, an additional plate fixation was the construct with the highest stiffness and least displacement, which could restore intact hinge rotational displacement. SAGE Publications 2023-11-07 2023-12 /pmc/articles/PMC10691291/ /pubmed/37936394 http://dx.doi.org/10.1177/03635465231206947 Text en © 2023 The Author(s) https://creativecommons.org/licenses/by/4.0/This article is distributed under the terms of the Creative Commons Attribution 4.0 License (https://creativecommons.org/licenses/by/4.0/) which permits any use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Articles
Peez, Christian
Grosse-Allermann, Arian
Deichsel, Adrian
Raschke, Michael J.
Glasbrenner, Johannes
Briese, Thorben
Wermers, Jens
Herbst, Elmar
Kittl, Christoph
Additional Plate Fixation of Hinge Fractures After Varisation Distal Femoral Osteotomies Provides Favorable Torsional Stability: A Biomechanical Study
title Additional Plate Fixation of Hinge Fractures After Varisation Distal Femoral Osteotomies Provides Favorable Torsional Stability: A Biomechanical Study
title_full Additional Plate Fixation of Hinge Fractures After Varisation Distal Femoral Osteotomies Provides Favorable Torsional Stability: A Biomechanical Study
title_fullStr Additional Plate Fixation of Hinge Fractures After Varisation Distal Femoral Osteotomies Provides Favorable Torsional Stability: A Biomechanical Study
title_full_unstemmed Additional Plate Fixation of Hinge Fractures After Varisation Distal Femoral Osteotomies Provides Favorable Torsional Stability: A Biomechanical Study
title_short Additional Plate Fixation of Hinge Fractures After Varisation Distal Femoral Osteotomies Provides Favorable Torsional Stability: A Biomechanical Study
title_sort additional plate fixation of hinge fractures after varisation distal femoral osteotomies provides favorable torsional stability: a biomechanical study
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10691291/
https://www.ncbi.nlm.nih.gov/pubmed/37936394
http://dx.doi.org/10.1177/03635465231206947
work_keys_str_mv AT peezchristian additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy
AT grosseallermannarian additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy
AT deichseladrian additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy
AT raschkemichaelj additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy
AT glasbrennerjohannes additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy
AT briesethorben additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy
AT wermersjens additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy
AT herbstelmar additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy
AT kittlchristoph additionalplatefixationofhingefracturesaftervarisationdistalfemoralosteotomiesprovidesfavorabletorsionalstabilityabiomechanicalstudy