Cargando…

Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software

This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combination of a...

Descripción completa

Detalles Bibliográficos
Autores principales: Both, Joseph A., Pawlicki, Todd
Formato: Texto
Lenguaje:English
Publicado: Ivyspring International Publisher 2004
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1074715/
https://www.ncbi.nlm.nih.gov/pubmed/15912199
_version_ 1782123404032212992
author Both, Joseph A.
Pawlicki, Todd
author_facet Both, Joseph A.
Pawlicki, Todd
author_sort Both, Joseph A.
collection PubMed
description This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combination of analytic and Monte Carlo methods generates beam models from dosimetric data collected in a water phantom. This study uses NXEGS to commission 6, 9, and 12 MeV electron beams of a Varian Clinac 2100C using three applicators with standard inserts. Central axis depth-dose, primary axis and diagonal beam profiles, and output factors are the measurements necessary for commissioning of the code. We present a comparison of measured dose distributions with the distributions generated by NXEGS, using confidence limits on seven measures of error. We find that confidence limits are typically less than 3% or 3 mm, but increase with increasing source to surface distance (SSD) and depth at or beyond R(50). We also investigate the dependence of NXEGS' performance on the size and composition of data used to commission the program, finding a weak dependence on number of dose profiles in the data set, but finding also that commissioning data need be measured at only two SSDs.
format Text
id pubmed-1074715
institution National Center for Biotechnology Information
language English
publishDate 2004
publisher Ivyspring International Publisher
record_format MEDLINE/PubMed
spelling pubmed-10747152005-05-18 Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software Both, Joseph A. Pawlicki, Todd Int J Med Sci Research Paper This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combination of analytic and Monte Carlo methods generates beam models from dosimetric data collected in a water phantom. This study uses NXEGS to commission 6, 9, and 12 MeV electron beams of a Varian Clinac 2100C using three applicators with standard inserts. Central axis depth-dose, primary axis and diagonal beam profiles, and output factors are the measurements necessary for commissioning of the code. We present a comparison of measured dose distributions with the distributions generated by NXEGS, using confidence limits on seven measures of error. We find that confidence limits are typically less than 3% or 3 mm, but increase with increasing source to surface distance (SSD) and depth at or beyond R(50). We also investigate the dependence of NXEGS' performance on the size and composition of data used to commission the program, finding a weak dependence on number of dose profiles in the data set, but finding also that commissioning data need be measured at only two SSDs. Ivyspring International Publisher 2004-06-01 /pmc/articles/PMC1074715/ /pubmed/15912199 Text en © Ivyspring International Publisher. This is an open access article licensed under a Creative Commons License (http://creativecommons.org/licenses/by-nc-nd/2.0/) which permits the distribution or copying for non-commercial purposes, provided that the article is in whole, unmodified, and properly cited.
spellingShingle Research Paper
Both, Joseph A.
Pawlicki, Todd
Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_full Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_fullStr Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_full_unstemmed Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_short Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_sort monte carlo commissioning of low energy electron radiotherapy beams using nxegs software
topic Research Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1074715/
https://www.ncbi.nlm.nih.gov/pubmed/15912199
work_keys_str_mv AT bothjosepha montecarlocommissioningoflowenergyelectronradiotherapybeamsusingnxegssoftware
AT pawlickitodd montecarlocommissioningoflowenergyelectronradiotherapybeamsusingnxegssoftware