Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening

INTRODUCTION: The success of vascular stents in the restoration of blood flow is limited by restenosis. Recent data generated from computational fluid dynamics (CFD) models suggest that the vascular geometry created by an implanted stent causes local alterations in wall shear stress (WSS) that are a...

Descripción completa

Detalles Bibliográficos
Autores principales: LaDisa, John F, Olson, Lars E, Hettrick, Douglas A, Warltier, David C, Kersten, Judy R, Pagel, Paul S
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2005
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1276824/
https://www.ncbi.nlm.nih.gov/pubmed/16250918
http://dx.doi.org/10.1186/1475-925X-4-59
_version_ 1782126016247889920
author LaDisa, John F
Olson, Lars E
Hettrick, Douglas A
Warltier, David C
Kersten, Judy R
Pagel, Paul S
author_facet LaDisa, John F
Olson, Lars E
Hettrick, Douglas A
Warltier, David C
Kersten, Judy R
Pagel, Paul S
author_sort LaDisa, John F
collection PubMed
description INTRODUCTION: The success of vascular stents in the restoration of blood flow is limited by restenosis. Recent data generated from computational fluid dynamics (CFD) models suggest that the vascular geometry created by an implanted stent causes local alterations in wall shear stress (WSS) that are associated with neointimal hyperplasia (NH). Foreshortening is a potential limitation of stent design that may affect stent performance and the rate of restenosis. The angle created between axially aligned stent struts and the principal direction of blood flow varies with the degree to which the stent foreshortens after implantation. METHODS: In the current investigation, we tested the hypothesis that stent foreshortening adversely influences the distribution of WSS and WSS gradients using time-dependent 3D CFD simulations of normal arteries based on canine coronary artery measurements of diameter and blood flow. WSS and WSS gradients were calculated using conventional techniques in ideal (16 mm) and progressively foreshortened (14 and 12 mm) stented computational vessels. RESULTS: Stent foreshortening increased the intrastrut area of the luminal surface exposed to low WSS and elevated spatial WSS gradients. Progressive degrees of stent foreshortening were also associated with strut misalignment relative to the direction of blood flow as indicated by analysis of near-wall velocity vectors. CONCLUSION: The current results suggest that foreshortening may predispose the stented vessel to a higher risk of neointimal hyperplasia.
format Text
id pubmed-1276824
institution National Center for Biotechnology Information
language English
publishDate 2005
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-12768242005-11-04 Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening LaDisa, John F Olson, Lars E Hettrick, Douglas A Warltier, David C Kersten, Judy R Pagel, Paul S Biomed Eng Online Research INTRODUCTION: The success of vascular stents in the restoration of blood flow is limited by restenosis. Recent data generated from computational fluid dynamics (CFD) models suggest that the vascular geometry created by an implanted stent causes local alterations in wall shear stress (WSS) that are associated with neointimal hyperplasia (NH). Foreshortening is a potential limitation of stent design that may affect stent performance and the rate of restenosis. The angle created between axially aligned stent struts and the principal direction of blood flow varies with the degree to which the stent foreshortens after implantation. METHODS: In the current investigation, we tested the hypothesis that stent foreshortening adversely influences the distribution of WSS and WSS gradients using time-dependent 3D CFD simulations of normal arteries based on canine coronary artery measurements of diameter and blood flow. WSS and WSS gradients were calculated using conventional techniques in ideal (16 mm) and progressively foreshortened (14 and 12 mm) stented computational vessels. RESULTS: Stent foreshortening increased the intrastrut area of the luminal surface exposed to low WSS and elevated spatial WSS gradients. Progressive degrees of stent foreshortening were also associated with strut misalignment relative to the direction of blood flow as indicated by analysis of near-wall velocity vectors. CONCLUSION: The current results suggest that foreshortening may predispose the stented vessel to a higher risk of neointimal hyperplasia. BioMed Central 2005-10-26 /pmc/articles/PMC1276824/ /pubmed/16250918 http://dx.doi.org/10.1186/1475-925X-4-59 Text en Copyright © 2005 LaDisa et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
LaDisa, John F
Olson, Lars E
Hettrick, Douglas A
Warltier, David C
Kersten, Judy R
Pagel, Paul S
Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening
title Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening
title_full Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening
title_fullStr Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening
title_full_unstemmed Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening
title_short Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening
title_sort axial stent strut angle influences wall shear stress after stent implantation: analysis using 3d computational fluid dynamics models of stent foreshortening
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1276824/
https://www.ncbi.nlm.nih.gov/pubmed/16250918
http://dx.doi.org/10.1186/1475-925X-4-59
work_keys_str_mv AT ladisajohnf axialstentstrutangleinfluenceswallshearstressafterstentimplantationanalysisusing3dcomputationalfluiddynamicsmodelsofstentforeshortening
AT olsonlarse axialstentstrutangleinfluenceswallshearstressafterstentimplantationanalysisusing3dcomputationalfluiddynamicsmodelsofstentforeshortening
AT hettrickdouglasa axialstentstrutangleinfluenceswallshearstressafterstentimplantationanalysisusing3dcomputationalfluiddynamicsmodelsofstentforeshortening
AT warltierdavidc axialstentstrutangleinfluenceswallshearstressafterstentimplantationanalysisusing3dcomputationalfluiddynamicsmodelsofstentforeshortening
AT kerstenjudyr axialstentstrutangleinfluenceswallshearstressafterstentimplantationanalysisusing3dcomputationalfluiddynamicsmodelsofstentforeshortening
AT pagelpauls axialstentstrutangleinfluenceswallshearstressafterstentimplantationanalysisusing3dcomputationalfluiddynamicsmodelsofstentforeshortening