Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots

BACKGROUND: Active magnetic resonance imaging implants, for example stents, stent grafts or vena cava filters, are constructed as wireless inductively coupled transmit and receive coils. They are built as a resonator tuned to the Larmor frequency of a magnetic resonance system. The resonator can be...

Descripción completa

Detalles Bibliográficos
Autores principales: Busch, Martin HJ, Vollmann, Wolfgang, Grönemeyer, Dietrich HW
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1513583/
https://www.ncbi.nlm.nih.gov/pubmed/16729878
http://dx.doi.org/10.1186/1475-925X-5-35
_version_ 1782128508500180992
author Busch, Martin HJ
Vollmann, Wolfgang
Grönemeyer, Dietrich HW
author_facet Busch, Martin HJ
Vollmann, Wolfgang
Grönemeyer, Dietrich HW
author_sort Busch, Martin HJ
collection PubMed
description BACKGROUND: Active magnetic resonance imaging implants, for example stents, stent grafts or vena cava filters, are constructed as wireless inductively coupled transmit and receive coils. They are built as a resonator tuned to the Larmor frequency of a magnetic resonance system. The resonator can be added to or incorporated within the implant. This technology can counteract the shielding caused by eddy currents inside the metallic implant structure. This may allow getting diagnostic information of the implant lumen (in stent stenosis or thrombosis for example). The electro magnetic rf-pulses during magnetic resonance imaging induce a current in the circuit path of the resonator. A by material fatigue provoked partial rupture of the circuit path or a broken wire with touching surfaces can set up a relatively high resistance on a very short distance, which may behave as a point-like power source, a hot spot, inside the body part the resonator is implanted to. This local power loss inside a small volume can reach ¼ of the total power loss of the intact resonating circuit, which itself is proportional to the product of the resonator volume and the quality factor and depends as well from the orientation of the resonator with respect to the main magnetic field and the imaging sequence the resonator is exposed to. METHODS: First an analytical solution of a hot spot for thermal equilibrium is described. This analytical solution with a definite hot spot power loss represents the worst case scenario for thermal equilibrium inside a homogeneous medium without cooling effects. Starting with this worst case assumptions additional conditions are considered in a numerical simulation, which are more realistic and may make the results less critical. The analytical solution as well as the numerical simulations use the experimental experience of the maximum hot spot power loss of implanted resonators with a definite volume during magnetic resonance imaging investigations. The finite volume analysis calculates the time developing temperature maps for the model of a broken linear metallic wire embedded in tissue. Half of the total hot spot power loss is assumed to diffuse into both wire parts at the location of a defect. The energy is distributed from there by heat conduction. Additionally the effect of blood perfusion and blood flow is respected in some simulations because the simultaneous appearance of all worst case conditions, especially the absence of blood perfusion and blood flow near the hot spot, is very unlikely for vessel implants. RESULTS: The analytical solution as worst case scenario as well as the finite volume analysis for near worst case situations show not negligible volumes with critical temperature increases for part of the modeled hot spot situations. MR investigations with a high rf-pulse density lasting below a minute can establish volumes of several cubic millimeters with temperature increases high enough to start cell destruction. Longer exposure times can involve volumes larger than 100 mm(3). Even temperature increases in the range of thermal ablation are reached for substantial volumes. MR sequence exposure time and hot spot power loss are the primary factors influencing the volume with critical temperature increases. Wire radius, wire material as well as the physiological parameters blood perfusion and blood flow inside larger vessels reduce the volume with critical temperature increases, but do not exclude a volume with critical tissue heating for resonators with a large product of resonator volume and quality factor. CONCLUSION: The worst case scenario assumes thermal equilibrium for a hot spot embedded in homogeneous tissue without any cooling due to blood perfusion or flow. The finite volume analysis can calculate the results for near and not close to worst case conditions. For both cases a substantial volume can reach a critical temperature increase in a short time. The analytical solution, as absolute worst case, points out that resonators with a small product of inductance volume and quality factor (Q V(ind )< 2 cm(3)) are definitely save. Stents for coronary vessels or resonators used as tracking devices for interventional procedures therefore have no risk of high temperature increases. The finite volume analysis shows for sure that also conditions not close to the worst case reach physiologically critical temperature increases for implants with a large product of inductance volume and quality factor (Q V(ind )> 10 cm(3)). Such resonators exclude patients from exactly the MRI investigation these devices are made for.
format Text
id pubmed-1513583
institution National Center for Biotechnology Information
language English
publishDate 2006
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-15135832006-07-22 Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots Busch, Martin HJ Vollmann, Wolfgang Grönemeyer, Dietrich HW Biomed Eng Online Research BACKGROUND: Active magnetic resonance imaging implants, for example stents, stent grafts or vena cava filters, are constructed as wireless inductively coupled transmit and receive coils. They are built as a resonator tuned to the Larmor frequency of a magnetic resonance system. The resonator can be added to or incorporated within the implant. This technology can counteract the shielding caused by eddy currents inside the metallic implant structure. This may allow getting diagnostic information of the implant lumen (in stent stenosis or thrombosis for example). The electro magnetic rf-pulses during magnetic resonance imaging induce a current in the circuit path of the resonator. A by material fatigue provoked partial rupture of the circuit path or a broken wire with touching surfaces can set up a relatively high resistance on a very short distance, which may behave as a point-like power source, a hot spot, inside the body part the resonator is implanted to. This local power loss inside a small volume can reach ¼ of the total power loss of the intact resonating circuit, which itself is proportional to the product of the resonator volume and the quality factor and depends as well from the orientation of the resonator with respect to the main magnetic field and the imaging sequence the resonator is exposed to. METHODS: First an analytical solution of a hot spot for thermal equilibrium is described. This analytical solution with a definite hot spot power loss represents the worst case scenario for thermal equilibrium inside a homogeneous medium without cooling effects. Starting with this worst case assumptions additional conditions are considered in a numerical simulation, which are more realistic and may make the results less critical. The analytical solution as well as the numerical simulations use the experimental experience of the maximum hot spot power loss of implanted resonators with a definite volume during magnetic resonance imaging investigations. The finite volume analysis calculates the time developing temperature maps for the model of a broken linear metallic wire embedded in tissue. Half of the total hot spot power loss is assumed to diffuse into both wire parts at the location of a defect. The energy is distributed from there by heat conduction. Additionally the effect of blood perfusion and blood flow is respected in some simulations because the simultaneous appearance of all worst case conditions, especially the absence of blood perfusion and blood flow near the hot spot, is very unlikely for vessel implants. RESULTS: The analytical solution as worst case scenario as well as the finite volume analysis for near worst case situations show not negligible volumes with critical temperature increases for part of the modeled hot spot situations. MR investigations with a high rf-pulse density lasting below a minute can establish volumes of several cubic millimeters with temperature increases high enough to start cell destruction. Longer exposure times can involve volumes larger than 100 mm(3). Even temperature increases in the range of thermal ablation are reached for substantial volumes. MR sequence exposure time and hot spot power loss are the primary factors influencing the volume with critical temperature increases. Wire radius, wire material as well as the physiological parameters blood perfusion and blood flow inside larger vessels reduce the volume with critical temperature increases, but do not exclude a volume with critical tissue heating for resonators with a large product of resonator volume and quality factor. CONCLUSION: The worst case scenario assumes thermal equilibrium for a hot spot embedded in homogeneous tissue without any cooling due to blood perfusion or flow. The finite volume analysis can calculate the results for near and not close to worst case conditions. For both cases a substantial volume can reach a critical temperature increase in a short time. The analytical solution, as absolute worst case, points out that resonators with a small product of inductance volume and quality factor (Q V(ind )< 2 cm(3)) are definitely save. Stents for coronary vessels or resonators used as tracking devices for interventional procedures therefore have no risk of high temperature increases. The finite volume analysis shows for sure that also conditions not close to the worst case reach physiologically critical temperature increases for implants with a large product of inductance volume and quality factor (Q V(ind )> 10 cm(3)). Such resonators exclude patients from exactly the MRI investigation these devices are made for. BioMed Central 2006-05-26 /pmc/articles/PMC1513583/ /pubmed/16729878 http://dx.doi.org/10.1186/1475-925X-5-35 Text en Copyright © 2006 Busch et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Busch, Martin HJ
Vollmann, Wolfgang
Grönemeyer, Dietrich HW
Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots
title Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots
title_full Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots
title_fullStr Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots
title_full_unstemmed Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots
title_short Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots
title_sort finite volume analysis of temperature effects induced by active mri implants: 2. defects on active mri implants causing hot spots
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1513583/
https://www.ncbi.nlm.nih.gov/pubmed/16729878
http://dx.doi.org/10.1186/1475-925X-5-35
work_keys_str_mv AT buschmartinhj finitevolumeanalysisoftemperatureeffectsinducedbyactivemriimplants2defectsonactivemriimplantscausinghotspots
AT vollmannwolfgang finitevolumeanalysisoftemperatureeffectsinducedbyactivemriimplants2defectsonactivemriimplantscausinghotspots
AT gronemeyerdietrichhw finitevolumeanalysisoftemperatureeffectsinducedbyactivemriimplants2defectsonactivemriimplantscausinghotspots