In vitro hydrodynamic properties of the Miethke proGAV hydrocephalus shunt
BACKGROUND: Adjustable shunts are very popular in the management of hydrocephalus and are believed to help in minimizing the number of surgical revisions. The drawback with almost all constructions is that they may be accidentally readjusted in relatively weak magnetic fields (around 30–40 mTesla) M...
Autores principales: | Allin, David M, Czosnyka, Zofia H, Czosnyka, Marek, Richards, Hugh K, Pickard, John D |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2006
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1552084/ https://www.ncbi.nlm.nih.gov/pubmed/16808836 http://dx.doi.org/10.1186/1743-8454-3-9 |
Ejemplares similares
-
Investigation of the hydrodynamic properties of a new MRI-resistant programmable hydrocephalus shunt
por: Allin, David M, et al.
Publicado: (2008) -
Efficacy and safety of the Miethke programmable differential pressure valve (proGAV®2.0): a single-centre retrospective analysis
por: Hall, Benjamin J., et al.
Publicado: (2021) -
Retrospective single-center historical comparative study between proGAV and proGAV2.0 for surgical revision and implant duration
por: Brunner, E, et al.
Publicado: (2022) -
Laboratory study on "intracranial hypotension" created by pumping the chamber of a hydrocephalus shunt
por: Bromby, Adam, et al.
Publicado: (2007) -
Programmable hydrocephalus shunt which cannot be unwillingly re-adjusted even in 3T MRI magnet
por: Allin, David, et al.