Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences

BACKGROUND: The purpose of the study was the clinical implementation of a kV cone beam CT (CBCT) for setup correction in radiotherapy. PATIENTS AND METHODS: For evaluation of the setup correction workflow, six tumor patients (lung cancer, sacral chordoma, head-and-neck and paraspinal tumor, and two...

Descripción completa

Detalles Bibliográficos
Autores principales: Thilmann, Christoph, Nill, Simeon, Tücking, Thomas, Höss, Angelika, Hesse, Bernd, Dietrich, Lars, Bendl, Rolf, Rhein, Bernhard, Häring, Peter, Thieke, Christian, Oelfke, Uwe, Debus, Juergen, Huber, Peter
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1557518/
https://www.ncbi.nlm.nih.gov/pubmed/16723023
http://dx.doi.org/10.1186/1748-717X-1-16
_version_ 1782129383430946816
author Thilmann, Christoph
Nill, Simeon
Tücking, Thomas
Höss, Angelika
Hesse, Bernd
Dietrich, Lars
Bendl, Rolf
Rhein, Bernhard
Häring, Peter
Thieke, Christian
Oelfke, Uwe
Debus, Juergen
Huber, Peter
author_facet Thilmann, Christoph
Nill, Simeon
Tücking, Thomas
Höss, Angelika
Hesse, Bernd
Dietrich, Lars
Bendl, Rolf
Rhein, Bernhard
Häring, Peter
Thieke, Christian
Oelfke, Uwe
Debus, Juergen
Huber, Peter
author_sort Thilmann, Christoph
collection PubMed
description BACKGROUND: The purpose of the study was the clinical implementation of a kV cone beam CT (CBCT) for setup correction in radiotherapy. PATIENTS AND METHODS: For evaluation of the setup correction workflow, six tumor patients (lung cancer, sacral chordoma, head-and-neck and paraspinal tumor, and two prostate cancer patients) were selected. All patients were treated with fractionated stereotactic radiotherapy, five of them with intensity modulated radiotherapy (IMRT). For patient fixation, a scotch cast body frame or a vacuum pillow, each in combination with a scotch cast head mask, were used. The imaging equipment, consisting of an x-ray tube and a flat panel imager (FPI), was attached to a Siemens linear accelerator according to the in-line approach, i.e. with the imaging beam mounted opposite to the treatment beam sharing the same isocenter. For dose delivery, the treatment beam has to traverse the FPI which is mounted in the accessory tray below the multi-leaf collimator. For each patient, a predefined number of imaging projections over a range of at least 200 degrees were acquired. The fast reconstruction of the 3D-CBCT dataset was done with an implementation of the Feldkamp-David-Kress (FDK) algorithm. For the registration of the treatment planning CT with the acquired CBCT, an automatic mutual information matcher and manual matching was used. RESULTS AND DISCUSSION: Bony landmarks were easily detected and the table shifts for correction of setup deviations could be automatically calculated in all cases. The image quality was sufficient for a visual comparison of the desired target point with the isocenter visible on the CBCT. Soft tissue contrast was problematic for the prostate of an obese patient, but good in the lung tumor case. The detected maximum setup deviation was 3 mm for patients fixated with the body frame, and 6 mm for patients positioned in the vacuum pillow. Using an action level of 2 mm translational error, a target point correction was carried out in 4 cases. The additional workload of the described workflow compared to a normal treatment fraction led to an extra time of about 10–12 minutes, which can be further reduced by streamlining the different steps. CONCLUSION: The cone beam CT attached to a LINAC allows the acquisition of a CT scan of the patient in treatment position directly before treatment. Its image quality is sufficient for determining target point correction vectors. With the presented workflow, a target point correction within a clinically reasonable time frame is possible. This increases the treatment precision, and potentially the complex patient fixation techniques will become dispensable.
format Text
id pubmed-1557518
institution National Center for Biotechnology Information
language English
publishDate 2006
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-15575182006-08-30 Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences Thilmann, Christoph Nill, Simeon Tücking, Thomas Höss, Angelika Hesse, Bernd Dietrich, Lars Bendl, Rolf Rhein, Bernhard Häring, Peter Thieke, Christian Oelfke, Uwe Debus, Juergen Huber, Peter Radiat Oncol Research BACKGROUND: The purpose of the study was the clinical implementation of a kV cone beam CT (CBCT) for setup correction in radiotherapy. PATIENTS AND METHODS: For evaluation of the setup correction workflow, six tumor patients (lung cancer, sacral chordoma, head-and-neck and paraspinal tumor, and two prostate cancer patients) were selected. All patients were treated with fractionated stereotactic radiotherapy, five of them with intensity modulated radiotherapy (IMRT). For patient fixation, a scotch cast body frame or a vacuum pillow, each in combination with a scotch cast head mask, were used. The imaging equipment, consisting of an x-ray tube and a flat panel imager (FPI), was attached to a Siemens linear accelerator according to the in-line approach, i.e. with the imaging beam mounted opposite to the treatment beam sharing the same isocenter. For dose delivery, the treatment beam has to traverse the FPI which is mounted in the accessory tray below the multi-leaf collimator. For each patient, a predefined number of imaging projections over a range of at least 200 degrees were acquired. The fast reconstruction of the 3D-CBCT dataset was done with an implementation of the Feldkamp-David-Kress (FDK) algorithm. For the registration of the treatment planning CT with the acquired CBCT, an automatic mutual information matcher and manual matching was used. RESULTS AND DISCUSSION: Bony landmarks were easily detected and the table shifts for correction of setup deviations could be automatically calculated in all cases. The image quality was sufficient for a visual comparison of the desired target point with the isocenter visible on the CBCT. Soft tissue contrast was problematic for the prostate of an obese patient, but good in the lung tumor case. The detected maximum setup deviation was 3 mm for patients fixated with the body frame, and 6 mm for patients positioned in the vacuum pillow. Using an action level of 2 mm translational error, a target point correction was carried out in 4 cases. The additional workload of the described workflow compared to a normal treatment fraction led to an extra time of about 10–12 minutes, which can be further reduced by streamlining the different steps. CONCLUSION: The cone beam CT attached to a LINAC allows the acquisition of a CT scan of the patient in treatment position directly before treatment. Its image quality is sufficient for determining target point correction vectors. With the presented workflow, a target point correction within a clinically reasonable time frame is possible. This increases the treatment precision, and potentially the complex patient fixation techniques will become dispensable. BioMed Central 2006-05-24 /pmc/articles/PMC1557518/ /pubmed/16723023 http://dx.doi.org/10.1186/1748-717X-1-16 Text en Copyright © 2006 Christoph et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Thilmann, Christoph
Nill, Simeon
Tücking, Thomas
Höss, Angelika
Hesse, Bernd
Dietrich, Lars
Bendl, Rolf
Rhein, Bernhard
Häring, Peter
Thieke, Christian
Oelfke, Uwe
Debus, Juergen
Huber, Peter
Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences
title Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences
title_full Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences
title_fullStr Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences
title_full_unstemmed Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences
title_short Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences
title_sort correction of patient positioning errors based on in-line cone beam cts: clinical implementation and first experiences
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1557518/
https://www.ncbi.nlm.nih.gov/pubmed/16723023
http://dx.doi.org/10.1186/1748-717X-1-16
work_keys_str_mv AT thilmannchristoph correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT nillsimeon correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT tuckingthomas correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT hossangelika correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT hessebernd correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT dietrichlars correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT bendlrolf correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT rheinbernhard correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT haringpeter correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT thiekechristian correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT oelfkeuwe correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT debusjuergen correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences
AT huberpeter correctionofpatientpositioningerrorsbasedoninlineconebeamctsclinicalimplementationandfirstexperiences