Dosimetric comparison of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery for single brain metastasis

BACKGROUND: Helical Tomotherapy (HT) integrates linear accelerator and computerized tomography (CT) technology to deliver IMRT. Targets are localized (i.e. outlined as gross tumor volume [GTV] and planning target volume [PTV]) on the planning kVCT study while daily MVCT is used for correction of pat...

Descripción completa

Detalles Bibliográficos
Autores principales: Peñagarícano, José A, Yan, Yulong, Shi, Chengyu, Linskey, Mark E, Ratanatharathorn, Vaneerat
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1557668/
https://www.ncbi.nlm.nih.gov/pubmed/16887031
http://dx.doi.org/10.1186/1748-717X-1-26
_version_ 1782129392210673664
author Peñagarícano, José A
Yan, Yulong
Shi, Chengyu
Linskey, Mark E
Ratanatharathorn, Vaneerat
author_facet Peñagarícano, José A
Yan, Yulong
Shi, Chengyu
Linskey, Mark E
Ratanatharathorn, Vaneerat
author_sort Peñagarícano, José A
collection PubMed
description BACKGROUND: Helical Tomotherapy (HT) integrates linear accelerator and computerized tomography (CT) technology to deliver IMRT. Targets are localized (i.e. outlined as gross tumor volume [GTV] and planning target volume [PTV]) on the planning kVCT study while daily MVCT is used for correction of patient's set-up and assessment of inter-fraction anatomy changes. Based on dosimetric comparisons, this study aims to find dosimetric equivalency between single fraction HT and Gamma Knife(® )stereotactic radiosurgery (GKSRS) for the treatment of single brain metastasis. METHODS: The targeting MRI data set from the GKSRS were used for tomotherapy planning. Five patients with single brain metastasis treated with GKSRS were re-planned in the HT planning station using the same prescribed doses. There was no expansion of the GTV to create the PTV. Sub-volumes were created within the PTV and prescribed to the maximum dose seen in the GKSRS plans to imitate the hot spot normally seen in GKSRS. The PTV objective was set as a region at risk in HT planning using the same prescribed dose to the PTV periphery as seen in the corresponding GKSRS plan. The tumor volumes ranged from 437–1840 mm(3). RESULTS: Conformality indices are inconsistent between HT and GKSRS. HT generally shows larger lower isodose line volumes, has longer treatment time than GKSRS and can treat a much larger lesion than GKSRS. Both HT and GKSRS single fraction dose-volume toxicity may be prohibitive in treating single or multiple lesions depending on the number and the sizes of the lesions. CONCLUSION: Based on the trend for larger lower dose volumes and more constricted higher dose volumes in HT as compared to GKSRS, dosimetric equivalency was not reached between HT and GKSRS.
format Text
id pubmed-1557668
institution National Center for Biotechnology Information
language English
publishDate 2006
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-15576682006-08-31 Dosimetric comparison of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery for single brain metastasis Peñagarícano, José A Yan, Yulong Shi, Chengyu Linskey, Mark E Ratanatharathorn, Vaneerat Radiat Oncol Research BACKGROUND: Helical Tomotherapy (HT) integrates linear accelerator and computerized tomography (CT) technology to deliver IMRT. Targets are localized (i.e. outlined as gross tumor volume [GTV] and planning target volume [PTV]) on the planning kVCT study while daily MVCT is used for correction of patient's set-up and assessment of inter-fraction anatomy changes. Based on dosimetric comparisons, this study aims to find dosimetric equivalency between single fraction HT and Gamma Knife(® )stereotactic radiosurgery (GKSRS) for the treatment of single brain metastasis. METHODS: The targeting MRI data set from the GKSRS were used for tomotherapy planning. Five patients with single brain metastasis treated with GKSRS were re-planned in the HT planning station using the same prescribed doses. There was no expansion of the GTV to create the PTV. Sub-volumes were created within the PTV and prescribed to the maximum dose seen in the GKSRS plans to imitate the hot spot normally seen in GKSRS. The PTV objective was set as a region at risk in HT planning using the same prescribed dose to the PTV periphery as seen in the corresponding GKSRS plan. The tumor volumes ranged from 437–1840 mm(3). RESULTS: Conformality indices are inconsistent between HT and GKSRS. HT generally shows larger lower isodose line volumes, has longer treatment time than GKSRS and can treat a much larger lesion than GKSRS. Both HT and GKSRS single fraction dose-volume toxicity may be prohibitive in treating single or multiple lesions depending on the number and the sizes of the lesions. CONCLUSION: Based on the trend for larger lower dose volumes and more constricted higher dose volumes in HT as compared to GKSRS, dosimetric equivalency was not reached between HT and GKSRS. BioMed Central 2006-08-03 /pmc/articles/PMC1557668/ /pubmed/16887031 http://dx.doi.org/10.1186/1748-717X-1-26 Text en Copyright © 2006 Peñagarícano et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Peñagarícano, José A
Yan, Yulong
Shi, Chengyu
Linskey, Mark E
Ratanatharathorn, Vaneerat
Dosimetric comparison of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery for single brain metastasis
title Dosimetric comparison of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery for single brain metastasis
title_full Dosimetric comparison of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery for single brain metastasis
title_fullStr Dosimetric comparison of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery for single brain metastasis
title_full_unstemmed Dosimetric comparison of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery for single brain metastasis
title_short Dosimetric comparison of Helical Tomotherapy and Gamma Knife Stereotactic Radiosurgery for single brain metastasis
title_sort dosimetric comparison of helical tomotherapy and gamma knife stereotactic radiosurgery for single brain metastasis
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1557668/
https://www.ncbi.nlm.nih.gov/pubmed/16887031
http://dx.doi.org/10.1186/1748-717X-1-26
work_keys_str_mv AT penagaricanojosea dosimetriccomparisonofhelicaltomotherapyandgammaknifestereotacticradiosurgeryforsinglebrainmetastasis
AT yanyulong dosimetriccomparisonofhelicaltomotherapyandgammaknifestereotacticradiosurgeryforsinglebrainmetastasis
AT shichengyu dosimetriccomparisonofhelicaltomotherapyandgammaknifestereotacticradiosurgeryforsinglebrainmetastasis
AT linskeymarke dosimetriccomparisonofhelicaltomotherapyandgammaknifestereotacticradiosurgeryforsinglebrainmetastasis
AT ratanatharathornvaneerat dosimetriccomparisonofhelicaltomotherapyandgammaknifestereotacticradiosurgeryforsinglebrainmetastasis