In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals
BACKGROUND: We need a sensor to measure the convective heat transfer coefficient during ablation of the heart or liver. METHODS: We built a minimally invasive instrument to measure the in vivo convective heat transfer coefficient, h in animals, using a Wheatstone-bridge circuit, similar to a hot-wir...
Autores principales: | , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2006
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1635717/ https://www.ncbi.nlm.nih.gov/pubmed/17067386 http://dx.doi.org/10.1186/1475-925X-5-57 |
_version_ | 1782130709016608768 |
---|---|
author | Tangwongsan, Chanchana Chachati, Louay Webster, John G Farrell, Patrick V |
author_facet | Tangwongsan, Chanchana Chachati, Louay Webster, John G Farrell, Patrick V |
author_sort | Tangwongsan, Chanchana |
collection | PubMed |
description | BACKGROUND: We need a sensor to measure the convective heat transfer coefficient during ablation of the heart or liver. METHODS: We built a minimally invasive instrument to measure the in vivo convective heat transfer coefficient, h in animals, using a Wheatstone-bridge circuit, similar to a hot-wire anemometer circuit. One arm is connected to a steerable catheter sensor whose tip is a 1.9 mm × 3.2 mm thin film resistive temperature detector (RTD) sensor. We used a circulation system to simulate different flow rates at 39°C for in vitro experiments using distilled water, tap water and saline. We heated the sensor approximately 5°C above the fluid temperature. We measured the power consumed by the sensor and the resistance of the sensor during the experiments and analyzed these data to determine the value of the convective heat transfer coefficient at various flow rates. RESULTS: From 0 to 5 L/min, experimental values of h in W/(m(2)·K) were for distilled water 5100 to 13000, for tap water 5500 to 12300, and for saline 5400 to 13600. Theoretical values were 1900 to 10700. CONCLUSION: We believe this system is the smallest, most accurate method of minimally invasive measurement of in vivo h in animals and provides the least disturbance of flow. |
format | Text |
id | pubmed-1635717 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2006 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-16357172006-11-14 In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals Tangwongsan, Chanchana Chachati, Louay Webster, John G Farrell, Patrick V Biomed Eng Online Research BACKGROUND: We need a sensor to measure the convective heat transfer coefficient during ablation of the heart or liver. METHODS: We built a minimally invasive instrument to measure the in vivo convective heat transfer coefficient, h in animals, using a Wheatstone-bridge circuit, similar to a hot-wire anemometer circuit. One arm is connected to a steerable catheter sensor whose tip is a 1.9 mm × 3.2 mm thin film resistive temperature detector (RTD) sensor. We used a circulation system to simulate different flow rates at 39°C for in vitro experiments using distilled water, tap water and saline. We heated the sensor approximately 5°C above the fluid temperature. We measured the power consumed by the sensor and the resistance of the sensor during the experiments and analyzed these data to determine the value of the convective heat transfer coefficient at various flow rates. RESULTS: From 0 to 5 L/min, experimental values of h in W/(m(2)·K) were for distilled water 5100 to 13000, for tap water 5500 to 12300, and for saline 5400 to 13600. Theoretical values were 1900 to 10700. CONCLUSION: We believe this system is the smallest, most accurate method of minimally invasive measurement of in vivo h in animals and provides the least disturbance of flow. BioMed Central 2006-10-26 /pmc/articles/PMC1635717/ /pubmed/17067386 http://dx.doi.org/10.1186/1475-925X-5-57 Text en Copyright © 2006 Tangwongsan et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Tangwongsan, Chanchana Chachati, Louay Webster, John G Farrell, Patrick V In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals |
title | In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals |
title_full | In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals |
title_fullStr | In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals |
title_full_unstemmed | In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals |
title_short | In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals |
title_sort | in vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1635717/ https://www.ncbi.nlm.nih.gov/pubmed/17067386 http://dx.doi.org/10.1186/1475-925X-5-57 |
work_keys_str_mv | AT tangwongsanchanchana invitrocalibrationofasystemformeasurementofinvivoconvectiveheattransfercoefficientinanimals AT chachatilouay invitrocalibrationofasystemformeasurementofinvivoconvectiveheattransfercoefficientinanimals AT websterjohng invitrocalibrationofasystemformeasurementofinvivoconvectiveheattransfercoefficientinanimals AT farrellpatrickv invitrocalibrationofasystemformeasurementofinvivoconvectiveheattransfercoefficientinanimals |