Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System

BACKGROUND: Intraoperative Radiotherapy (IORT) is a method by which a critical radiation dose is delivered to the tumour bed immediately after surgical excision. It is being investigated whether a single high dose of radiation will impart the same clinical benefit as a standard course of external be...

Descripción completa

Detalles Bibliográficos
Autores principales: Armoogum, Kris S, Parry, John M, Souliman, Salam K, Sutton, David G, Mackay, Colin D
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1808461/
https://www.ncbi.nlm.nih.gov/pubmed/17326838
http://dx.doi.org/10.1186/1748-717X-2-11
_version_ 1782132539994931200
author Armoogum, Kris S
Parry, John M
Souliman, Salam K
Sutton, David G
Mackay, Colin D
author_facet Armoogum, Kris S
Parry, John M
Souliman, Salam K
Sutton, David G
Mackay, Colin D
author_sort Armoogum, Kris S
collection PubMed
description BACKGROUND: Intraoperative Radiotherapy (IORT) is a method by which a critical radiation dose is delivered to the tumour bed immediately after surgical excision. It is being investigated whether a single high dose of radiation will impart the same clinical benefit as a standard course of external beam therapy. Our centre has four Photon Radiosurgery Systems (PRS) currently used to irradiate breast and neurological sites. MATERIALS AND METHODS: The PRS comprises an x-ray generator, control console, quality assurance tools and a mobile gantry. We investigated the dosimetric characteristics of each source and its performance stability over a period of time. We investigated half value layer, output diminution factor, internal radiation monitor (IRM) reproducibility and depth-doses in water. The half value layer was determined in air by the broad beam method, using high purity aluminium attenuators. To quantify beam hardening at clinical depths, solid water attenuators of 5 and 10 mm were placed between the x-ray probe and attenuators. The ion chamber current was monitored over 30 minutes to deduce an output diminution factor. IRM reproducibility was investigated under various exposures. Depth-dose curves in water were obtained at distances up to 35 mm from the probe. RESULTS: The mean energies for the beam attenuated by 5 and 10 mm of solid water were derived from ICRU Report 17 and found to be 18 and 24 keV. The average output level over a period of 30 minutes was found to be 99.12%. The average difference between the preset IRM limit and the total IRM count was less than 0.5%. For three x-ray sources, the average difference between the calculated and actual treatment times was found to be 0.62% (n = 30). The beam attenuation in water varied by approximately 1/r(3). CONCLUSION: The x-ray sources are stable over time. Most measurements were found to lie within the manufacturer's tolerances and an intercomparison of these checks suggests that the four x-ray sources have similar performance characteristics.
format Text
id pubmed-1808461
institution National Center for Biotechnology Information
language English
publishDate 2007
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-18084612007-03-03 Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System Armoogum, Kris S Parry, John M Souliman, Salam K Sutton, David G Mackay, Colin D Radiat Oncol Research BACKGROUND: Intraoperative Radiotherapy (IORT) is a method by which a critical radiation dose is delivered to the tumour bed immediately after surgical excision. It is being investigated whether a single high dose of radiation will impart the same clinical benefit as a standard course of external beam therapy. Our centre has four Photon Radiosurgery Systems (PRS) currently used to irradiate breast and neurological sites. MATERIALS AND METHODS: The PRS comprises an x-ray generator, control console, quality assurance tools and a mobile gantry. We investigated the dosimetric characteristics of each source and its performance stability over a period of time. We investigated half value layer, output diminution factor, internal radiation monitor (IRM) reproducibility and depth-doses in water. The half value layer was determined in air by the broad beam method, using high purity aluminium attenuators. To quantify beam hardening at clinical depths, solid water attenuators of 5 and 10 mm were placed between the x-ray probe and attenuators. The ion chamber current was monitored over 30 minutes to deduce an output diminution factor. IRM reproducibility was investigated under various exposures. Depth-dose curves in water were obtained at distances up to 35 mm from the probe. RESULTS: The mean energies for the beam attenuated by 5 and 10 mm of solid water were derived from ICRU Report 17 and found to be 18 and 24 keV. The average output level over a period of 30 minutes was found to be 99.12%. The average difference between the preset IRM limit and the total IRM count was less than 0.5%. For three x-ray sources, the average difference between the calculated and actual treatment times was found to be 0.62% (n = 30). The beam attenuation in water varied by approximately 1/r(3). CONCLUSION: The x-ray sources are stable over time. Most measurements were found to lie within the manufacturer's tolerances and an intercomparison of these checks suggests that the four x-ray sources have similar performance characteristics. BioMed Central 2007-02-27 /pmc/articles/PMC1808461/ /pubmed/17326838 http://dx.doi.org/10.1186/1748-717X-2-11 Text en Copyright © 2007 Armoogum et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Armoogum, Kris S
Parry, John M
Souliman, Salam K
Sutton, David G
Mackay, Colin D
Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System
title Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System
title_full Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System
title_fullStr Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System
title_full_unstemmed Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System
title_short Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System
title_sort functional intercomparison of intraoperative radiotherapy equipment – photon radiosurgery system
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC1808461/
https://www.ncbi.nlm.nih.gov/pubmed/17326838
http://dx.doi.org/10.1186/1748-717X-2-11
work_keys_str_mv AT armoogumkriss functionalintercomparisonofintraoperativeradiotherapyequipmentphotonradiosurgerysystem
AT parryjohnm functionalintercomparisonofintraoperativeradiotherapyequipmentphotonradiosurgerysystem
AT soulimansalamk functionalintercomparisonofintraoperativeradiotherapyequipmentphotonradiosurgerysystem
AT suttondavidg functionalintercomparisonofintraoperativeradiotherapyequipmentphotonradiosurgerysystem
AT mackaycolind functionalintercomparisonofintraoperativeradiotherapyequipmentphotonradiosurgerysystem