Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study

BACKGROUND: Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the...

Descripción completa

Detalles Bibliográficos
Autores principales: van Sörnsen de Koste, John R, Cuijpers, Johan P, de Geest, Frank GM, Lagerwaard, Frank J, Slotman, Ben J, Senan, Suresh
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2075522/
https://www.ncbi.nlm.nih.gov/pubmed/17760960
http://dx.doi.org/10.1186/1748-717X-2-32
_version_ 1782138074811072512
author van Sörnsen de Koste, John R
Cuijpers, Johan P
de Geest, Frank GM
Lagerwaard, Frank J
Slotman, Ben J
Senan, Suresh
author_facet van Sörnsen de Koste, John R
Cuijpers, Johan P
de Geest, Frank GM
Lagerwaard, Frank J
Slotman, Ben J
Senan, Suresh
author_sort van Sörnsen de Koste, John R
collection PubMed
description BACKGROUND: Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the position of internal structures during RGRT delivery METHODS: TI-EPI portals were generated by continuously collecting exit dose data (aSi500 EPID, Portal vision, Varian Medical Systems) when a respiratory motion phantom was irradiated during expiration, inspiration and free breathing phases. RGRT was delivered using the Varian RPM system, and grey value profile plots over a fixed trajectory were used to study object positions. Time-related positional information was derived by subtracting grey values from TI-EPI portals sharing the pixel matrix. TI-EPI portals were also collected in 2 patients undergoing RPM-triggered RGRT for a lung and hepatic tumor (with fiducial markers), and corresponding planning 4-dimensional CT (4DCT) scans were analyzed for motion amplitude. RESULTS: Integral grey values of phantom TI-EPI portals correlated well with mean object position in all respiratory phases. Cranio-caudal motion of internal structures ranged from 17.5–20.0 mm on planning 4DCT scans. TI-EPI of bronchial images reproduced with a mean value of 5.3 mm (1 SD 3.0 mm) located cranial to planned position. Mean hepatic fiducial markers reproduced with 3.2 mm (SD 2.2 mm) caudal to planned position. After bony alignment to exclude set-up errors, mean displacement in the two structures was 2.8 mm and 1.4 mm, respectively, and corresponding reproducibility in anatomy improved to 1.6 mm (1 SD). CONCLUSION: TI-EPI appears to be a promising method for verifying delivery of RGRT. The RPM system was a good indirect surrogate of internal anatomy, but use of TI-EPI allowed for a direct link between anatomy and breathing patterns.
format Text
id pubmed-2075522
institution National Center for Biotechnology Information
language English
publishDate 2007
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-20755222007-11-13 Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study van Sörnsen de Koste, John R Cuijpers, Johan P de Geest, Frank GM Lagerwaard, Frank J Slotman, Ben J Senan, Suresh Radiat Oncol Research BACKGROUND: Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the position of internal structures during RGRT delivery METHODS: TI-EPI portals were generated by continuously collecting exit dose data (aSi500 EPID, Portal vision, Varian Medical Systems) when a respiratory motion phantom was irradiated during expiration, inspiration and free breathing phases. RGRT was delivered using the Varian RPM system, and grey value profile plots over a fixed trajectory were used to study object positions. Time-related positional information was derived by subtracting grey values from TI-EPI portals sharing the pixel matrix. TI-EPI portals were also collected in 2 patients undergoing RPM-triggered RGRT for a lung and hepatic tumor (with fiducial markers), and corresponding planning 4-dimensional CT (4DCT) scans were analyzed for motion amplitude. RESULTS: Integral grey values of phantom TI-EPI portals correlated well with mean object position in all respiratory phases. Cranio-caudal motion of internal structures ranged from 17.5–20.0 mm on planning 4DCT scans. TI-EPI of bronchial images reproduced with a mean value of 5.3 mm (1 SD 3.0 mm) located cranial to planned position. Mean hepatic fiducial markers reproduced with 3.2 mm (SD 2.2 mm) caudal to planned position. After bony alignment to exclude set-up errors, mean displacement in the two structures was 2.8 mm and 1.4 mm, respectively, and corresponding reproducibility in anatomy improved to 1.6 mm (1 SD). CONCLUSION: TI-EPI appears to be a promising method for verifying delivery of RGRT. The RPM system was a good indirect surrogate of internal anatomy, but use of TI-EPI allowed for a direct link between anatomy and breathing patterns. BioMed Central 2007-08-30 /pmc/articles/PMC2075522/ /pubmed/17760960 http://dx.doi.org/10.1186/1748-717X-2-32 Text en Copyright © 2007 van Sörnsen de Koste et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
van Sörnsen de Koste, John R
Cuijpers, Johan P
de Geest, Frank GM
Lagerwaard, Frank J
Slotman, Ben J
Senan, Suresh
Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
title Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
title_full Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
title_fullStr Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
title_full_unstemmed Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
title_short Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
title_sort verifying 4d gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2075522/
https://www.ncbi.nlm.nih.gov/pubmed/17760960
http://dx.doi.org/10.1186/1748-717X-2-32
work_keys_str_mv AT vansornsendekostejohnr verifying4dgatedradiotherapyusingtimeintegratedelectronicportalimagingaphantomandclinicalstudy
AT cuijpersjohanp verifying4dgatedradiotherapyusingtimeintegratedelectronicportalimagingaphantomandclinicalstudy
AT degeestfrankgm verifying4dgatedradiotherapyusingtimeintegratedelectronicportalimagingaphantomandclinicalstudy
AT lagerwaardfrankj verifying4dgatedradiotherapyusingtimeintegratedelectronicportalimagingaphantomandclinicalstudy
AT slotmanbenj verifying4dgatedradiotherapyusingtimeintegratedelectronicportalimagingaphantomandclinicalstudy
AT senansuresh verifying4dgatedradiotherapyusingtimeintegratedelectronicportalimagingaphantomandclinicalstudy