Cargando…

Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters

Resolution for 125I-labeled specimens under electron microscope (EM) autoradiographic conditions was assessed experimentally. With this isotope the size of the silver halide crystal was the most important resolution-limiting factor. Heavy metal staining such as is routinely used in preparing animal...

Descripción completa

Detalles Bibliográficos
Formato: Texto
Lenguaje:English
Publicado: The Rockefeller University Press 1977
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2110981/
https://www.ncbi.nlm.nih.gov/pubmed/63463
Descripción
Sumario:Resolution for 125I-labeled specimens under electron microscope (EM) autoradiographic conditions was assessed experimentally. With this isotope the size of the silver halide crystal was the most important resolution-limiting factor. Heavy metal staining such as is routinely used in preparing animal tissues for EM autoradiography produced an improvement in resolution of approximately 15-20%. For a 500-1,000-A biological tissue section fixed with OsO4 and stained with uranyl acetate, we obtained resolution (half distance, HD) values of approximately 800 +/- 120 A using Ilford L4 emulsion and 500 +/- 70 A using a Kodak NTE-type emulsion. General aspects of resolution-limiting factors and comparison with 3H and 14C values are discussed.