Cargando…

Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters

Resolution for 125I-labeled specimens under electron microscope (EM) autoradiographic conditions was assessed experimentally. With this isotope the size of the silver halide crystal was the most important resolution-limiting factor. Heavy metal staining such as is routinely used in preparing animal...

Descripción completa

Detalles Bibliográficos
Formato: Texto
Lenguaje:English
Publicado: The Rockefeller University Press 1977
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2110981/
https://www.ncbi.nlm.nih.gov/pubmed/63463
_version_ 1782139701694562304
collection PubMed
description Resolution for 125I-labeled specimens under electron microscope (EM) autoradiographic conditions was assessed experimentally. With this isotope the size of the silver halide crystal was the most important resolution-limiting factor. Heavy metal staining such as is routinely used in preparing animal tissues for EM autoradiography produced an improvement in resolution of approximately 15-20%. For a 500-1,000-A biological tissue section fixed with OsO4 and stained with uranyl acetate, we obtained resolution (half distance, HD) values of approximately 800 +/- 120 A using Ilford L4 emulsion and 500 +/- 70 A using a Kodak NTE-type emulsion. General aspects of resolution-limiting factors and comparison with 3H and 14C values are discussed.
format Text
id pubmed-2110981
institution National Center for Biotechnology Information
language English
publishDate 1977
publisher The Rockefeller University Press
record_format MEDLINE/PubMed
spelling pubmed-21109812008-05-01 Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters J Cell Biol Articles Resolution for 125I-labeled specimens under electron microscope (EM) autoradiographic conditions was assessed experimentally. With this isotope the size of the silver halide crystal was the most important resolution-limiting factor. Heavy metal staining such as is routinely used in preparing animal tissues for EM autoradiography produced an improvement in resolution of approximately 15-20%. For a 500-1,000-A biological tissue section fixed with OsO4 and stained with uranyl acetate, we obtained resolution (half distance, HD) values of approximately 800 +/- 120 A using Ilford L4 emulsion and 500 +/- 70 A using a Kodak NTE-type emulsion. General aspects of resolution-limiting factors and comparison with 3H and 14C values are discussed. The Rockefeller University Press 1977-01-01 /pmc/articles/PMC2110981/ /pubmed/63463 Text en This article is distributed under the terms of an Attribution–Noncommercial–Share Alike–No Mirror Sites license for the first six months after the publication date (see http://www.rupress.org/terms). After six months it is available under a Creative Commons License (Attribution–Noncommercial–Share Alike 4.0 Unported license, as described at http://creativecommons.org/licenses/by-nc-sa/4.0/).
spellingShingle Articles
Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters
title Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters
title_full Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters
title_fullStr Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters
title_full_unstemmed Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters
title_short Resolution in electron microscope autoradiography. III. Iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters
title_sort resolution in electron microscope autoradiography. iii. iodine-125, the effect of heavy metal staining, and a reassessment of critical parameters
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2110981/
https://www.ncbi.nlm.nih.gov/pubmed/63463