HUMORAL RHEOLOGY : I. VISCOSITY STUDIES AND ANOMALOUS FLOW PROPERTIES OF HUMAN BLOOD SYSTEMS WITH HEPARIN AND OTHER ANTICOAGULANTS

1. A modified falling ball viscometer (rolling ball viscometer) for blood and other humors is presented. It is capable of easily measuring flow properties at several stresses, as is required to define satisfactorily the properties of anomalous flow systems. At high shearing stresses, apparent specif...

Descripción completa

Detalles Bibliográficos
Autores principales: Copley, Alfred L., Krchma, Ludwig C., Whitney, Mary E.
Formato: Texto
Lenguaje:English
Publicado: The Rockefeller University Press 1942
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2142051/
https://www.ncbi.nlm.nih.gov/pubmed/19873328
_version_ 1782144286226120704
author Copley, Alfred L.
Krchma, Ludwig C.
Whitney, Mary E.
author_facet Copley, Alfred L.
Krchma, Ludwig C.
Whitney, Mary E.
author_sort Copley, Alfred L.
collection PubMed
description 1. A modified falling ball viscometer (rolling ball viscometer) for blood and other humors is presented. It is capable of easily measuring flow properties at several stresses, as is required to define satisfactorily the properties of anomalous flow systems. At high shearing stresses, apparent specific viscosity values of 2.5 + are observed, corresponding to 2.2 ±0.2 reported as possible with the biological viscometer of Whittaker and Winton. 2. Previous references to the anomalous flow properties of blood were verified. It was demonstrated that these systems conform to the Bingham concept of anomalous flow. To define completely the flow properties of such systems it is necessary to make determinations at at least two shearing stresses, preferably more. Data are reported for the pseudoviscosity and yield value, the latter being possibly the most specific property of the three bloods studied. 3. Heparin in increasing amounts tended to decrease the apparent viscosity, pseudoviscosity, and yield value of blood. Similar increases of heparin also reduced the viscosites of the serum and plasma. 4. The ratio of the apparent viscosity of blood and its plasma was found to be reasonably constant as reported by Trevan. However, as the apparent viscosity is a function of the shearing stress, it is believed that the relationship for the calculations of corpuscular concentrations, such as the Whittaker and Winton modification of the Hatschek formula, is specific for the instrument and conditions of tests by which it was determined. 5. Heparinized blood was found to exhibit thixotropy, dilatancy, and age-hardening phenomena.
format Text
id pubmed-2142051
institution National Center for Biotechnology Information
language English
publishDate 1942
publisher The Rockefeller University Press
record_format MEDLINE/PubMed
spelling pubmed-21420512008-04-23 HUMORAL RHEOLOGY : I. VISCOSITY STUDIES AND ANOMALOUS FLOW PROPERTIES OF HUMAN BLOOD SYSTEMS WITH HEPARIN AND OTHER ANTICOAGULANTS Copley, Alfred L. Krchma, Ludwig C. Whitney, Mary E. J Gen Physiol Article 1. A modified falling ball viscometer (rolling ball viscometer) for blood and other humors is presented. It is capable of easily measuring flow properties at several stresses, as is required to define satisfactorily the properties of anomalous flow systems. At high shearing stresses, apparent specific viscosity values of 2.5 + are observed, corresponding to 2.2 ±0.2 reported as possible with the biological viscometer of Whittaker and Winton. 2. Previous references to the anomalous flow properties of blood were verified. It was demonstrated that these systems conform to the Bingham concept of anomalous flow. To define completely the flow properties of such systems it is necessary to make determinations at at least two shearing stresses, preferably more. Data are reported for the pseudoviscosity and yield value, the latter being possibly the most specific property of the three bloods studied. 3. Heparin in increasing amounts tended to decrease the apparent viscosity, pseudoviscosity, and yield value of blood. Similar increases of heparin also reduced the viscosites of the serum and plasma. 4. The ratio of the apparent viscosity of blood and its plasma was found to be reasonably constant as reported by Trevan. However, as the apparent viscosity is a function of the shearing stress, it is believed that the relationship for the calculations of corpuscular concentrations, such as the Whittaker and Winton modification of the Hatschek formula, is specific for the instrument and conditions of tests by which it was determined. 5. Heparinized blood was found to exhibit thixotropy, dilatancy, and age-hardening phenomena. The Rockefeller University Press 1942-09-20 /pmc/articles/PMC2142051/ /pubmed/19873328 Text en Copyright © Copyright, 1942, by The Rockefeller Institute for Medical Research This article is distributed under the terms of an Attribution–Noncommercial–Share Alike–No Mirror Sites license for the first six months after the publication date (see http://www.rupress.org/terms). After six months it is available under a Creative Commons License (Attribution–Noncommercial–Share Alike 4.0 Unported license, as described at http://creativecommons.org/licenses/by-nc-sa/4.0/).
spellingShingle Article
Copley, Alfred L.
Krchma, Ludwig C.
Whitney, Mary E.
HUMORAL RHEOLOGY : I. VISCOSITY STUDIES AND ANOMALOUS FLOW PROPERTIES OF HUMAN BLOOD SYSTEMS WITH HEPARIN AND OTHER ANTICOAGULANTS
title HUMORAL RHEOLOGY : I. VISCOSITY STUDIES AND ANOMALOUS FLOW PROPERTIES OF HUMAN BLOOD SYSTEMS WITH HEPARIN AND OTHER ANTICOAGULANTS
title_full HUMORAL RHEOLOGY : I. VISCOSITY STUDIES AND ANOMALOUS FLOW PROPERTIES OF HUMAN BLOOD SYSTEMS WITH HEPARIN AND OTHER ANTICOAGULANTS
title_fullStr HUMORAL RHEOLOGY : I. VISCOSITY STUDIES AND ANOMALOUS FLOW PROPERTIES OF HUMAN BLOOD SYSTEMS WITH HEPARIN AND OTHER ANTICOAGULANTS
title_full_unstemmed HUMORAL RHEOLOGY : I. VISCOSITY STUDIES AND ANOMALOUS FLOW PROPERTIES OF HUMAN BLOOD SYSTEMS WITH HEPARIN AND OTHER ANTICOAGULANTS
title_short HUMORAL RHEOLOGY : I. VISCOSITY STUDIES AND ANOMALOUS FLOW PROPERTIES OF HUMAN BLOOD SYSTEMS WITH HEPARIN AND OTHER ANTICOAGULANTS
title_sort humoral rheology : i. viscosity studies and anomalous flow properties of human blood systems with heparin and other anticoagulants
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2142051/
https://www.ncbi.nlm.nih.gov/pubmed/19873328
work_keys_str_mv AT copleyalfredl humoralrheologyiviscositystudiesandanomalousflowpropertiesofhumanbloodsystemswithheparinandotheranticoagulants
AT krchmaludwigc humoralrheologyiviscositystudiesandanomalousflowpropertiesofhumanbloodsystemswithheparinandotheranticoagulants
AT whitneymarye humoralrheologyiviscositystudiesandanomalousflowpropertiesofhumanbloodsystemswithheparinandotheranticoagulants