Cargando…
Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study
Many middle-aged patients are affected by localized cartilage defects that are neither appropriate for primary, nor repeat biological repair methods, nor for conventional arthroplasty. This in vitro study aims to determine the peak contact pressure in the tibiofemoral joint with a partial femoral re...
Autores principales: | , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Springer-Verlag
2007
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2190783/ https://www.ncbi.nlm.nih.gov/pubmed/17934718 http://dx.doi.org/10.1007/s00167-007-0416-7 |
_version_ | 1782146852169187328 |
---|---|
author | Becher, Christoph Huber, Roland Thermann, Hajo Paessler, Hans H. Skrbensky, Gobert |
author_facet | Becher, Christoph Huber, Roland Thermann, Hajo Paessler, Hans H. Skrbensky, Gobert |
author_sort | Becher, Christoph |
collection | PubMed |
description | Many middle-aged patients are affected by localized cartilage defects that are neither appropriate for primary, nor repeat biological repair methods, nor for conventional arthroplasty. This in vitro study aims to determine the peak contact pressure in the tibiofemoral joint with a partial femoral resurfacing device (HemiCAP(®), Arthrosurface Inc., Franklin, MA, USA). Peak contact pressure was determined in eight fresh-frozen cadaveric specimens using a Tekscan sensor placed in the medial compartment above the menisci. A closed loop robotic knee simulator was used to test each knee in static stance positions (5°/15°/30°/45°) with body weight ground reaction force (GRF), 30° flexion with twice the body weight (2tBW) GRF and dynamic knee-bending cycles with body weight GRF. The ground reaction force was adjusted to the living body weight of the cadaver donor and maintained throughout all cycles. Each specimen was tested under four different conditions: Untreated, flush HemiCAP(®) implantation, 1-mm proud implantation and 20-mm defect. A paired sampled t test to compare means (significance, P ≤ 0.05) was used for statistical analysis. On average, no statistically significant differences were found in any testing condition comparing the normal knee with flush device implantation. With the 1-mm proud implant, statistically significant increase of peak contact pressures of 217% (5° stance), 99% (dynamic knee bending) and 90% (30° stance with 2tBW) compared to the untreated condition was seen. No significant increase of peak contact pressure was evaluated with the 20-mm defect. The data suggests that resurfacing with the HemiCAP(®) does not lead to increased peak contact pressure with flush implantation. However, elevated implantation results in increased peak contact pressure and might be biomechanically disadvantageous in an in vivo application. |
format | Text |
id | pubmed-2190783 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2007 |
publisher | Springer-Verlag |
record_format | MEDLINE/PubMed |
spelling | pubmed-21907832008-01-12 Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study Becher, Christoph Huber, Roland Thermann, Hajo Paessler, Hans H. Skrbensky, Gobert Knee Surg Sports Traumatol Arthrosc Knee Many middle-aged patients are affected by localized cartilage defects that are neither appropriate for primary, nor repeat biological repair methods, nor for conventional arthroplasty. This in vitro study aims to determine the peak contact pressure in the tibiofemoral joint with a partial femoral resurfacing device (HemiCAP(®), Arthrosurface Inc., Franklin, MA, USA). Peak contact pressure was determined in eight fresh-frozen cadaveric specimens using a Tekscan sensor placed in the medial compartment above the menisci. A closed loop robotic knee simulator was used to test each knee in static stance positions (5°/15°/30°/45°) with body weight ground reaction force (GRF), 30° flexion with twice the body weight (2tBW) GRF and dynamic knee-bending cycles with body weight GRF. The ground reaction force was adjusted to the living body weight of the cadaver donor and maintained throughout all cycles. Each specimen was tested under four different conditions: Untreated, flush HemiCAP(®) implantation, 1-mm proud implantation and 20-mm defect. A paired sampled t test to compare means (significance, P ≤ 0.05) was used for statistical analysis. On average, no statistically significant differences were found in any testing condition comparing the normal knee with flush device implantation. With the 1-mm proud implant, statistically significant increase of peak contact pressures of 217% (5° stance), 99% (dynamic knee bending) and 90% (30° stance with 2tBW) compared to the untreated condition was seen. No significant increase of peak contact pressure was evaluated with the 20-mm defect. The data suggests that resurfacing with the HemiCAP(®) does not lead to increased peak contact pressure with flush implantation. However, elevated implantation results in increased peak contact pressure and might be biomechanically disadvantageous in an in vivo application. Springer-Verlag 2007-10-13 2008-01 /pmc/articles/PMC2190783/ /pubmed/17934718 http://dx.doi.org/10.1007/s00167-007-0416-7 Text en © Springer-Verlag 2007 |
spellingShingle | Knee Becher, Christoph Huber, Roland Thermann, Hajo Paessler, Hans H. Skrbensky, Gobert Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study |
title | Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study |
title_full | Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study |
title_fullStr | Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study |
title_full_unstemmed | Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study |
title_short | Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study |
title_sort | effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study |
topic | Knee |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2190783/ https://www.ncbi.nlm.nih.gov/pubmed/17934718 http://dx.doi.org/10.1007/s00167-007-0416-7 |
work_keys_str_mv | AT becherchristoph effectsofacontouredarticularprostheticdeviceontibiofemoralpeakcontactpressureabiomechanicalstudy AT huberroland effectsofacontouredarticularprostheticdeviceontibiofemoralpeakcontactpressureabiomechanicalstudy AT thermannhajo effectsofacontouredarticularprostheticdeviceontibiofemoralpeakcontactpressureabiomechanicalstudy AT paesslerhansh effectsofacontouredarticularprostheticdeviceontibiofemoralpeakcontactpressureabiomechanicalstudy AT skrbenskygobert effectsofacontouredarticularprostheticdeviceontibiofemoralpeakcontactpressureabiomechanicalstudy |