Intracellular calibration of a pH-sensitive dye in isolated, perfused salamander proximal tubules

We evaluated the dye 4',5'-dimethyl-5-(and -6-) carboxyfluorescein (Me2CF) for determining the intracellular pH(pHi) of isolated, perfused proximal tubules of the salamander. The intracellular absorbance spectrum, corrected for the intrinsic absorbance of the tubule, was obtained once per...

Descripción completa

Detalles Bibliográficos
Formato: Texto
Lenguaje:English
Publicado: The Rockefeller University Press 1985
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2228795/
https://www.ncbi.nlm.nih.gov/pubmed/4078557
_version_ 1782149977216122880
collection PubMed
description We evaluated the dye 4',5'-dimethyl-5-(and -6-) carboxyfluorescein (Me2CF) for determining the intracellular pH(pHi) of isolated, perfused proximal tubules of the salamander. The intracellular absorbance spectrum, corrected for the intrinsic absorbance of the tubule, was obtained once per second. The dye was incorporated into tubule cells by exposing them to the membrane-permeable precursor 4',5'-dimethyl-5- (and -6-) carboxyfluorescein diacetate. The introduction of the dye had no significant effect on either pHi or cell voltage transients. Compared with dye contained in a cuvette, intracellular dye had a peak absorbance that was red-shifted by approximately 5 nm, and an apparent pK that was increased by approximately 0.3. These differences precluded an accurate calculation of pHi by the comparison of intracellular spectra with in vitro calibration spectra. However, when Me2CF was calibrated intracellularly, using the K-H exchanger nigericin to equalize external pH and pHi, the dye-derived, steady state pHi was within approximately 0.1 of the value obtained with pH-sensitive microelectrodes. Furthermore, when pHi was simultaneously measured with dye and microelectrodes during rapid pHi transients, the pHi time courses measured by the two methods were very similar. We conclude that the intracellular absorbance spectrum of Me2CF can be used to measure steady state pHi and rapid pHi transients reliably, provided the dye is calibrated intracellularly.
format Text
id pubmed-2228795
institution National Center for Biotechnology Information
language English
publishDate 1985
publisher The Rockefeller University Press
record_format MEDLINE/PubMed
spelling pubmed-22287952008-04-23 Intracellular calibration of a pH-sensitive dye in isolated, perfused salamander proximal tubules J Gen Physiol Articles We evaluated the dye 4',5'-dimethyl-5-(and -6-) carboxyfluorescein (Me2CF) for determining the intracellular pH(pHi) of isolated, perfused proximal tubules of the salamander. The intracellular absorbance spectrum, corrected for the intrinsic absorbance of the tubule, was obtained once per second. The dye was incorporated into tubule cells by exposing them to the membrane-permeable precursor 4',5'-dimethyl-5- (and -6-) carboxyfluorescein diacetate. The introduction of the dye had no significant effect on either pHi or cell voltage transients. Compared with dye contained in a cuvette, intracellular dye had a peak absorbance that was red-shifted by approximately 5 nm, and an apparent pK that was increased by approximately 0.3. These differences precluded an accurate calculation of pHi by the comparison of intracellular spectra with in vitro calibration spectra. However, when Me2CF was calibrated intracellularly, using the K-H exchanger nigericin to equalize external pH and pHi, the dye-derived, steady state pHi was within approximately 0.1 of the value obtained with pH-sensitive microelectrodes. Furthermore, when pHi was simultaneously measured with dye and microelectrodes during rapid pHi transients, the pHi time courses measured by the two methods were very similar. We conclude that the intracellular absorbance spectrum of Me2CF can be used to measure steady state pHi and rapid pHi transients reliably, provided the dye is calibrated intracellularly. The Rockefeller University Press 1985-12-01 /pmc/articles/PMC2228795/ /pubmed/4078557 Text en This article is distributed under the terms of an Attribution–Noncommercial–Share Alike–No Mirror Sites license for the first six months after the publication date (see http://www.rupress.org/terms). After six months it is available under a Creative Commons License (Attribution–Noncommercial–Share Alike 4.0 Unported license, as described at http://creativecommons.org/licenses/by-nc-sa/4.0/).
spellingShingle Articles
Intracellular calibration of a pH-sensitive dye in isolated, perfused salamander proximal tubules
title Intracellular calibration of a pH-sensitive dye in isolated, perfused salamander proximal tubules
title_full Intracellular calibration of a pH-sensitive dye in isolated, perfused salamander proximal tubules
title_fullStr Intracellular calibration of a pH-sensitive dye in isolated, perfused salamander proximal tubules
title_full_unstemmed Intracellular calibration of a pH-sensitive dye in isolated, perfused salamander proximal tubules
title_short Intracellular calibration of a pH-sensitive dye in isolated, perfused salamander proximal tubules
title_sort intracellular calibration of a ph-sensitive dye in isolated, perfused salamander proximal tubules
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2228795/
https://www.ncbi.nlm.nih.gov/pubmed/4078557