Cargando…
A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions
INTRODUCTION: Characterizing and differentiating between malignant tumors, benign tumors, and normal breast tissue is increasingly important in the patient presenting with breast problems. Near-infrared diffuse optical imaging and spectroscopy is capable of measuring multiple physiologic parameters...
Autores principales: | , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2007
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2246191/ https://www.ncbi.nlm.nih.gov/pubmed/18088411 http://dx.doi.org/10.1186/bcr1837 |
_version_ | 1782150736726982656 |
---|---|
author | Xu, Ronald X Young, Donn C Mao, Jimmy J Povoski, Stephen P |
author_facet | Xu, Ronald X Young, Donn C Mao, Jimmy J Povoski, Stephen P |
author_sort | Xu, Ronald X |
collection | PubMed |
description | INTRODUCTION: Characterizing and differentiating between malignant tumors, benign tumors, and normal breast tissue is increasingly important in the patient presenting with breast problems. Near-infrared diffuse optical imaging and spectroscopy is capable of measuring multiple physiologic parameters of biological tissue systems and may have clinical applications for assessing the development and progression of neoplastic processes, including breast cancer. The currently available application of near-infrared imaging technology for the breast, however, is compromised by low spatial resolution, tissue heterogeneity, and interpatient variation. MATERIALS AND METHODS: We tested a dynamic near-infrared imaging schema for the characterization of suspicious breast lesions identified on diagnostic clinical ultrasound. A portable handheld near-infrared tissue imaging device (P-Scan; ViOptix Inc., Fremont, CA, USA) was utilized. An external mechanical compression force was applied to breast tissue. The tissue oxygen saturation and hemoglobin concentration were recorded simultaneously by the handheld near-infrared imaging device. Twelve categories of dynamic tissue parameters were derived based on real-time measurements of the tissue hemoglobin concentration and the oxygen saturation. RESULTS: Fifty suspicious breast lesions were evaluated in 48 patients. Statistical analyses were carried out on 36 out of 50 datasets that satisfied our inclusion criteria. Suspicious breast lesions identified on diagnostic clinical ultrasound had lower oxygenation and higher hemoglobin concentration than the surrounding normal breast tissue. Furthermore, histopathologic-proven malignant breast tumors had a lower differential hemoglobin contrast (that is, the difference of hemoglobin concentration variability between the suspicious breast lesion and the normal breast parenchyma located remotely elsewhere within the ipsilateral breast) as compared with histopathologic-proven benign breast lesions. CONCLUSION: The proposed dynamic near-infrared imaging schema has the potential to differentiate benign processes from those of malignant breast tumors. Further development and refinement of the dynamic imaging device and additional subsequent clinical testing are necessary for optimizing the accuracy of detection. |
format | Text |
id | pubmed-2246191 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2007 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-22461912008-06-04 A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions Xu, Ronald X Young, Donn C Mao, Jimmy J Povoski, Stephen P Breast Cancer Res Research Article INTRODUCTION: Characterizing and differentiating between malignant tumors, benign tumors, and normal breast tissue is increasingly important in the patient presenting with breast problems. Near-infrared diffuse optical imaging and spectroscopy is capable of measuring multiple physiologic parameters of biological tissue systems and may have clinical applications for assessing the development and progression of neoplastic processes, including breast cancer. The currently available application of near-infrared imaging technology for the breast, however, is compromised by low spatial resolution, tissue heterogeneity, and interpatient variation. MATERIALS AND METHODS: We tested a dynamic near-infrared imaging schema for the characterization of suspicious breast lesions identified on diagnostic clinical ultrasound. A portable handheld near-infrared tissue imaging device (P-Scan; ViOptix Inc., Fremont, CA, USA) was utilized. An external mechanical compression force was applied to breast tissue. The tissue oxygen saturation and hemoglobin concentration were recorded simultaneously by the handheld near-infrared imaging device. Twelve categories of dynamic tissue parameters were derived based on real-time measurements of the tissue hemoglobin concentration and the oxygen saturation. RESULTS: Fifty suspicious breast lesions were evaluated in 48 patients. Statistical analyses were carried out on 36 out of 50 datasets that satisfied our inclusion criteria. Suspicious breast lesions identified on diagnostic clinical ultrasound had lower oxygenation and higher hemoglobin concentration than the surrounding normal breast tissue. Furthermore, histopathologic-proven malignant breast tumors had a lower differential hemoglobin contrast (that is, the difference of hemoglobin concentration variability between the suspicious breast lesion and the normal breast parenchyma located remotely elsewhere within the ipsilateral breast) as compared with histopathologic-proven benign breast lesions. CONCLUSION: The proposed dynamic near-infrared imaging schema has the potential to differentiate benign processes from those of malignant breast tumors. Further development and refinement of the dynamic imaging device and additional subsequent clinical testing are necessary for optimizing the accuracy of detection. BioMed Central 2007 2007-12-18 /pmc/articles/PMC2246191/ /pubmed/18088411 http://dx.doi.org/10.1186/bcr1837 Text en Copyright © 2008 Xu et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an open access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Xu, Ronald X Young, Donn C Mao, Jimmy J Povoski, Stephen P A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions |
title | A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions |
title_full | A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions |
title_fullStr | A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions |
title_full_unstemmed | A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions |
title_short | A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions |
title_sort | prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2246191/ https://www.ncbi.nlm.nih.gov/pubmed/18088411 http://dx.doi.org/10.1186/bcr1837 |
work_keys_str_mv | AT xuronaldx aprospectivepilotclinicaltrialevaluatingtheutilityofadynamicnearinfraredimagingdeviceforcharacterizingsuspiciousbreastlesions AT youngdonnc aprospectivepilotclinicaltrialevaluatingtheutilityofadynamicnearinfraredimagingdeviceforcharacterizingsuspiciousbreastlesions AT maojimmyj aprospectivepilotclinicaltrialevaluatingtheutilityofadynamicnearinfraredimagingdeviceforcharacterizingsuspiciousbreastlesions AT povoskistephenp aprospectivepilotclinicaltrialevaluatingtheutilityofadynamicnearinfraredimagingdeviceforcharacterizingsuspiciousbreastlesions AT xuronaldx prospectivepilotclinicaltrialevaluatingtheutilityofadynamicnearinfraredimagingdeviceforcharacterizingsuspiciousbreastlesions AT youngdonnc prospectivepilotclinicaltrialevaluatingtheutilityofadynamicnearinfraredimagingdeviceforcharacterizingsuspiciousbreastlesions AT maojimmyj prospectivepilotclinicaltrialevaluatingtheutilityofadynamicnearinfraredimagingdeviceforcharacterizingsuspiciousbreastlesions AT povoskistephenp prospectivepilotclinicaltrialevaluatingtheutilityofadynamicnearinfraredimagingdeviceforcharacterizingsuspiciousbreastlesions |