Correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration

A popular technique to reduce respiratory motion for cardiovascular magnetic resonance is to perform a multi-slice acquisition in which a patient holds their breath multiple times during the scan. The feasibility of rigid slice-to-volume registration to correct for misalignments of slice stacks in s...

Descripción completa

Detalles Bibliográficos
Autores principales: Chandler, Adam G, Pinder, Richard J, Netsch, Thomas, Schnabel, Julia A, Hawkes, David J, Hill, Derek LG, Razavi, Reza
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2292180/
https://www.ncbi.nlm.nih.gov/pubmed/18312619
http://dx.doi.org/10.1186/1532-429X-10-13
_version_ 1782152477991239680
author Chandler, Adam G
Pinder, Richard J
Netsch, Thomas
Schnabel, Julia A
Hawkes, David J
Hill, Derek LG
Razavi, Reza
author_facet Chandler, Adam G
Pinder, Richard J
Netsch, Thomas
Schnabel, Julia A
Hawkes, David J
Hill, Derek LG
Razavi, Reza
author_sort Chandler, Adam G
collection PubMed
description A popular technique to reduce respiratory motion for cardiovascular magnetic resonance is to perform a multi-slice acquisition in which a patient holds their breath multiple times during the scan. The feasibility of rigid slice-to-volume registration to correct for misalignments of slice stacks in such images due to differing breath-hold positions is explored. Experimental results indicate that slice-to-volume registration can compensate for the typical misalignments expected. Correction of slice misalignment results in anatomically more correct images, as well as improved left ventricular volume measurements. The interstudy reproducibility has also been improved reducing the number of samples needed for cardiac MR studies.
format Text
id pubmed-2292180
institution National Center for Biotechnology Information
language English
publishDate 2008
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-22921802008-04-11 Correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration Chandler, Adam G Pinder, Richard J Netsch, Thomas Schnabel, Julia A Hawkes, David J Hill, Derek LG Razavi, Reza J Cardiovasc Magn Reson Research A popular technique to reduce respiratory motion for cardiovascular magnetic resonance is to perform a multi-slice acquisition in which a patient holds their breath multiple times during the scan. The feasibility of rigid slice-to-volume registration to correct for misalignments of slice stacks in such images due to differing breath-hold positions is explored. Experimental results indicate that slice-to-volume registration can compensate for the typical misalignments expected. Correction of slice misalignment results in anatomically more correct images, as well as improved left ventricular volume measurements. The interstudy reproducibility has also been improved reducing the number of samples needed for cardiac MR studies. BioMed Central 2008-02-29 /pmc/articles/PMC2292180/ /pubmed/18312619 http://dx.doi.org/10.1186/1532-429X-10-13 Text en Copyright © 2008 Chandler et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Chandler, Adam G
Pinder, Richard J
Netsch, Thomas
Schnabel, Julia A
Hawkes, David J
Hill, Derek LG
Razavi, Reza
Correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration
title Correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration
title_full Correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration
title_fullStr Correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration
title_full_unstemmed Correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration
title_short Correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration
title_sort correction of misaligned slices in multi-slice cardiovascular magnetic resonance using slice-to-volume registration
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2292180/
https://www.ncbi.nlm.nih.gov/pubmed/18312619
http://dx.doi.org/10.1186/1532-429X-10-13
work_keys_str_mv AT chandleradamg correctionofmisalignedslicesinmultislicecardiovascularmagneticresonanceusingslicetovolumeregistration
AT pinderrichardj correctionofmisalignedslicesinmultislicecardiovascularmagneticresonanceusingslicetovolumeregistration
AT netschthomas correctionofmisalignedslicesinmultislicecardiovascularmagneticresonanceusingslicetovolumeregistration
AT schnabeljuliaa correctionofmisalignedslicesinmultislicecardiovascularmagneticresonanceusingslicetovolumeregistration
AT hawkesdavidj correctionofmisalignedslicesinmultislicecardiovascularmagneticresonanceusingslicetovolumeregistration
AT hilldereklg correctionofmisalignedslicesinmultislicecardiovascularmagneticresonanceusingslicetovolumeregistration
AT razavireza correctionofmisalignedslicesinmultislicecardiovascularmagneticresonanceusingslicetovolumeregistration