An Optimized Spline-Based Registration of a 3D CT to a Set of C-Arm Images

We have developed an algorithm for the rigid-body registration of a CT volume to a set of C-arm images. The algorithm uses a gradient-based iterative minimization of a least-squares measure of dissimilarity between the C-arm images and projections of the CT volume. To compute projections, we use a n...

Descripción completa

Detalles Bibliográficos
Autores principales: Jonić, S., Thévenaz, P., Zheng, G., Nolte, L. -P., Unser, M.
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2324037/
https://www.ncbi.nlm.nih.gov/pubmed/23165033
http://dx.doi.org/10.1155/IJBI/2006/47197
Descripción
Sumario:We have developed an algorithm for the rigid-body registration of a CT volume to a set of C-arm images. The algorithm uses a gradient-based iterative minimization of a least-squares measure of dissimilarity between the C-arm images and projections of the CT volume. To compute projections, we use a novel method for fast integration of the volume along rays. To improve robustness and speed, we take advantage of a coarse-to-fine processing of the volume/image pyramids. To compute the projections of the volume, the gradient of the dissimilarity measure, and the multiresolution data pyramids, we use a continuous image/volume model based on cubic B-splines, which ensures a high interpolation accuracy and a gradient of the dissimilarity measure that is well defined everywhere. We show the performance of our algorithm on a human spine phantom, where the true alignment is determined using a set of fiducial markers.