Clinical Evaluation of the Peroral Cholangioscopy Using a New Videoscope

Peroral cholangioscopy (PCS) has been performed in 22 cases using XCHF-B200 (Olympus Optical Co.) since June 1995 and in 77 cases using CHF-B20 (Olympus Optical Co.) after EST from Jan. 1989. XCHF-B200 has a longer rigid portion of distal end and a smaller channel diameter than CHF-B20. The successf...

Descripción completa

Detalles Bibliográficos
Autores principales: Igarashi, Yoshinori, Ukita, Takeo, Inoue, Hirokazu, Ishiguro, Jun, Ogawa, Satoshi, Satou, Masahiro, Maetani, Iruru, Sakai, Yoshihiro
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 1999
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2362641/
https://www.ncbi.nlm.nih.gov/pubmed/18493506
http://dx.doi.org/10.1155/DTE.5.231
Descripción
Sumario:Peroral cholangioscopy (PCS) has been performed in 22 cases using XCHF-B200 (Olympus Optical Co.) since June 1995 and in 77 cases using CHF-B20 (Olympus Optical Co.) after EST from Jan. 1989. XCHF-B200 has a longer rigid portion of distal end and a smaller channel diameter than CHF-B20. The successful rate of PCS using XCHF-B200 (82%) was lower than that of CHF-B20 (89%). The vascular pattern and fine vertical groove of the bile duct mucosa were shown more clearly on the photographs obtained with XCHF-B200 than those obtained with CHF-B20. However, not enough biopsy specimens could be obtained because the channel diameter of XCHF-B200 was rather small. If the length of rigid portion and biopsy channel of XCHF-B200 are improved, PCS using XCHF-B200 will be more useful for the diagnosis of bile duct disorders.