Cargando…

Population-based study of diagnostic assays for Borrelia infection: comparison of purified flagella antigen assay (Ideia™, Dako Cytomation) and recombinant antigen assay (Liaison(®), DiaSorin)

BACKGROUND: Testing for Borrelia-specific IgM and IgG-antibodies are often performed on a variety of poorly defined symptoms, and isolated IgM results are a frequent finding, which results in diagnostic uncertainty and further testing. We wanted to test the hypothesis that Borrelia-specific assays u...

Descripción completa

Detalles Bibliográficos
Autores principales: Petersen, Eskild, Tolstrup, Martin, Capuano, Francesco, Ellermann-Eriksen, Svend
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2365948/
https://www.ncbi.nlm.nih.gov/pubmed/18423001
http://dx.doi.org/10.1186/1472-6890-8-4
Descripción
Sumario:BACKGROUND: Testing for Borrelia-specific IgM and IgG-antibodies are often performed on a variety of poorly defined symptoms, and isolated IgM results are a frequent finding, which results in diagnostic uncertainty and further testing. We wanted to test the hypothesis that Borrelia-specific assays using recombinant antigens perform differently from assays based on purified flagella antigen. METHODS: We compared the use of recombinant antigens (LIAISON(® )DiaSorin, Saluggia, Italy) and purified flagella antigen (IDEIA™ Borrelia, DakoCytomation, Glostrup, Denmark) in the assay for Borrelia-specific IgM and IgG-antibodies. The assays were tested on an unselected population of serum samples submitted from general practice. A total of 357 consecutive samples for analysis of Borrelia IgM and IgG antibodies. Furthermore, we analysed 540 samples for Borrelia-specific IgM or IgG antibodies first by the IDEIA™ and, if they were positive, the samples were further analysed using the LIAISON(® )assay. To verify the correctness of the patient's serological status, discrepant samples were analysed by line blots (EcoLine, Virotech). RESULTS: In the consecutive series of 357 samples, the IgM assays detected 308 negative and 3 positive samples with concordant results. Compared with the line blot, the IDEIA™ system produced 21 false-positive IgM results, whereas the LIAISON(® )system produced only one false-positive IgM result. The IgG assays showed 1 positive and 328 negative concordant results. The LIAISON(® )system produced 9 true IgG-positive samples that were not detected by the IDEIA™ system, but the former produced 4 positive IgG results that were negative by line blot. CONCLUSION: Diagnostic assays based on flagella antigen seem to show more false-positive IgM and false-negative IgG results than assays based on recombinant antigens. The latter may reduce the number of presumably false-positive IgM results and identify more IgG-positive subjects, but this system also produces more false-positive IgG results.