Cargando…
Alcohol based surgical prep solution and the risk of fire in the operating room: a case report
A few cases of fire in the operating room are reported in the literature. The factors that may initiate these fires are many and include alcohol based surgical prep solutions, electrosurgical equipment, flammable drapes etc. We are reporting a case of fire in the operating room while operating on a...
Autores principales: | , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2377238/ https://www.ncbi.nlm.nih.gov/pubmed/18439304 http://dx.doi.org/10.1186/1754-9493-2-10 |
_version_ | 1782154797388922880 |
---|---|
author | Batra, Sumit Gupta, Rajiv |
author_facet | Batra, Sumit Gupta, Rajiv |
author_sort | Batra, Sumit |
collection | PubMed |
description | A few cases of fire in the operating room are reported in the literature. The factors that may initiate these fires are many and include alcohol based surgical prep solutions, electrosurgical equipment, flammable drapes etc. We are reporting a case of fire in the operating room while operating on a patient with burst fracture C6 vertebra with quadriplegia. The cause of the fire was due to incomplete drying of the covering drapes with an alcohol based surgical prep solution. This paper discusses potential preventive measures to minimize the incidence of fire in the operating room. |
format | Text |
id | pubmed-2377238 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2008 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-23772382008-05-13 Alcohol based surgical prep solution and the risk of fire in the operating room: a case report Batra, Sumit Gupta, Rajiv Patient Saf Surg Case Report A few cases of fire in the operating room are reported in the literature. The factors that may initiate these fires are many and include alcohol based surgical prep solutions, electrosurgical equipment, flammable drapes etc. We are reporting a case of fire in the operating room while operating on a patient with burst fracture C6 vertebra with quadriplegia. The cause of the fire was due to incomplete drying of the covering drapes with an alcohol based surgical prep solution. This paper discusses potential preventive measures to minimize the incidence of fire in the operating room. BioMed Central 2008-04-26 /pmc/articles/PMC2377238/ /pubmed/18439304 http://dx.doi.org/10.1186/1754-9493-2-10 Text en Copyright © 2008 Batra and Gupta; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Batra, Sumit Gupta, Rajiv Alcohol based surgical prep solution and the risk of fire in the operating room: a case report |
title | Alcohol based surgical prep solution and the risk of fire in the operating room: a case report |
title_full | Alcohol based surgical prep solution and the risk of fire in the operating room: a case report |
title_fullStr | Alcohol based surgical prep solution and the risk of fire in the operating room: a case report |
title_full_unstemmed | Alcohol based surgical prep solution and the risk of fire in the operating room: a case report |
title_short | Alcohol based surgical prep solution and the risk of fire in the operating room: a case report |
title_sort | alcohol based surgical prep solution and the risk of fire in the operating room: a case report |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2377238/ https://www.ncbi.nlm.nih.gov/pubmed/18439304 http://dx.doi.org/10.1186/1754-9493-2-10 |
work_keys_str_mv | AT batrasumit alcoholbasedsurgicalprepsolutionandtheriskoffireintheoperatingroomacasereport AT guptarajiv alcoholbasedsurgicalprepsolutionandtheriskoffireintheoperatingroomacasereport |