Performance of an electron gun for a high-brightness X-ray generator

A prototype thermionic electron gun for a high-brightness X-ray generator has been developed. Its extraction voltage and design current are 60 kV and 100 mA (DC), respectively. The X-ray generator aims towards a maximum brilliance of 60 kW mm(−2). The beam sizes at the rotating anticathode must ther...

Descripción completa

Detalles Bibliográficos
Autores principales: Sugimura, Takashi, Ohsawa, Satoshi, Ikeda, Mitsuo
Formato: Texto
Lenguaje:English
Publicado: International Union of Crystallography 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2394796/
https://www.ncbi.nlm.nih.gov/pubmed/18421153
http://dx.doi.org/10.1107/S0909049507066769
Descripción
Sumario:A prototype thermionic electron gun for a high-brightness X-ray generator has been developed. Its extraction voltage and design current are 60 kV and 100 mA (DC), respectively. The X-ray generator aims towards a maximum brilliance of 60 kW mm(−2). The beam sizes at the rotating anticathode must therefore be within 1.0 mm × 0.1 mm and a small beam emittance is required. The fabricated electron gun optimizes an aperture grid and a Whenelt electrode. The performance of the prototype electron gun measured using pulsed-beam tests is as follows: maximum beam current, 85.7 mA; beam focus size at the rotating anticathode, 0.79 mm × 0.13 mm. In DC beam tests, FWHM beam sizes were measured to be 0.65 mm × 0.08 mm at the rotating anticathode with a beam current of 45 mA. The beam current recently reached ∼60 mA with some thermal problems.