Affordable pediatric CD4 counting by flow cytometry in Malawi

BACKGROUND: Rapid expansion of antiretroviral therapy in Malawi has occurred in the relative absence of suitable pediatric CD4 counting facilities. We have recently validated in adults a simplified affordable flow cytometric CD4 counting method, the Blantyre count. There is a need for this technolog...

Descripción completa

Detalles Bibliográficos
Autores principales: MacLennan, Calman A, Dzumani, Felix, Namarika, Alinane, Moons, Peter, Senga, Edward, Molyneux, Malcolm E, Drayson, Mark T, Bunn, James E G
Formato: Texto
Lenguaje:English
Publicado: Wiley Subscription Services, Inc., A Wiley Company 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2408853/
https://www.ncbi.nlm.nih.gov/pubmed/18307249
http://dx.doi.org/10.1002/cyto.b.20411
Descripción
Sumario:BACKGROUND: Rapid expansion of antiretroviral therapy in Malawi has occurred in the relative absence of suitable pediatric CD4 counting facilities. We have recently validated in adults a simplified affordable flow cytometric CD4 counting method, the Blantyre count. There is a need for this technology to transfer to government laboratories run by local staff, and to be validated in children, where %CD4/lymphocyte values are required. METHODS: We assessed agreement of %CD4/lymphocyte values determined by the Blantyre count and Panleucogate methods on an EPICS XL-MCL flow cytometer on 113 venous blood samples from HIV-seropositive children in Blantyre, Malawi. All assays were performed by two Malawian laboratory technicians. RESULTS: Overall bias between the two methods was −0.13% (95% CI −0.37 to 0.11) and limits of agreement were −2.69 to 2.43% (95% CI −3.11 to −2.27 and 2.01 to 2.85). Limits of agreement were within −3.00 and 3.00 for each laboratory technician. Coefficient of variation for the Blantyre count assay was 2.0% and samples showed good stability over 5 days. CONCLUSIONS: The Blantyre count method can accurately determine %CD4/lymphocyte values in blood of HIV-seropositive children on an EPIC XL-MCL flow cytometer at a reagent cost of US $0.21 per test or less. The assay can be competently carried out by local laboratory technicians.