Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla

PURPOSE: To evaluate MRI issues at 3-Tesla for a ventricular assist device (VAD). METHODS: The AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached (Abiomed, Inc., Danvers, MA) was evaluated for magnetic field interactions, heating, and artifacts at...

Descripción completa

Detalles Bibliográficos
Autores principales: Shellock, Frank G, Valencerina, Samuel
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2440740/
https://www.ncbi.nlm.nih.gov/pubmed/18495028
http://dx.doi.org/10.1186/1532-429X-10-23
_version_ 1782156568548081664
author Shellock, Frank G
Valencerina, Samuel
author_facet Shellock, Frank G
Valencerina, Samuel
author_sort Shellock, Frank G
collection PubMed
description PURPOSE: To evaluate MRI issues at 3-Tesla for a ventricular assist device (VAD). METHODS: The AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached (Abiomed, Inc., Danvers, MA) was evaluated for magnetic field interactions, heating, and artifacts at 3-Tesla. MRI-related heating was assessed with the device in a gelled-saline-filled, head/torso phantom using a transmit/received RF body coil while performing MRI at a whole body averaged SAR of 3-W/kg for 15-min. Artifacts were assessed for the main metallic component of this VAD (atrial cannula) using T1-weighted, spin echo and gradient echo pulse sequences. RESULTS: The AB5000 Ventricle with the prototype In-Flow Cannula and Out-Flow Cannula attached showed relatively minor magnetic field interactions that will not cause movement in situ. Heating was not excessive (highest temperature change, +0.8°C). Artifacts may create issues for diagnostic imaging if the area of interest is in the same area or close to the implanted metallic component of this VAD (i.e., the venous cannula). CONCLUSION: The results of this investigation demonstrated that it would be acceptable for a patient with this VAD (AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached) to undergo MRI at 3-Tesla or less. Notably, it is likely that the operation console for this device requires positioning a suitable distance (beyond the 100 Gauss line or in the MR control room) from the 3-Tesla MR system to ensure proper function of the VAD.
format Text
id pubmed-2440740
institution National Center for Biotechnology Information
language English
publishDate 2008
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-24407402008-06-27 Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla Shellock, Frank G Valencerina, Samuel J Cardiovasc Magn Reson Technical Notes PURPOSE: To evaluate MRI issues at 3-Tesla for a ventricular assist device (VAD). METHODS: The AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached (Abiomed, Inc., Danvers, MA) was evaluated for magnetic field interactions, heating, and artifacts at 3-Tesla. MRI-related heating was assessed with the device in a gelled-saline-filled, head/torso phantom using a transmit/received RF body coil while performing MRI at a whole body averaged SAR of 3-W/kg for 15-min. Artifacts were assessed for the main metallic component of this VAD (atrial cannula) using T1-weighted, spin echo and gradient echo pulse sequences. RESULTS: The AB5000 Ventricle with the prototype In-Flow Cannula and Out-Flow Cannula attached showed relatively minor magnetic field interactions that will not cause movement in situ. Heating was not excessive (highest temperature change, +0.8°C). Artifacts may create issues for diagnostic imaging if the area of interest is in the same area or close to the implanted metallic component of this VAD (i.e., the venous cannula). CONCLUSION: The results of this investigation demonstrated that it would be acceptable for a patient with this VAD (AB5000 Ventricle with a prototype Nitinol wire-reinforced In-Flow Cannula and Out-Flow Cannula attached) to undergo MRI at 3-Tesla or less. Notably, it is likely that the operation console for this device requires positioning a suitable distance (beyond the 100 Gauss line or in the MR control room) from the 3-Tesla MR system to ensure proper function of the VAD. BioMed Central 2008-05-21 /pmc/articles/PMC2440740/ /pubmed/18495028 http://dx.doi.org/10.1186/1532-429X-10-23 Text en Copyright © 2008 Shellock and Valencerina; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Shellock, Frank G
Valencerina, Samuel
Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla
title Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla
title_full Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla
title_fullStr Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla
title_full_unstemmed Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla
title_short Ventricular Assist Device implant (AB 5000) prototype cannula: In vitro assessment of MRI issues at 3-Tesla
title_sort ventricular assist device implant (ab 5000) prototype cannula: in vitro assessment of mri issues at 3-tesla
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2440740/
https://www.ncbi.nlm.nih.gov/pubmed/18495028
http://dx.doi.org/10.1186/1532-429X-10-23
work_keys_str_mv AT shellockfrankg ventricularassistdeviceimplantab5000prototypecannulainvitroassessmentofmriissuesat3tesla
AT valencerinasamuel ventricularassistdeviceimplantab5000prototypecannulainvitroassessmentofmriissuesat3tesla